改良Goldmann棱镜眼压测量准确性及其与角膜生物力学指标的相关性:多中心随机临床试验

Modified Goldmann prism intraocular pressure measurement accuracy and correlation to corneal biomechanical metrics: multicentre randomised clinical trial.
PubMed 文献 · Br J Ophthalmol · 2019-02-22

这项前瞻性、开放标签、随机对照多中心参考设备准确性分析在243只眼上比较标准GAT棱镜与改良表面Goldmann棱镜的眼压测量差异,并分析其与中央角膜厚度(CCT)和角膜滞后量(CH)的相关性。结果显示两种棱镜测量差值CCT显著相关(R2=0.404,p=0.007),CH亦显著相关(R2=0.125,p=0.039),总体平均眼压偏倚无统计学意义(改良棱镜+0.43 mm Hg,p=0.19)。提示改良替换棱镜可能通过减少CCT和CH相关角膜生物力学误差提高眼压测量准确性。

要点

  • 标准与改良Goldmann棱镜眼压测量差值CCT显著相关(R2=0.404,p=0.007)。
  • CH与两棱镜测量差值亦显著相关(R2=0.125,p=0.039)。
  • 两棱镜总体平均眼压偏倚无统计学意义(+0.43 mm Hg,p=0.19)。

核心结论

  • 标准GAT与改良表面Goldmann棱镜的眼压测量差值CCT和CH均呈统计学显著相关。
  • 简单改良替换棱镜可能通过最小化CCT和CH相关角膜生物力学误差改善眼压测量准确性。

研究设计与样本

  • 前瞻性、开放标签、随机对照、多中心参考设备准确性分析。
  • 243只独特眼接受GAT和改良表面GAT棱镜眼压测量,设计与方法遵循ISO眼压计评估指南,但纳入薄(<500 µm)和厚(>600 µm)角膜。
  • 所有眼随机分配至五种标准Goldmann棱镜和五种改良棱镜之一测量,由六名研究者每人每棱镜测量两次,共1936次眼压测量。

主要发现

  • 标准与改良Goldmann棱镜眼压测量差值CCT相关良好,尤其在薄(<500 µm)和厚(>600 µm)角膜(R2=0.404,p=0.007)。
  • CH亦与棱镜测量差值显著相关(R2=0.125,p=0.039)。
  • 两棱镜间总体平均眼压偏倚无统计学意义(改良棱镜+0.43 mm Hg,p=0.19)。

关键数值与推荐

  • 薄角膜定义为<500 µm,厚角膜定义为>600 µm。
  • CCT与测量差值回归R2=0.404,p=0.007;CH与测量差值回归R2=0.125,p=0.039。
  • 改良棱镜平均眼压偏倚+0.43 mm Hg,p=0.19。

临床怎么用

  • 对于薄(<500 µm)或厚(>600 µm)角膜患者,改良表面Goldmann棱镜可能减少CCT相关眼压测量误差。
  • CH异常眼可考虑使用改良棱镜以降低角膜生物力学对眼压测量的影响。
  • 改良棱镜可作为Goldmann型眼压计的简单替换部件用于临床。

局限性

  • 研究为开放标签设计,未设盲。
  • 研究纳入薄和厚角膜,偏离ISO指南标准纳入范围,可能影响结果外推。
  • 摘要未报告长期随访或不同操作者间重复性数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Clinically evaluate intraocular pressure (IOP) measurements taken with a Goldmann applanation tonometer (GAT) prism and a modified surface Goldmann prism examining measurement differences correlated to central corneal thickness (CCT) and corneal hysteresis (CH) values. DESIGN: Prospective, open-label, randomised, controlled, multicentre reference device accuracy analysis. METHODS: A GAT and a modified surface GAT prism measured IOP on 243 unique eyes. The study design and methodology complied with International Standard Organization (ISO) tonometer evaluation guidelines, except the inclusion of thin (<500 µm) and thick (>600 µm) corneas. All eyes were randomised to IOP measurement by one of five standard Goldmann prisms and five modified prisms. Pressures were measured by six investigators, two times with each prism for a total of 1936 IOP measurements. Analysis included a multiple linear regression including CCT and CH correlation. RESULTS: The difference in IOP measurements of the standard and modified Goldmann prisms correlated well to CCT particularly in thin (<500 µm) and thick (>600 µm) corneas (R2=0.404, p=0.007). Corneal hysteresis (CH) also significantly correlated to the difference in prism measurements (R2=0.125, p=0.039). There was no significant overall mean IOP bias between the two prisms (+0.43 mm Hg in modified, p=0.19). DISCUSSION: The paired IOP measurement difference between GAT and a modified surface Goldmann replacement prism indicated a statistically significant correlation to CCT and CH. A simple modified replacement prism for any Goldmann-type tonometer may significantly improve IOP measurement accuracy by minimising corneal biomechanical errors associated with CCT and CH. TRIAL REGISTRATION NUMBER: NCT02990169 and NCT02989909.