OCT血管成像鉴别视乳头水肿与假性视乳头水肿

Optical Coherence Tomography Angiography in Papilledema Compared With Pseudopapilledema.
PubMed 文献 · Invest Ophthalmol Vis Sci · 2019-01-01

一项前瞻性比较研究纳入视乳头水肿41眼、假性视乳头水肿27眼及健康对照44眼,使用OCT-A测量毛细血管密度。商业软件显示的视乳头周围血管值在两组患者中均低于健康眼,但无法区分二者;定制软件显示假性视乳头水肿的全图像及鼻侧毛细血管密度显著低于视乳头水肿,提示OCT-A有鉴别诊断价值。

要点

  • 商业OCT-A血管密度图无法区分视乳头水肿与假性视乳头水肿
  • 定制软件测量的全图像和鼻侧毛细血管密度可鉴别真性与假性视盘肿胀
  • 假性视乳头水肿的视乳头周围毛细血管密度显著低于健康对照

核心结论

  • 使用OCT-A测量的全图像和鼻侧视乳头周围毛细血管密度对鉴别真性与假性视盘肿胀具有诊断准确性。

研究设计与样本

  • 前瞻性比较研究纳入视乳头水肿41眼(21例)、假性视乳头水肿27眼(15例)和健康对照44眼(44例),均行OCT-A成像。

主要发现

  • 视乳头水肿眼平均RNFL厚度大于假性视乳头水肿和对照组,三组间GCC厚度无差异。
  • 商业软件测量的视乳头周围血管值在视乳头水肿(58.5±6.1%)和假性视乳头水肿(58.9±4.7%)均显著低于健康眼(63.2±3.1%),但两组间无差异。
  • 定制软件测量的视乳头周围毛细血管密度在视乳头水肿为29.8±9.4%,与健康受试者(31.8±7.4%;P=0.94)无显著差异;假性视乳头水肿为25.5±8.3%,显著低于对照眼(P=0.01)。
  • 假性视乳头水肿眼内视网膜全图像和鼻侧视乳头周围毛细血管密度显著低于视乳头水肿眼(分别为P=0.03和P=0.02)。

关键数值与推荐

  • 视乳头水肿眼平均RNFL厚度大于假性视乳头水肿和对照组。
  • 三组间GCC厚度无差异。
  • 商业软件视乳头周围血管值:视乳头水肿58.5±6.1%,假性视乳头水肿58.9±4.7%,健康眼63.2±3.1%。
  • 定制软件视乳头周围毛细血管密度:视乳头水肿29.8±9.4%,健康受试者31.8±7.4%(P=0.94);假性视乳头水肿25.5±8.3%,与对照眼比较P=0.01。
  • 假性视乳头水肿与视乳头水肿比较:全图像毛细血管密度P=0.03,鼻侧视乳头周围毛细血管密度P=0.02。

临床怎么用

  • 临床可借助OCT-A定制软件分析全图像和鼻侧视乳头周围毛细血管密度,辅助鉴别视乳头水肿与假性视乳头水肿。
  • 商业OCT-A血管密度图在两组患者中均降低且重叠,不宜单独用于区分真性与假性视盘肿胀。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The purpose of this study is to evaluate differences in optical coherence tomography angiography (OCT-A) findings between patients with papilledema and pseudopapilledema. METHODS: In this prospective, comparative study, 41 eyes of 21 subjects with papilledema, 27 eyes of 15 subjects with pseudopapilledema, and 44 eyes of 44 healthy normal subjects were included and were imaged using OCT-A. In addition to peripapillary total vasculature maps obtained with commercial vessel density mapping, major vessel removal using customized image analysis software was also used to measure whole image capillary density and peripapillary capillary density (PCD). Peripapiilary retinal nerve fiber layer (RNFL) and macular ganglion cell complex (GCC) were recorded. RESULTS: Average RNFL thicknesses were greater in papilledema eyes than in pseudopapilledema and control subjects. GCC thickness was not different among three groups. Peripapillary vasculature values were significantly lower in papilledema (58.5 ± 6.1%) and pseudopapilledema (58.9 ± 4.7%) eyes compared with healthy eyes (63.2 ± 3.1%) using commercial machine software, without a difference between papilledema and pseudopapilledema eyes. However, using our customized software, peripapillary "capillary" density of papilledema eyes was 29.8 ± 9.4%, which was not significantly different from healthy subjects (31.8 ± 7.4%; P = 0.94). Pseudopapilledema eyes with peripapillary density of 25.5 ± 8.3% had significantly lower capillary values compared with control eyes (P = 0.01). There was a significantly lower whole image and nasal sector peripapillary capillary density of inner retina in pseudopapilledema eyes than papilledema eyes (P = 0.03 and P = 0.02, respectively). CONCLUSIONS: Whole image and nasal peripapillary sector capillary densities using OCT-A had diagnostic accuracy for differentiating true and pseudo-disc swelling.