四维显微镜集成光学相干断层扫描在模型眼视网膜下手术中的引导作用
本研究在离体猪眼上评估四维显微镜集成光学相干断层扫描(4D-MIOCT)引导视网膜下注射、视网膜切开、膜植入及器械操作的价值。结果显示4D-MIOCT可显示常规显微镜下不明显的工具、膜、视网膜与视网膜色素上皮之间的相互作用,有助于术者调整操作以避免并发症。作者认为4D-MIOCT可用于模型视网膜下手术的连续监测,并可能有助于避免技术相关并发症。
要点
- 4D-MIOCT以每秒3.3–10个体积的速度提供抬头显示横断面和立体体积视图。
- 在离体猪眼中,4D-MIOCT可显示常规显微镜下常无法检测的工具、膜、视网膜与视网膜色素上皮之间的相互作用。
- 4D-MIOCT有助于监测诱导的视网膜脱离范围、引导套管尖端和器械操作、识别细胞流入玻璃体腔,使术者调整操作以避免并发症。
核心结论
- 4D-MIOCT可用于模型视网膜下手术的连续监测,并可能有助于避免技术相关并发症。
- 4D-MIOCT可促进正在开发的视网膜下治疗递送技术的进步。
- 4D-MIOCT是引导模型视网膜下手术操作的有价值工具,是对常规显微镜的重要补充,有助于提高视网膜下治疗递送的精确性并帮助避免手术技术相关并发症。
研究设计与样本
- 本研究为实验性研究,在离体猪眼上评估4D-MIOCT引导视网膜下手术操作的价值。
- 4D-MIOCT系统通过显微镜目镜显示抬头显示横断面和立体体积视图,速度为每秒3.3–10个体积。
主要发现
- 在离体猪眼中,工具、膜、视网膜与视网膜色素上皮之间的相互作用在常规显微镜视图下常无法检测,但在4D-MIOCT下清晰可见。
- 4D-MIOCT可视化对于监测诱导的视网膜脱离范围、引导套管尖端和器械在视网膜下空间的操作、识别细胞流入玻璃体腔是必要的。
- 4D-MIOCT提供的信息使术者能够调整操作以避免并发症,例如剪刀刮擦视网膜色素上皮。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE:: Subretinal delivery of stem cells and gene therapy vectors hold great promise in clinical care. Delivery techniques are evolving. We investigate the utility of 4-dimensional (volumes over time) microscope-integrated optical coherence tomography (4D-MIOCT) to guide such maneuvers. METHODS:: This investigational 4D-MIOCT system displayed heads-up cross-sectional and stereoscopic volumetric views through the microscope oculars at 3.3–10 volumes/second. In enucleated porcine eyes, we evaluated novel information available to the surgeon during subretinal: 1) injection of fluid and cell suspensions, 2) retinotomy creation, 3) insertion of membranes with a retinal pigment epithelium (RPE) cell monolayer and 4) subretinal instrument maneuvers, all in postmortem porcine eyes. RESULTS:: Interactions between tools, membranes, retina and RPE were often not detectable on the conventional microscope view but were obvious with 4D-MIOCT. 4D-MIOCT visualization was essential to: monitor the extent of induced retinal detachment, guide the cannula tip and instrument maneuvers in the subretinal space, and identify efflux of cells into the vitreous cavity. This allowed adjustment by the surgeon to avoid complications—e.g. scissors scraping RPE. CONCLUSIONS:: 4D-MIOCT is useful for continuous monitoring of model subretinal surgery and may help avoid technique-related complications. 4D-MIOCT can contribute to the advancement of subretinal delivery of therapies under development. 4-dimensional (volumes over time) microscope-integrated optical coherence tomography (4D MIOCT) is valuable tool for guidance of model subretinal surgical maneuvers. It is an important addition to the conventional microscope that helps with precision of subretinal therapy delivery and helps avoid surgical technique-related complications.
引用
- Vajzovic L, Sleiman K, Viehland C, et al. Four-Dimensional Microscope-Integrated Optical Coherence Tomography Guidance in a Model Eye Subretinal Surgery. Retina. 2019. PMID: 31090685. doi:10.1097/IAE.0000000000002518
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。