swept-source OCT 评估圆周房角关闭的社区研究
在新加坡社区≥50岁无青光眼等病史的2027名晶状体眼中,SS-OCT(CASIA SS-1000)以房角镜为参考标准检测房角关闭,最终分析1857人(91.6%)。SS-OCT手动分级总体AUC为0.84(95%CI 0.81-0.88),≥35%圆周关闭定义的敏感度82.5%、特异度78.5%。结论认为SS-OCT检测房角关闭性能中等。
要点
- SS-OCT手动分级检测房角关闭总体AUC为0.84(95%CI 0.81-0.88)。
- ≥35%圆周关闭定义敏感度82.5%(75.3%-88.4%),特异度78.5%(76.5%-80.4%)。
- 与2象限房角镜定义的协议系数在0.88-0.89之间,一致性良好。
核心结论
- 在大型社区研究中,SS-OCT以房角镜为参考标准检测房角关闭表现为中等性能。
研究设计与样本
- 可靠性分析,连续纳入2027名≥50岁晶状体眼、无青光眼/激光/内眼手术/眼外伤史者,最终分析1857人(91.6%)。
- 房角镜由单一 trained ophthalmologist 完成,SS-OCT图像由单一看片者盲法分析。
主要发现
- 几乎90%受试者为华人,平均年龄61.8±6.7岁,女性占63.5%。
- SS-OCT手动分级对比房角镜的总体AUC为0.84(95%CI 0.81-0.88)。
- SS-OCT≥35%定义下房角关闭患病率为26.1%,AUC 0.80(0.77-0.84)。
关键数值与推荐
- SS-OCT图像上房角关闭定义为虹膜与巩膜突前方任何房角壁接触。
- 分析采用圆周房角关闭程度≥35%、≥50%、≥75%三个切点。
- 2象限房角镜定义与SS-OCT≥35%、≥50%、≥75%定义的协议系数分别为0.89(0.83-0.93)、0.88(0.842-0.93)、0.88(0.831-0.99)。
临床怎么用
- SS-OCT可作为社区房角关闭检测的辅助工具,但性能中等,不能完全替代房角镜。
局限性
- 受试者以华人为主,外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To evaluate the diagnostic performance of swept-source optical coherence tomography (SS-OCT, CASIA SS-1000; Tomey Corporation, Nagoya, Japan) for angle closure detection, in comparison with gonioscopy, in a community setting. DESIGN: Reliability analysis. METHODS: A total of 2027 phakic subjects aged ≥50 years, with no previous history of glaucoma, laser (including peripheral iridotomy), intraocular surgery, or ocular trauma, were consecutively recruited from a community polyclinic in Singapore. Gonioscopy was performed by a single trained ophthalmologist. SS-OCT angle scans, which obtain radial scans for the entire circumference of the angle, were analyzed by a single examiner, masked to the subject's clinical details. On SS-OCT images, angle closure was defined as contact between the iris and any part of the angle wall anterior to the scleral spur. Different cutoff values of the degree of circumferential angle closure (≥35%, ≥50%, and ≥75%) were taken for analysis to assess SS-OCT performance in detecting angle closure. RESULTS: A total of 1857 subjects (91.6%) were included in the final analysis after excluding poor-quality SS-OCT scans. Almost 90% of the subjects were Chinese, with a mean age of 61.8 ± 6.7 years, and more than half were women (63.5%). The overall AUC of SS-OCT manual grading against gonioscopy was 0.84 (95% confidence interval, 0.81-0.88). The prevalence of angle closure on SS-OCT was 26.1% for the ≥35% definition, with an area under the curve of 0.80 (0.77-0.84), sensitivity of 82.5% (75.3%-88.4%), and specificity of 78.5% (76.5%-80.4%). The first-order agreement coefficient statistics for the 2-quadrant gonioscopic definition of angle-closure with corresponding ≥35%, ≥50%, and ≥75% angle closure definitions for SS-OCT were good at 0.89 (0.83-0.93), 0.88 (0.842-0.93), and 0.88 (0.831-0.99), respectively. CONCLUSIONS: In this large community-based study, SS-OCT exhibited moderate performance for angle closure detection compared to gonioscopy as the reference standard.
引用
- Porporato N, Baskaran M, Tun TA, et al. Assessment of Circumferential Angle Closure with Swept-Source Optical Coherence Tomography: a Community Based Study. Am J Ophthalmol. 2018. PMID: 30502338. doi:10.1016/j.ajo.2018.11.015