自动检测圆锥角膜交联后基质分界线的OCT软件评估

Automated Detection of the Stromal Demarcation Line Using Optical Coherence Tomography in Keratoconus Eyes After Corneal Cross-linking.
PubMed 文献 · Am J Ophthalmol · 2018-12-06

本研究比较自动检测软件与人工操作者在圆锥角膜交联术后3个月OCT图像上评估基质分界线存在与深度的表现,纳入25眼。自动软件与两名操作者测得的平均深度分别为321.54±47.71、322.86±45.77和309.21±40.98 μm,组内相关系数(ICC)分别为0.884和0.847(P<.001)。结论认为人工检测可重复,自动软件可进一步优化和标准化分界线检测。

要点

  • 自动软件与人工操作者检测分界线深度的ICC分别为0.884和0.847(P<.001)。
  • 操作者间重现性ICC为0.890,操作者1和2的重复性ICC分别为0.922和0.925。
  • 软件间重复性ICC为1,Bland-Altman图显示软件与观察者一致性良好。

核心结论

  • 人工操作者检测分界线具有可重复性和重现性,但超快准确的自动软件可进一步优化和标准化检测,为治疗成功提供可靠指标。

研究设计与样本

  • 可靠性分析研究,纳入25例圆锥角膜交联术后3个月患者(25眼),采用Dresden方案,由两名独立操作者和一个自动检测软件评估角膜OCT图像。
  • 操作者在间隔1周的两次独立场合评估OCT图像(128个切面)中分界线的存在并测量深度,自动软件检查每次OCT测量的全部128个切面。

主要发现

  • 自动检测软件和人工操作者计算的平均角膜分界线深度分别为321.54±47.71、322.86±45.77和309.21±40.98 μm。
  • 自动检测软件与观察者1和2之间的ICC分别为0.884和0.847(P<.001)。
  • 操作者间重现性ICC为0.890,观察者1和2的操作者内重复性ICC分别为0.922和0.925。
  • 软件间重复性ICC为1,Bland-Altman图显示两名观察者与软件之间一致性良好,95%一致性界限适当。

临床怎么用

  • 自动检测软件可作为快速、准确、标准化的工具用于圆锥角膜交联术后OCT图像中基质分界线的评估。
  • 人工检测分界线具有可重复性和重现性,但自动软件可能进一步优化和标准化该评估过程。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate the role of a novel automated detection software as compared to human operators in assessing the presence and depth of stromal demarcation line on optical coherence tomography (OCT) in keratoconus eyes post cross-linking. DESIGN: Reliability analysis study. METHODS: Two independent operators and an automated detection software examined corneal OCTs of 25 eyes of 25 patients post corneal cross-linking using the Dresden protocol, at 3 months postoperatively. Operators evaluated the presence of the demarcation line and measured its depth by looking at OCT images (128 cuts) on 2 separate occasions 1 week apart. The automated software examined all 128 cuts of each OCT measurement. RESULTS: The mean corneal demarcation line depth was 321.54 ± 47.71, 322.86 ± 45.77, and 309.21 ± 40.98 μm, as computed by the automated detection software and the human operators, respectively. The intraclass correlation coefficients (ICC) between the automated detection software and Observers #1 and #2 were 0.884 and 0.847, respectively (P < .001). The ICC for interoperator reproducibility was 0.890, and for intraoperator repeatability for Operator #1 and Observer #2 were 0.922 and 0.925, respectively. The ICC for intersoftware repeatability was 1. Bland-Altman plots showed a good agreement between both observers and the software, with adequate 95% limits of agreement. CONCLUSION: Detection of the demarcation line by human operators is repeatable and reproducible, but it can be further optimized and standardized by an ultrafast and accurate automated software detection tool, providing a reliable indicator for treatment success.