用光学生物测量法测量气流脉冲诱导的眼部各组件动力学
研究纳入20名健康受试者的20只眼,用扫频光源光学相干断层扫描生物测量仪联合气流脉冲系统测量角膜、晶状体和视网膜的动力学,并用Goldmann压平眼压计测量眼压。结果显示角膜厚度、玻璃体深度和眼轴长度与眼压呈统计学显著相关,角膜和晶状体的形变幅度与眼压成反比,但仅角膜达到统计学显著性;晶状体无压缩位移,回复呈摆动形式。使用0.2%酒石酸溴莫尼定滴眼液降眼压后,提取参数发生相应变化。
要点
- 角膜厚度、玻璃体深度和眼轴长度与眼压呈统计学显著相关。
- 角膜和晶状体形变幅度与眼压成反比,仅角膜达到统计学显著性。
- 晶状体无压缩位移,回复呈摆动形式;降眼压后提取参数发生相应变化。
核心结论
- 光学生物测量联合气流脉冲可提供眼部所有组件(包括晶状体)在活体中的行为信息。
- 测量形变过程中的眼轴长度动力学可校正眼球回缩对形变的影响。
研究设计与样本
- 研究纳入20名健康受试者的20只眼,使用扫频光源光学相干断层扫描生物测量仪联合气流脉冲系统,在同一位置随时间采集系列轴向扫描,无横向扫描。
- 眼压用Goldmann压平眼压计测量,在给予0.2%酒石酸溴莫尼定滴眼液降眼压前和2小时后进行测量。
主要发现
- 角膜厚度、玻璃体深度和眼轴长度与眼压呈统计学显著相关。
- 角膜和晶状体的形变幅度与眼压成反比,但仅角膜达到统计学显著性。
- 晶状体无压缩位移,回复呈摆动形式。
- 眼压水平降低引起提取参数发生相应变化。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To analyze the dynamics of all optical components of the eye and the behavior of the eyeball under air-puff conditions in vivo. To determine the impact of the intraocular pressure (IOP) on the air-puff-induced deformation of the eye. METHODS: Twenty eyes of 20 healthy subjects were included in this study. The dynamics of the ocular components, such as the cornea, the crystalline lens, and the retina, was measured by a prototype swept source optical coherence tomography biometer integrated with the air-puff system. The system allows to acquire a series of axial scans at the same location as a function of time with no transverse scanning. Several parameters were extracted from optical coherence tomography data. The IOP was measured using a Goldmann applanation tonometry. The measurements of the eyes were performed before and 2 hours after administration of IOP-reducing drops, namely, 0.2 % brimonidine tartrate. RESULTS: There is a statistically significant correlation of corneal thickness, vitreous depth, and eye length with IOP. The deformation amplitudes of the cornea and the crystalline lens are inversely proportional to the IOP, but statistical significance is achieved only for the cornea. The crystalline lens is displaced without compression, and the return has the form of wobbling. The reduction of IOP level induces corresponding changes in the extracted parameters. CONCLUSIONS: Optical biometry combined with air puff provides comprehensive information on the in vivo behavior of all ocular components, including the crystalline lens. Measurement of the axial length dynamics of during deformation enables correcting the deformation for eye retraction.
引用
- Maczynska E, Rzeszewska-Zamiara J, Jimenez Villar A, et al. Air-Puff-Induced Dynamics of Ocular Components Measured with Optical Biometry. Invest Ophthalmol Vis Sci. 2019. PMID: 31050724. doi:10.1167/iovs.19-26681