文献速递角膜与眼表疾病

圆锥角膜角膜上皮厚度测量重复性及其与视力受限的关系

Corneal Epithelial Thickness Intrasubject Repeatability and its Relation With Visual Limitation in Keratoconus.
PubMed 文献 · Am J Ophthalmol · 2019-01-26

本研究评估AS-OCT联合Placido盘MS-39测量角膜上皮厚度的个体内重复性,并分析圆锥角膜患者上皮厚度与最佳矫正视力的相关性。共纳入167眼(圆锥角膜107眼、健康60眼),重复性高(组内相关系数≥0.90),上皮厚度指标与BCVA及圆锥角膜分级显著相关。

要点

  • MS-39测量角膜上皮厚度在圆锥角膜与健康眼均具有高重复性(组内相关系数≥0.90)。
  • 中央3mm变薄、上方8mm增厚、下方8mm变薄与更差BCVA相关。
  • 上皮厚度指标联合K2判别圆锥角膜的曲线下面积为0.92。

核心结论

  • MS-39在圆锥角膜与健康眼均可实现高重复性的上皮厚度测量,上皮厚度指标与BCVA显著相关,可能有助于判别圆锥角膜与健康眼。

研究设计与样本

  • 可靠性分析与横断面研究,地点为西班牙阿利坎特的Vissum,共纳入167眼,其中圆锥角膜患者107眼、健康眼60眼。
  • 连续三次重复测量上皮厚度,包括中央3.0mm及鼻侧、颞侧、上方、下方3.0-6.0mm和6.0-8.0mm区域。
  • 主要结局为上皮指标个体内重复性及其与最佳矫正视力(BCVA)的相关性。

主要发现

  • 组内相关系数≥0.90,提示圆锥角膜与健康眼测量均高度可重复。
  • 多因素分析中与更差BCVA相关的上皮指标为:中央3mm更薄(P = .04)、上方8mm更厚(P < .001)、下方8mm更薄(P < .001)。
  • 中央3mm厚度随圆锥角膜分级升高而减小(P = .002),8mm上方-下方比值随分级升高而显著增大(P < .001)。
  • 多因素分析中中央3mm更薄(R2 = 6.19%,P = .04)、8mm上方-下方比值更大(R2 = 8.37%,P = .004)、6mm上方-下方比值更大(R2 = 2.67%,P = .02)及K2更陡(R2 = 21.51%,P < .001)与圆锥角膜相关,联合曲线下面积为0.92。

关键数值与推荐

  • 组内相关系数≥0.90为高度可重复的判定界值。
  • 中央3mm更薄P = .04、上方8mm更厚P < .001、下方8mm更薄P < .001与更差BCVA相关。
  • 中央3mm厚度随分级升高而减小P = .002,8mm上方-下方比值随分级升高而增大P < .001。
  • 联合判别模型曲线下面积为0.92。

临床怎么用

  • 可参考MS-39测量的上皮厚度指标辅助评估圆锥角膜患者的视力受限程度与分级。
  • 上皮厚度指标联合K2或有助于在临床中判别圆锥角膜与健康眼。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate intra-subject repeatability of anterior segment optical coherence tomography (AS-OCT) combined with placido disc MS-39 (CSO, Firenze, Italy) and correlate epithelial thickness measurements and the degree of visual limitation in keratoconus patients. DESIGN: Reliability analysis and cross-sectional study. METHODS: Setting: Vissum, Alicante, Spain. SUBJECTS: Total of 167 eyes, of which 107 were from keratoconus patients and 60 were healthy. PROCEDURES: Three repeated consecutive epithelial thickness measurements were taken: central 3.0 mm and the nasal, temporal, superior, and inferior regions at 3.0-6.0 mm and 6.0-8.0 mm. MAIN OUTCOME MEASURES: Intrasubject repeatability of epithelial indices was assessed and correlated to the degree of best-corrected visual acuity (BCVA). RESULTS: The intraclass correlation coefficient was ≥0.90, indicating highly repeatable measurements for both keratoconus and healthy eyes. In multivariate analysis, epithelial indices associated with worse BCVA were as follows: thinner 3 mm central (P = .04), thicker 8 mm superior (P < .001), and thinner 8 mm inferior (P < .001). Thickness central 3 mm decreased as keratoconus grading increased (P = .002). The superior-inferior ratio at 8 mm significantly increased as keratoconus grading increased (P < .001). In multivariate analysis thinner 3 mm central (R2 = 6.19%, P = .04), greater superior-inferior 8 mm ratio (R2 = 8.37%, P = .004), greater superior-inferior 6mm ratio (R2 = 2.67%, P = .02), and steeper K2 (R2 = 21.51%, P < .001) were associated with keratoconus with a combined area under the curve of 0.92. CONCLUSION: High repeatability of epithelial thickness measurements may be achieved in both keratoconus and healthy eyes with the MS-39. There is a significant correlation between epithelial thickness measurements and BCVA. These measurements may be of additional value in discriminating between keratoconus and healthy eyes.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Vega-Estrada A, Mimouni M, Espla E, et al. Corneal Epithelial Thickness Intrasubject Repeatability and its Relation With Visual Limitation in Keratoconus. Am J Ophthalmol. 2019. PMID: 30689987. doi:10.1016/j.ajo.2019.01.015

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。