改良Goldmann棱镜眼压测量准确性与可重复性多中心随机试验

Intraocular Pressure Measurement Accuracy and Repeatability of a Modified Goldmann Prism: Multicenter Randomized Clinical Trial.
PubMed 文献 · Am J Ophthalmol · 2018-09-07

这项前瞻性、开放标签、随机对照多中心研究在173只 unique 眼上比较了标准Goldmann压平眼压计(GAT)棱镜与改良表面GAT棱镜的眼压测量准确性和可重复性。结果显示两种棱镜的Bland-Altman分析均无超出±5 mmHg指南的测量对,操作者与生产批次重复性无统计学差异(所有P > .10),改良棱镜与Dresdner中央角膜厚度(CCT)校正的相关性有统计学意义(P = .01)。结论认为改良表面替换棱镜与平面棱镜在统计学上等效,且对CCT的校正有统计学意义,需在ANSI标准限值外(0.500 mm < CCT < 0.600 mm)进一步测试。

要点

  • 改良表面GAT棱镜与标准棱镜眼压测量在±5 mmHg指南内一致,无超出配对的Bland-Altman分析。
  • 操作者内和操作者间以及生产批次重复性F检验和单因素方差分析均无统计学差异(所有P > .10)。
  • 标准与改良棱镜眼压测量差异与Dresdner CCT校正显著相关(P = .01)。

核心结论

  • 改良表面替换棱镜与平面棱镜在统计学上等效。
  • 改良表面棱镜对CCT的校正有统计学意义,需在ANSI标准限值外(0.500 mm < CCT < 0.600 mm)进一步测试以考察其全部潜力。

研究设计与样本

  • 前瞻性、开放标签、随机对照多中心参考器械信度与效度分析。
  • 研究设计和分析符合FDA 510(k)和ANSI Z80.10-2014指南。
  • 共173只 unique 眼随机分配至5个标准棱镜或5个改良棱镜(各来自不同生产批次)进行眼压测量。
  • 由6名研究者使用每种棱镜测量2次,共1384次眼压测量。

主要发现

  • Bland-Altman分析显示无眼压测量对超出±5 mmHg指南。
  • 操作者与生产批次重复性F检验和单因素方差分析显示标准与改良棱镜间无统计学差异(所有P > .10)。
  • 标准与改良棱镜眼压测量差异与Dresdner中央角膜厚度(CCT)校正良好相关(P = .01)。

关键数值与推荐

  • Bland-Altman分析以±5 mmHg为指南界限,无测量对超出。
  • 重复性F检验和单因素方差分析所有P > .10。
  • 标准与改良棱镜眼压测量差异与Dresdner CCT校正相关P = .01。
  • 需进一步测试的CCT范围为0.500 mm < CCT < 0.600 mm(ANSI标准限值外)。

临床怎么用

  • 改良表面GAT棱镜可作为标准平面棱镜的等效替换用于眼压测量。
  • 在CCT处于0.500 mm至0.600 mm范围外时,改良棱镜的CCT校正潜力需进一步验证。

局限性

  • 需在ANSI标准限值外(0.500 mm < CCT < 0.600 mm)进一步测试以考察改良棱镜的全部潜力。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To clinically evaluate a modified surface Goldmann applanation tomometer (GAT) prism for intraocular pressure (IOP) accuracy, repeatability, and safety. DESIGN: Prospective, open-label, randomized, controlled, multicenter reference device reliability and validity analysis. METHODS: A GAT and a modified surface GAT prism measured IOP on 173 unique eyes. The study design and analysis complied with FDA 510(k) and ANSIZ80.10-2014 guidelines. All eyes were randomized to IOP measurement by 1 of 5 standard prisms or 5 modified prisms, each from a different manufacturing lot. Pressures were measured by 6 investigators, 2 times with each prism, for a total of 1384 IOP measurements. Analysis included Bland-Altman difference accuracy, intraoperator and interoperator IOP measurement, and manufactured lot repeatability. RESULTS: Bland-Altman indicated no IOP measurements pairs outside the ±5 mm Hg guidelines. Operator and manufactured lot repeatability F tests and 1-way ANOVAs indicated no statistical difference between the standard and modified prisms (all P > .10). The difference in IOP measurements of the standard and modified prisms correlated well to Dresdner central corneal thickness (CCT) correction (P = .01). CONCLUSION: A modified surface replacement prism is statistically equivalent to a flat-surfaced prism. The modified surface prism indicated statistically significant correction for CCT requiring further testing outside the ANSI standard limits (0.500 mm < CCT < 0.600 mm) to examine its full potential.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。