正常与病变角膜眼iVue SD-OCT角膜上皮厚度成像的可重复性与再现性
研究纳入60只眼(正常12只,干眼综合征、接触镜配戴、激光屈光术后及圆锥角膜各12只),用3台iVue SD-OCT重复扫描评估角膜上皮厚度成像。59只眼纳入最终分析,598次扫描中11.9%因图像质量被排除。iVue SD-OCT在正常及病变角膜眼各图区均显示良好的可重复性与再现性,但干眼综合征组再现性显著较低。
要点
- iVue SD-OCT在正常及病变角膜眼各图区均显示良好可重复性与再现性。
- 干眼综合征组再现性显著低于正常组(3.2%-5.5% vs 1.1%-2.9%)。
核心结论
- iVue SD-OCT在正常和病变角膜眼中,所有图区均提供良好的角膜上皮厚度成像可重复性与再现性。
- 干眼综合征患者的再现性显著低于正常受试者。
研究设计与样本
- 由设备指定操作者在3台iVue SD-OCT上对每名受试者进行3次重复扫描,基于以瞳孔为中心的6 mm直径角膜上皮厚度图划分空间区域评估可重复性与再现性。
主要发现
- 合格扫描527次中40次(7.6%)需手动编辑分割线。
- 可重复性与再现性结果相似,提示两组设备/操作者变异极小。
- 除干眼综合征组外,各角膜病患者亚组的可重复性与再现性均相似;干眼综合征组再现性显著较低(范围3.2%-5.5%),正常受试者为1.1%-2.9%。
关键数值与推荐
- 以瞳孔为中心6 mm直径角膜上皮厚度图划分空间区域评估可重复性(基于重复扫描随机误差)和再现性(包括随机误差及设备/操作者变异)。
- 扫描排除标准为图像质量问题,598次扫描中11.9%被排除。
临床怎么用
- iVue SD-OCT可用于正常及病变角膜眼的角膜上皮厚度成像评估。
- 解读干眼综合征患者的上皮厚度图时应考虑其再现性较低。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To evaluate the performance of the epithelial thickness mapping (ETM) of the iVue spectral-domain optical coherence tomography (SD-OCT) device (Optovue Inc, Fremont, California, USA) in normal and diseased cornea eyes. DESIGN: Reliability and validity analysis. METHODS: Sixty eyes of 60 subjects were recruited for the study, which included normal subjects (n = 12) and patients with corneal diseases (12 patients each: dry eye syndrome [DES], contact lens wear, post-laser refractive surgery, and keratoconus). Three repeated scans were acquired on 3 iVue SD-OCTs with device-designated operators from consented subjects. Each subject was scanned on each device. Repeatability (based on random error of repeated scans) and reproducibility (including the random error and the instrument/operator variability) were assessed based on spatial zones derived from a 6-mm-diameter corneal ETM centered on the pupil and compared between the groups. RESULTS: Fifty-nine eyes qualified for final analysis. Seventy-one of 598 acquired scans (11.9%) were excluded owing to scan quality concerns. The percentage of disqualified scans was similar across normal (10.7%) and diseased eyes (12.1%). Of 527 qualified scans, 40 (7.6%) scans required manual edits of the segmentation lines. Repeatability and reproducibility results were similar, indicating minimal device/operator variability for both groups. Repeatability and reproducibility were similar in all subgroups of cornea patients, excluding the DES group, for which reproducibility was significantly lower (range 3.2%-5.5% for DES patients and 1.1%-2.9% for normal subjects). CONCLUSIONS: The iVue SD-OCT provides good corneal ETM repeatability and reproducibility in normal and diseased cornea eyes through all map zones.
引用
- Sella R, Zangwill LM, Weinreb RN, et al. Repeatability and Reproducibility of Corneal Epithelial Thickness Mapping With Spectral-Domain Optical Coherence Tomography in Normal and Diseased Cornea Eyes. Am J Ophthalmol. 2018. PMID: 30240724. doi:10.1016/j.ajo.2018.09.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。