文献速递白内障与屈光手术青光眼

扫频源前节OCT与UBM测量老年性白内障前房及房角参数的比较

[Comparison of swept source anterior segment optical coherence tomography and ultrasound biomicroscopy in measurement of anterior chamber depth and anterior chamber angle data in age-related cataract patients].
PubMed 文献 · Zhonghua Yan Ke Za Zhi · 2018-01-01

前瞻性研究纳入45例(45只眼)老年性白内障患者,比较SS-AS-OCT与UBM测量前房深度及鼻、颞、上、下四方位AOD500和TIA500的差异与一致性。结果显示两种方法各参数差异均有统计学意义但相关性良好,Bland-Altman分析示一致性较好。

要点

  • SS-AS-OCT与UBM测量前房深度及四方位房角参数差异有统计学意义
  • 两法各参数均呈正相关(前房深度r=0.989,AOD500与TIA500各方位r均>0.95)
  • Bland-Altman示两法一致性较好,仅2.22%~6.67%测量值超出95%一致性界限

核心结论

  • SS-AS-OCT与UBM在老年性白内障患者前房深度和房角参数测量中存在差别,但两者相关性和一致性均较好。

研究设计与样本

  • 前瞻性研究,选取2016年11至12月首都医科大学附属北京同仁医院北京同仁眼科中心单纯老年性白内障患者45例(45只眼),男23例、女22例,年龄(51±19)岁。
  • 分别用SS-AS-OCT和UBM测量前房深度及鼻、颞、上、下4个方位的AOD500和TIA500。
  • 统计采用配对样本t检验、Pearson直线相关分析和Bland-Altman分析。

主要发现

  • SS-AS-OCT与UBM测得前房深度分别为(2.750±0.544)mm和(2.775±0.531)mm,差异有统计学意义(t=2.109,P=0.040),呈正相关(r=0.989,P<0.001)。
  • 鼻、颞、上、下4个方位AOD500和TIA500两法测量差异均有统计学意义(均P<0.001),且均呈正相关(AOD500 r=0.971、0.975、0.979、0.973;TIA500 r=0.972、0.957、0.956、0.970;均P<0.001)。

关键数值与推荐

  • AOD500:SS-AS-OCT鼻、颞、上、下方位分别为(0.451±0.201)mm、(0.499±0.229)mm、(0.390±0.215)mm、(0.441±0.226)mm,UBM分别为(0.421±0.194)mm、(0.465±0.207)mm、(0.355±0.208)mm、(0.423±0.215)mm,t=4.262、4.350、5.355、4.415。
  • TIA500:SS-AS-OCT鼻、颞、上、下方位分别为39.587°±14.698°、42.149°±14.138°、36.482°±15.887°、38.504°±14.805°,UBM分别为36.944°±13.790°、40.289°±13.041°、35.600°±15.215°、36.818°±14.931°,t=5.131、4.658、4.174、5.731。
  • Bland-Altman分析显示2.22%~6.67%的测量值在95%一致性界限外。

临床怎么用

  • 两法测量前房深度和房角参数相关性及一致性较好,临床上可互为参考,但结果存在系统差异,不宜直接混用数值。
  • 随访或对比同一患者房角参数时宜固定使用同一测量设备。

