文献速递角膜与眼表疾病青光眼

Boston角膜假体术后Eyemate-IO遥测眼压第一年动态变化

Telemetric Intraocular Pressure Monitoring after Boston Keratoprosthesis Surgery Using the Eyemate-IO Sensor: Dynamics in the First Year.
PubMed 文献 · Am J Ophthalmol · 2019-03-05

这项前瞻性多中心单臂临床试验纳入12例接受Boston I型角膜假体植入并植入Eyemate-IO遥测眼压系统的患者,随访12个月、计划13次访视。遥测眼压在术后第1天达峰值23.1±16.5 mmHg,第5天为16.0±5.2 mmHg,6-12个月为20.95±6.5 mmHg;遥测眼压与两位专家指测分组间差异有统计学意义(P<.001),提示遥测眼压可识别术后眼压峰值与纵向升高。

要点

  • 遥测眼压术后第1天峰值23.1±16.5 mmHg,第5天16.0±5.2 mmHg。
  • 6-12个月遥测眼压均值20.95±6.5 mmHg,可识别日间波动与峰值。
  • 指测四分组间遥测眼压差异显著(P<.001),一致性满意。

核心结论

  • 遥测眼压评估似能识别BI-KPro术后眼压峰值及纵向升高,并与两位专家指测结果满意一致。

研究设计与样本

  • 前瞻性、开放标签、多中心、单臂临床试验(ARGOS,NCT02945176),12例受试者接受Eyemate-IO眼内系统植入,术后随访12个月、计划每人13次访视。

主要发现

  • 术前Goldmann眼压为13.4±6.2 mmHg,遥测眼压在术后第1天达峰值23.1±16.5 mmHg。
  • 术后第5天遥测眼压均值16.0±5.2 mmHg,6-12个月为20.95±6.5 mmHg。
  • 指测A组遥测眼压18.2±6.1 mmHg,B组8.9±2.8 mmHg,C组22.4±4.9 mmHg,D组34.3±11.0 mmHg,组间差异有统计学意义(P<.001)。
  • 遥测系统可识别日间眼压波动与峰值。

关键数值与推荐

  • 指测分组:正常为A,软/低眼压为B,临界为C,高眼压为D。
  • 各指测分组间遥测眼压均值差异有统计学意义(P<.001)。

临床怎么用

  • BI-KPro术后可考虑使用Eyemate-IO遥测系统辅助识别眼压峰值与纵向升高。
  • 遥测眼压与两位专家指测结果满意一致,可作为指测的补充评估手段。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To analyze the dynamics of telemetrically measured intraocular pressure (IOP) during the first year after implantation of a Boston keratoprosthesis type I (BI-KPro) cornea and to compare agreement of telemetric IOP measurements with finger palpations. DESIGN: Prospective, open-label, multicenter, single-arm clinical trial. METHODS: In the ARGOS (NCT02945176) study, 12 individuals underwent implantation of an Eyemate-IO intraocular system. Follow-up after surgery took place 12 months later with 13 visits planned per patient. During BI-KPro surgery, an electromagnetic induction sensor ring enabling telemetric IOP data transfer to a hand-held reading device outside the eye was implanted into the ciliary sulcus with or without trans-scleral suture fixation. Comprehensive ophthalmic examinations and IOP assessments through the telemetric system were compared to IOP assessed by finger palpation by 2 experts. RESULTS: Preoperative IOP measured by Goldmann tonometry was 13.4 ± 6.2 mm Hg. Telemetric IOP peaked at 23.1 ± 16.5 mm Hg at the first postoperative day. On day 5, mean IOP was 16.0 ± 5.2 mm Hg and 20.95 ± 6.5 mm Hg after 6-12 months. IOP estimation by finger palpation was grouped in 4 categories: normal, A; soft/hypotonic, B; borderline, C; and hypertonic, D. Mean telemetric IOP was 18.2 ± 6.1 mm Hg in category A, 8.9 ± 2.8 mm Hg in B, 22.4 ± 4.9 mm Hg in C, and 34.3 ± 11.0 mm Hg in D. Differences in mean telemetric IOPs per category were statistically significant (P < .001). Daily IOP fluctuations and peaks could be identified. CONCLUSIONS: Telemetric IOP assessment seems to be able to identify postoperative IOP peaks and a longitudinal increase of IOP after BI-KPro surgery. IOP measurements using the telemetric Eyemate-IO sensor showed a satisfactory agreement with those of finger palpations by 2 experts.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Enders P, Hall J, Bornhauser M, et al. Telemetric Intraocular Pressure Monitoring after Boston Keratoprosthesis Surgery Using the Eyemate-IO Sensor: Dynamics in the First Year. Am J Ophthalmol. 2019. PMID: 30849343. doi:10.1016/j.ajo.2019.02.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。