不同眼压计测量的眼压与视野进展速率的关联
这项前瞻性观察性队列研究纳入125例青光眼患者的213只眼,平均随访2.4±0.6年,比较Goldmann压平眼压计(GAT)、Ocular Response Analyzer(ORA)的角膜补偿眼压(IOPcc)和ICare回弹式眼压计(RBT)测量的眼压与标准自动视野计平均偏差(MD)进展速率的关系。线性混合模型校正年龄、种族、中央角膜厚度和角膜滞后后显示,平均IOPcc与视野MD丢失的关系最强(R2=24.5%),显著优于GAT(R2=11.1%)和RBT(R2=5.8%)。
要点
- 平均IOPcc预测视野进展速率优于GAT和RBT。
- IOPcc与视野MD丢失的R2为24.5%,GAT为11.1%,RBT为5.8%。
- IOPcc校正角膜影响后可能更利于预测青光眼临床结局。
核心结论
- 平均ORA IOPcc对视野丢失速率的预测优于GAT或RBT测量的平均眼压。
- 校正角膜所致伪影后,IOPcc测量可能对预测青光眼患者临床相关结局有显著优势。
研究设计与样本
- 前瞻性观察性队列研究,纳入125例青光眼患者的213只眼,平均随访2.4±0.6年。
- 每次随访使用GAT、ORA(IOPcc)和RBT测量眼压,并用标准自动视野计MD评估视野丢失速率。
- 采用线性混合模型校正年龄、种族、中央角膜厚度和角膜滞后,分析各眼压计平均眼压与视野丢失速率的关系。
主要发现
- GAT、ORA和RBT的平均眼压分别为14.4±3.3、15.2±4.2和13.4±4.2 mmHg。
- 平均IOPcc与视野MD丢失的关系最强(R2=24.5%)。
- IOPcc模型与GAT模型(R2=11.1%)的R2差异为6.6-19.6(95%CI)。
- IOPcc模型与RBT模型(R2=5.8%)的R2差异为11.1-25.0(95%CI)。
临床怎么用
- 在青光眼随访中,可考虑使用ORA测量的IOPcc辅助评估视野进展风险。
- 解读眼压与视野进展关系时需考虑测量方法差异及角膜特性影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To investigate the associations between intraocular pressure (IOP) measurements obtained by different tonometric methods and rates of visual field loss in a cohort of patients with glaucoma followed over time. DESIGN: Prospective, observational cohort study. PARTICIPANTS: This study included 213 eyes of 125 glaucomatous patients who were followed for an average of 2.4±0.6 years. METHODS: At each visit, IOP measurements were obtained using Goldmann applanation tonometry (GAT), the Ocular Response Analyzer (ORA) (Reichert, Inc., Depew, NY), corneal-compensated IOP (IOPcc), and the ICare Rebound Tonometer (RBT) (Tiolat, Oy, Helsinki, Finland). Rates of visual field loss were assessed by standard automated perimetry (SAP) mean deviation (MD). Linear mixed models were used to investigate the relationship between mean IOP by each tonometer and rates of visual field loss over time, while adjusting for age, race, central corneal thickness, and corneal hysteresis. MAIN OUTCOME MEASURES: Strength of associations (R2) between IOP measurements from each tonometer and rates of SAP MD change over time. RESULTS: Average values for mean IOP over time measured by GAT, ORA, and RBT were 14.4±3.3, 15.2±4.2, and 13.4±4.2 mmHg, respectively. Mean IOPcc had the strongest relationship with SAP MD loss over time (R2 = 24.5%) and was significantly different from the models using mean GAT IOP (R2 = 11.1%; 95% confidence interval [CI] of the difference, 6.6-19.6) and mean RBT IOP (R2= 5.8%; 95% CI of the difference, 11.1-25.0). CONCLUSIONS: Mean ORA IOPcc was more predictive of rates of visual field loss than mean IOP obtained by GAT or RBT. By correcting for corneal-induced artifacts, IOPcc measurements may present significant advantages for predicting clinically relevant outcomes in patients with glaucoma.