文献速递角膜与眼表疾病

使用新型泪膜成像仪(TFI)动态评估泪膜黏液-水层与脂质层

Dynamic assessment of the tear film muco-aqueous and lipid layers using a novel tear film imager (TFI).
PubMed 文献 · Br J Ophthalmol · 2019-04-18

前瞻性初步病例系列纳入49例干眼与非干眼个体,用新型泪膜成像仪(TFI)连续两次成像评估黏液-水层厚度(MALT)与脂质层破裂时间(LBUT)的可重复性及其与临床干眼指标的相关性。MALT可重复性极好(r=0.88),与Schirmer评分显著相关(r=0.31),LBUT与临床泪膜破裂时间显著相关(r=0.73);干眼组MALT更薄、LBUT更短,设备敏感性87%、特异性88%。

要点

  • TFI是首台能在人体可重复测量黏液-水层厚度的设备,MALT可重复性r=0.88。
  • TFI测得的LBUT与临床TBUT显著相关(r=0.73),MALT与Schirmer评分显著相关(r=0.31)。
  • 干眼组(轻中度与重度)MALT较对照组更薄、LBUT较对照组更短,设备敏感性87%、特异性88%。

核心结论

  • TFI是首台能在人体受试者中可重复测量黏液-水层厚度且与Schirmer评分相关的设备。
  • TFI同时评估泪膜功能的其他重要方面,这些方面与临床医生评估的干眼指标相关。

研究设计与样本

  • 前瞻性初步病例系列,纳入有或无干眼的个体。
  • 使用TFI连续获取两张图像,评估测量可重复性并与临床来源的干眼指标比较。
  • 根据干眼体征将个体分组为干眼类别。

主要发现

  • 共49例患者参与,平均年龄58.8岁(SD 15.9),女性占多数(69%)。
  • 黏液-水层厚度(MALT)可重复性极好(r=0.88)。
  • MALT测量与Schirmer评分显著相关(r=0.31)。
  • TFI测得的脂质层破裂时间(LBUT)与临床泪膜破裂时间(TBUT)显著相关(r=0.73)。
  • 与对照组相比,干眼组(轻中度与重度)MALT显著更薄、LBUT显著更短。

关键数值与推荐

  • TFI参数与临床评估体征比较,设备敏感性为87%、特异性为88%。
  • MALT与Schirmer评分相关(r=0.31),LBUT与TBUT相关(r=0.73)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The objective of the study was to assess a new technology, the tear film imager (TFI), which can dynamically image the muco-aqueous and lipid layers. METHODS: Prospective pilot case series of individuals with and without dry eye (DE). Two sequential images were obtained with the TFI. Measurements were assessed for reproducibility and compared with clinically derived DE metrics. Individuals were grouped into DE categories based on signs of DE. RESULTS: 49 patients participated in the study with a mean age of 58.8 years (SD 15.9) and a female majority (69%). Reproducibility of the muco-aqueous layer thickness (MALT) was excellent (r=0.88). MALT measurements significantly correlated with the Schirmer score (r=0.31). Lipid break up time (LBUT) as measured by the TFI significantly correlated with the clinical measure of tear break up time (TBUT) (r=0.73). MALT and LBUT were significantly thinner and shorter, respectively, in the DE groups (mild-moderate and severe) compared with the control group. When comparing TFI parameters to clinically assessed signs, sensitivity of the device was 87% and specificity was 88%. CONCLUSION: The TFI is the first machine capable of reproducibly measuring muco-aqueous thickness in human subjects which correlates with Schirmer score. In parallel, it assesses other important aspects of tear film function which correlate with clinician assessed DE metrics.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Segev F, Geffen N, Galor A, et al. Dynamic assessment of the tear film muco-aqueous and lipid layers using a novel tear film imager (TFI). Br J Ophthalmol. 2019. PMID: 31000512. doi:10.1136/bjophthalmol-2018-313379

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。