利用锥形扫描偏振敏感OCT绘制角膜各层厚度图

Mapping of Corneal Layer Thicknesses With Polarization-Sensitive Optical Coherence Tomography Using a Conical Scan Pattern.
PubMed 文献 · Invest Ophthalmol Vis Sci · 2018-01-01

本研究使用中心波长1045 nm的定制偏振敏感OCT(PS-OCT)配合锥形扫描,对20名健康志愿者双眼进行至少三次成像,分割并绘制全角膜(角膜缘到角膜缘)的上皮、Bowman层和基质厚度图,并分析重复性。结果显示各层厚度测量重复性良好(上皮CV<3%,Bowman层CV<5%,基质CV<2%),上皮呈现上方变薄、下方增厚,Bowman层在中央7至8 mm直径内均匀,向周边变薄直至距中心4至5.5 mm偏心处消失。

要点

  • PS-OCT锥形扫描可高重复性绘制全角膜上皮、Bowman层和基质厚度图
  • 上皮厚度呈上方变薄、下方增厚(P<0.001)
  • Bowman层中央7至8 mm均匀,距中心4至5.5 mm偏心处消失

核心结论

  • PS-OCT结合锥形扫描是大视野、高重复性测定角膜各层厚度图的可行方法。

研究设计与样本

  • 使用中心波长1045 nm的定制PS-OCT系统配合锥形扫描进行图像采集。
  • 对20名健康志愿者的双眼各至少成像三次,以测试方法并量化重复性。

主要发现

  • 上皮厚度图显示上方显著变薄、下方增厚(P<0.001)。
  • Bowman层在角膜中央7至8 mm直径内均匀,向周边变薄直至距中心4至5.5 mm偏心处消失。
  • 受试者间厚度变异:上皮CV约8%,Bowman层CV约25%,基质CV约10%。
  • 厚度测量重复性:上皮CV<3%,Bowman层CV<5%,基质CV<2%。
  • 同一受试者左右眼角膜各层厚度间存在显著相关(P<0.001)。

关键数值与推荐

  • 上皮受试者间变异系数(CV)约8%。
  • Bowman层受试者间CV约25%。
  • 基质受试者间CV约10%。
  • 上皮厚度测量重复性CV<3%。
  • Bowman层厚度测量重复性CV<5%,基质CV<2%。

临床怎么用

  • 可用于健康角膜各层厚度的全视野、高重复性定量评估。
  • 上皮上方变薄、下方增厚及Bowman层周边消失的分布特征可作为参考基线。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: We demonstrate segmentation and mapping of corneal layers (epithelium, Bowman's layer, and stroma) across the entire cornea (limbus to limbus), using additional contrast provided by polarization-sensitive optical coherence tomography (PS-OCT) and analyze the reproducibility of the procedure. METHODS: A custom built PS-OCT system operating at 1045 nm central wavelength with conical scanning was used for image acquisition. Conical scanning allows for almost perpendicular beam incidence on the corneal surface and provides good signal quality over the entire field of view. Epithelium, Bowman's layer, and stroma were segmented using the additional contrast provided by PS-OCT. Thickness maps were computed and analyzed in sectors. Both eyes of 20 healthy volunteers were imaged at least three times to test this method and to quantify reproducibility. RESULTS: Thickness maps of the epithelium show significant (P < 0.001) superior thinning and an inferior thickening. Bowman's layer appears homogeneous within the central 7 to 8 mm diameter of the cornea and gets thinner toward the periphery until this layer disappears between 4 and 5.5 mm eccentricity from the center. Intersubject variations of the measured thicknesses of epithelium (coefficient of variation [CV] ∼8%), Bowman's layer (CV∼25%), and stroma (CV∼10%) were observed. Very good reproducibility of thickness measurements of epithelium (CV < 3%), Bowman's layer (CV < 5%), and stroma (CV < 2%) was found. Furthermore, a significant correlation (P < 0.001) between layer thicknesses of the right and left eyes of the same subject was found. CONCLUSIONS: PS-OCT with conical scanning is a feasible approach for determining thickness maps of corneal layers on a large field of view with high reproducibility.

来源:Invest Ophthalmol Vis Sci,查看原文。版权归原作者所有。