巩膜接触镜配戴者镜后泪膜午间雾浊与白细胞相关
46%的眼中出现午间雾浊,雾浊眼的巩膜镜角膜间隙为308±98 μm,高于无雾浊眼的246±61 μm(P<0.05)。镜碗中回收的白细胞数(9551±18,926)多于眼表冲洗液(2195±4384,P<0.02),角膜间隙与雾浊存在关联(OR=2.24,95%CI=1.48-3.38,P<0.001),镜后泪膜白细胞以中性粒细胞为主。
要点
- 巩膜镜配戴者镜后泪膜中存在白细胞,以中性粒细胞为主。
- 午间雾浊眼巩膜镜角膜间隙大于无雾浊眼(308±98 μm vs 246±61 μm,P<0.05)。
- 角膜间隙增大与午间雾浊存在关联(OR=2.24,95%CI=1.48-3.38,P<0.001)。
核心结论
- 巩膜接触镜配戴者镜后泪膜中存在以中性粒细胞为主的白细胞,且角膜间隙较大的配戴者出现午间雾浊的 odds 更高。
主要发现
- 46%的眼中出现午间雾浊。
- 无雾浊眼巩膜镜角膜间隙为246±61 μm,雾浊眼为308±98 μm(P<0.05)。
- 镜碗回收白细胞数平均为9551±18,926,多于眼表冲洗液的2195±4384(P<0.02)。
- 巩膜镜角膜间隙与午间雾浊存在关联,OR为2.24(95%CI=1.48-3.38,P<0.001)。
临床怎么用
- 对巩膜接触镜配戴者出现午间雾浊时,可关注镜后泪膜白细胞及角膜间隙情况。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Midday fogging is a frequent complaint among scleral contact lens (ScCL) wearers, and the mechanism and cause of this is unknown. The purpose of this investigation was to understand the relation between midday fogging, ocular surface leukocytes, and ScCL fitting characteristics. METHODS: Subjects arrived at a clinical exam having worn ScCLs for at least 4 hours. ScCL were removed, and 150 μL of phosphate-buffered saline (PBS) was used to wash the bowl of the ScCL. Eyes were washed post-ScCL removal with 5 mL PBS per eye. Wash solutions were collected and leukocytes were then isolated and counted, followed by assessment with flow cytometry. Samples from the post-lens tear fluid were stained with fluorescently labeled antibodies to detect leukocyte distributions. RESULTS: Thirty-nine eyes from 19 adapted, full-time, ScCL wearers were included, and 46% presented with midday fogging. ScCL corneal clearance was 246 ± 61 μm for nonfoggers, while it was 308 ± 98 μm for those with fogging (P < 0.05). On average, the number of leukocytes collected from the ScCL bowl (9551 ± 18,926) was greater than the number of leukocytes recovered from the eye wash (2195 ± 4384, P < 0.02). ScCL corneal clearance was associated with the presence of fogging, with an odds ratio of 2.24 (95% confidence interval = 1.48-3.38, P < 0.001). CONCLUSIONS: Leukocytes, predominated by neutrophils, are present in the post-lens tear film of ScCL wearers, and in particular wearers with greater ScCL corneal clearance have greater odds of having midday fogging.
引用
- Postnikoff CK, Pucker AD, Laurent J, et al. Identification of Leukocytes Associated With Midday Fogging in the Post-Lens Tear Film of Scleral Contact Lens Wearers. Invest Ophthalmol Vis Sci. 2019. PMID: 30646011. doi:10.1167/iovs.18-24664
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。