局限性

  • 仅纳入单纯老年性白内障患者45例(45只眼),样本量有限,结论外推需谨慎。
  • 原文未报告两种测量方法的具体操作细节及重复性数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objective: To explore the difference and agreement in the measurement of anterior chamber depth (ACD) and anterior chamber angle data in age-related cataract patients between swept source anterior segment optical coherence tomography (SS-AS-OCT) and ultrasound biomicroscopy (UBM). Methods: A prospective study. Forty-five eyes from 45 age-related cataract patients were selected at Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University from November to December 2016. There were 23 males and 22 females aged (51±19) years. SS-AS-OCT and UBM were used to measure the ACD, angle opening distance (AOD 500) and trabecular-iris angle (TIA 500). The paired samples t-test and Pearson linear correlation were used for statistical analysis, and the Bland-Altman analysis was used to evaluate the measurement consistency of the two tools. Results: The ACD measured by SS-AS-OCT and UBM was (2.750±0.544) mm and (2.775±0.531) mm, respectively, the difference was statistically significant (t=2.109, P=0.040), and the measurement results were positively correlated (r=0.989, P<0.001). In the nasal, temporal, upper and lower positions, the AOD 500 and TIA 500 values measured by SS-AS-OCT and UBM had a significant difference [AOD 500: (0.451±0.201) mm vs.(0.421±0.194) mm, (0.499±0.229) mm vs. (0.465±0.207) mm, (0.390±0.215) mm vs. (0.355±0.208) mm, (0.441±0.226) mm vs. (0.423±0.215) mm, t=4.262, 4.350, 5.355, 4.415; TIA 500: 39.587°±14.698°vs. 36.944°±13.790°, 42.149°±14.138°vs. 40.289°±13.041°, 36.482°±15.887°vs. 35.600°±15.215°, 38.504°±14.805°vs. 36.818°±14.931°, t=5.131, 4.658, 4.174, 5.731; all P<0.001] and a significant linear correlation (AOD 500: r=0.971, 0.975, 0.979, 0.973; TIA 500: r=0.972, 0.957, 0.956, 0.970; all P<0.001). The Bland-Altman method showed that the two modalities had comparable results for the ACD, AOD 500 and TIA 500 (2.22%-6.67% of the measured value outside the 95% consistency boundaries). Conclusion: SS-OCT and UBM show a good agreement in the measurement of ACD and anterior chamber angle data in age-related cataract patients, though there is difference between them in measurement results. (Chin J Ophthalmol, 2018, 54: 678-682). 目的: 探讨扫频源前节相干光层析成像术(SS-AS-OCT)与超声生物显微镜(UBM)测量前房深度及房角参数结果的一致性。 方法: 前瞻性研究。选取2016年11至12月在首都医科大学附属北京同仁医院北京同仁眼科中心就诊的单纯老年性白内障患者45例(45只眼),男性23例,女性22例,年龄(51±19)岁。分别用SS-AS-OCT和UBM测量其前房深度及鼻、颞、上、下4个方位的房角参数,包括距巩膜突500 μm的房角开放距离(AOD500)和小梁网-虹膜夹角(TIA500)。采用配对样本t检验比较两种仪器前房深度、AOD500和TIA500的测量结果,采用Pearson直线相关分析法分析两种仪器测量结果的相关性,采用Blant-Altman分析评价两种仪器测量结果的一致性。 结果: SS-AS-OCT和UBM测得的前房深度分别为(2.750±0.544)mm和(2.775±0.531)mm,差异有统计学意义(t=2.109,P=0.040),二者测量结果呈正相关(r=0.989,P<0.001)。SS-AS-OCT与UBM对鼻、颞、上、下4个方位的AOD500和TIA500测量结果差异均有统计学意义[AOD500:SS-AS-OCT检测值分别为(0.451±0.201)mm、(0.499±0.229)mm、(0.390±0.215)mm、(0.441±0.226)mm,UBM检测值分别为(0.421±0.194)mm、(0.465±0.207)mm、(0.355±0.208)mm、(0.423±0.215)mm,二者比较的t值分别为4.262、4.350、5.355、4.415;TIA500:SS-AS-OCT检测值分别为39.587°±14.698°、42.149°±14.138°、36.482°±15.887°、38.504°±14.805°,UBM检测值分别为36.944°±13.790°、40.289°±13.041°、35.600°±15.215°、36.818°±14.931°,二者比较的t值分别为5.131、4.658、4.174、5.731;均P<0.001],且二者测量结果呈正相关(AOD500:r值分别为0.971、0.975、0.979、0.973;TIA500:r值分别为0.972、0.957、0.956、0.970;均P<0.001)。Blant-Altman分析显示两种测量方法有较好的一致性(2.22%~6.67%的测量值在95%的一致性界限外)。 结论: SS-AS-OCT与UBM在测量前房深度和房角参数中存在差别,但两者的一致性和相关性均较好。(中华眼科杂志,2018,54:678-682).

来源:Zhonghua Yan Ke Za Zhi,查看原文。版权归原作者所有。

引用

  1. Shen L, Wang XN, Li DJ, et al. [Comparison of swept source anterior segment optical coherence tomography and ultrasound biomicroscopy in measurement of anterior chamber depth and anterior chamber angle data in age-related cataract patients]. Zhonghua Yan Ke Za Zhi. 2018. PMID: 30220183. doi:10.3760/cma.j.issn.0412-4081.2018.09.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。