正常角膜与圆锥角膜的布里渊光谱研究

Brillouin Spectroscopy of Normal and Keratoconus Corneas.
PubMed 文献 · Am J Ophthalmol · 2019-02-15

回顾性病例对照研究纳入47只健康眼和36只圆锥角膜眼,用布里渊光谱测量全角膜厚度平均频移。

要点

  • 布里渊光谱检测圆锥角膜的敏感性和特异性不及Kmax和最薄点厚度

核心结论

  • 布里渊光谱可无创提供角膜垂直方向的生物力学信息,但当前区分圆锥角膜与正常角膜的敏感性和特异性不足。

主要发现

  • 最大后表面隆起点的频移与几何学圆锥角膜指数相关性最好。
  • 布里渊频移的ROC曲线分析显示其检测圆锥角膜的敏感性和特异性明显差于Kmax和最薄点厚度。

临床怎么用

  • 布里渊光谱可作为角膜生物力学状态的补充信息,但尚不能单独用于圆锥角膜的精准分型。

局限性

  • 方法目前敏感性和特异性不足,需标准化 mapping 和新指标开发以提升潜力。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To investigate the age dependence of Brillouin spectroscopy of the cornea and to compare normal and keratoconus corneas. DESIGN: Retrospective case-control study. METHODS: Study Population: Healthy patients and patients suffering from keratoconus seen at the Institut für Refraktive und Ophthalmo-Chirurgie (IROC) between December 2016 and March 2017. Brillouin frequency shifts of patients of 2 different groups were examined with Brillouin spectroscopy perpendicular to the corneal surface. Group 1 consisted of 47 healthy eyes, whereas Group 2 included 36 eyes with keratoconus of unclear progression. Besides Brillouin examinations, corneal tomographies were acquired so that correlations and comparisons between geometric parameters and Brillouin frequency shifts could be evaluated. MAIN OUTCOME MEASURES: Corneal Brillouin frequency shifts averaged over full corneal thickness. RESULTS: A significant correlation between age and central Brillouin frequency shift was identified (P = .011) with an increase in Brillouin frequency shift of 4 MHz per decade in normal corneas. Keratoconus corneas have a significantly reduced Brillouin frequency shift at the thinnest point compared to normal corneas (5.7072 ± 0.0214 vs 5.7236 ± 0.0146 GHz, P < .001). The Brillouin frequency shift at the point of maximum posterior elevation showed best correlation with geometry-derived keratoconus indices. The receiver operating characteristic curve analysis of Brillouin frequency shift showed substantially worse sensitivity and specificity compared to Kmax and thinnest pachymetry for keratoconus detection. CONCLUSION: Noninvasive Brillouin spectroscopy adds clinical information about the biomechanical state of the cornea perpendicular to the surface. An age-dependent stiffening of the cornea has been found and keratoconus corneas are statistically significantly different from normal corneas, but for precise differentiating of keratoconus stages (including normal corneas) the method is currently neither specific nor sensitive enough. Further development including standardized mapping and establishment of new indices may increase the potential of Brillouin spectroscopy substantially.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Seiler TG, Shao P, Eltony A, et al. Brillouin Spectroscopy of Normal and Keratoconus Corneas. Am J Ophthalmol. 2019. PMID: 30772345. doi:10.1016/j.ajo.2019.02.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。