未治疗高眼压眼垂直杯盘比测量变异及其对5年青光眼风险估计的影响

Variability of vertical cup to disc ratio measurement and the effects of glaucoma 5-year risk estimation in untreated ocular hypertensive eyes.
PubMed 文献 · Br J Ophthalmol · 2018-06-01

该研究比较了OCT、HRT和视盘立体照相在未治疗高眼压眼中测量的垂直杯盘比(VCDR),并计算对应的5年青光眼发生风险。结果显示不同方法测量的VCDR及对应5年风险存在差异,VCDR最大差异为0.64,5年风险最大差异为24.02%。结论认为VCDR测量不一致会影响5年青光眼风险估计。

要点

  • OCT、HRT和立体照相测量的VCDR分别为0.60±0.14、0.53±0.23、0.44±0.13(SP1)和0.49±0.10(SP2)。
  • 对应5年青光眼风险分别为19.54%±16.60%、18.13%±16.96%、15.64%±14.35%和16.70%±14.49%。
  • VCDR测量最大差异为0.64,对应5年风险最大差异为24.02%。

核心结论

  • OCT、HRT和立体照相在未治疗高眼压眼中测量的VCDR不一致,并延伸至其5年青光眼风险估计的差异。

主要发现

  • OCT、HRT、SP1和SP2测量的VCDR分别为0.60±0.14、0.53±0.23、0.44±0.13和0.49±0.10。
  • 对应5年青光眼发生风险分别为19.54%±16.60%、18.13%±16.96%、15.64%±14.35%和16.70%±14.49%。
  • 立体照相、OCT和HRT获得的VCDR测量中观察到不同程度的比例偏倚。
  • VCDR测量最大差异为0.64,对应5年风险最大差异为24.02%。

关键数值与推荐

  • OCT测量的VCDR为0.60±0.14,对应5年风险19.54%±16.60%。
  • HRT测量的VCDR为0.53±0.23,对应5年风险18.13%±16.96%。
  • SP1立体照相测量的VCDR为0.44±0.13,对应5年风险15.64%±14.35%。
  • SP2立体照相测量的VCDR为0.49±0.10,对应5年风险16.70%±14.49%。

临床怎么用

  • 在未治疗高眼压眼中,OCT、HRT和立体照相测量的VCDR及对应5年青光眼风险估计存在差异。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To compare the vertical cup to disc ratio (VCDR) measurements obtained with optical coherence tomography (OCT), Heidelberg retina tomography (HRT) and stereophotography of the optic nerve head (ONH) in patients with ocular hypertension (OHT), and their corresponding estimated 5-year risk for development of glaucoma. METHODS: ONH images of patients with OHT were taken by fundus camera (stereoscopic images), OCT and HRT. Optic disc stereophotographs were evaluated with a stereo-viewer by two glaucoma specialists (SP1 and SP2) and the VCDR was measured with the ImageJ software. VCDR measurements obtained with stereophotography, OCT and HRT were used to calculate the estimated 5-year risk. RESULTS: One hundred and forty eyes of 75 patients with OHT were included. The VCDR values measured by OCT, HRT, SP1 and SP2 were 0.60±0.14, 0.53±0.23, 0.44±0.13 and 0.49±0.10, respectively. The corresponding 5-year risk for development of glaucoma was 19.54%±16.60%, 18.13%±16.96%, 15.64%±14.35% and 16.70%±14.49%, respectively. Different degrees of proportional biases were observed in VCDR measurements obtained with stereophotography, OCT and HRT. The maximum difference of VCDR measurement was 0.64, while the maximum difference of the corresponding 5-year risk was 24.02%. CONCLUSION: The disagreement in VCDR measured by OCT, HRT and stereophotography in untreated OHT eyes extends to their 5-year risk estimation of glaucoma development.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。