短波长蓝光自发荧光与常规蓝光自发荧光在地图样萎缩中的比较
结果显示450-FAF阅片者间一致性最低、450-RF最高;488-FAF所测GA面积大于450-FAF(中位数2.1 mm² vs 1.0 mm²,p<0.0001),而488-FAF与450-RF面积测量无显著差异(p=1.0)。
要点
- 450-RF来自450-FAF红发射荧光成分分离,其GA定量可重复性最高(ICC 0.997)。
- 450-FAF阅片者间一致性最低(ICC 0.711),488-FAF为0.994。
- 488-FAF所测GA面积大于450-FAF(中位数2.1 mm² vs 1.0 mm²,p<0.0001)。
核心结论
- 从450-FAF图像中分离红发射荧光成分(REFC)可对GA进行可重复量化,且与488-FAF图像所得评估具有良好可比性。
研究设计与样本
- 另获取并分级第三种图像模态:分离并突出红发射荧光成分的450-RF图像;每幅图像由两名阅片者分级以评估阅片者间变异,每种模态取单幅图像评估模态间变异。
主要发现
- 95%重复性系数:488-FAF为1.35 mm²,450-FAF为8.13 mm²,450-RF为1.08 mm²;变异系数:488-FAF为1.11,450-FAF为2.05,450-RF为0.92。
- 组内相关系数(ICC):488-FAF为0.994,450-FAF为0.711,450-RF为0.997,提示450-FAF与450-RF的阅片者间一致性分别为最低与最高。
- GA面积:488-FAF图像中位数(IQR)为2.1 mm²(0.8-6.4 mm²),大于450-FAF图像的1.0 mm²(0.3-4.3 mm²),p<0.0001。
- 488-FAF与450-RF分级之间病灶面积测量无显著差异,中位数(IQR)2.6 mm²(0.8-6.8 mm²),p=1.0。
关键数值与推荐
- 95%重复性系数:488-FAF为1.35 mm²,450-FAF为8.13 mm²,450-RF为1.08 mm²。
- 变异系数:488-FAF为1.11,450-FAF为2.05,450-RF为0.92。
- 组内相关系数(ICC):488-FAF为0.994,450-FAF为0.711,450-RF为0.997。
- GA面积中位数(IQR):488-FAF为2.1 mm²(0.8-6.4 mm²),450-FAF为1.0 mm²(0.3-4.3 mm²),p<0.0001;488-FAF与450-RF比较为2.6 mm²(0.8-6.8 mm²),p=1.0。
临床怎么用
- 从450-FAF图像中分离红发射荧光成分(450-RF)可对GA进行可重复量化,且与488-FAF图像所得评估具有良好可比性。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND/AIMS: To systematically compare the intermodality and inter-reader agreement for two blue-light confocal fundus autofluorescence (FAF) systems. METHODS: Thirty eyes (21 patients) with a diagnosis of geographic atrophy (GA) were enrolled. Eyes were imaged using two confocal blue-light FAF devices: (1) Spectralis device with a 488 nm excitation wavelength (488-FAF); (2) EIDON device with 450 nm excitation wavelength and the capability for 'colour' FAF imaging including both the individual red and green components of the emission spectrum. Furthermore, a third imaging modality (450-RF image) isolating and highlighting the red emission fluorescence component (REFC) was obtained and graded. Each image was graded by two readers to assess inter-reader variability and a single image for each modality was used to assess the intermodality variability. RESULTS: The 95% coefficient of repeatability (1.35 mm2 for the 488-FAF-based grading, 8.13 mm2 for the 450-FAF-based grading and 1.08 mm2 for the 450-RF-based grading), the coefficient of variation (1.11 for 488-FAF, 2.05 for 450-FAF, 0.92 for 450-RF) and the intraclass correlation coefficient (0.994 for 488-FAF, 0.711 for 450-FAF, 0.997 for 450-RF) indicated that 450-FAF-based and 450-RF-based grading have the lowest and highest inter-reader agreements, respectively. The GA area was larger for 488-FAF images (median (IQR) 2.1 mm2 (0.8-6.4 mm2)) than for 450-FAF images (median (IQR) 1.0 mm2 (0.3-4.3 mm2); p<0.0001). There was no significant difference in lesion area measurement between 488-FAF-based and 450-RF-based grading (median (IQR) 2.6 mm2 (0.8-6.8 mm2); p=1.0). CONCLUSION: The isolation of the REFC from the 450-FAF images allowed for a reproducible quantification of GA. This assessment had good comparability with that obtained with 488-FAF images.