文献速递角膜与眼表疾病

比较角膜交联术后前表面曲率变化:Scheimpflug与扫描裂隙技术

Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology.
PubMed 文献 · Am J Ophthalmol · 2018-05-03

本研究基于随机临床试验数据,评估Scheimpflug断层扫描与Placido扫描裂隙地形图在进展性圆锥角膜行标准CXL后前轴向曲率差异图的相关性。51眼在术前及术后3、6、12月成像,两设备绝对Kmax显著不同但强相关,ΔKmax和RI无显著差异但相关性差,95%一致性界限宽。结论为两设备相关但不可互换,分别监测CXL反应合理,比较结果需谨慎。

要点

  • 两设备绝对Kmax各时间点显著不同但强相关(r=0.90-0.93)。
  • ΔKmax和RI两设备无显著差异但相关性差(r=0.41-0.53和0.29-0.48)。
  • 12月95%一致性界限:绝对Kmax 7.51D,ΔKmax 8.61D,RI 19.86D。

核心结论

  • Scheimpflug与扫描裂隙Placido技术测量相关但不可互换,分别监测CXL反应合理,跨技术比较结果需谨慎。

研究设计与样本

  • 基于随机临床试验数据的设备间可靠性分析,单中心收集术前及术后3、6、12月角膜成像,标准CXL方案为3 mW/cm²持续30分钟。
  • 共纳入36例患者51眼,所有时间点均完成评估。

主要发现

  • 绝对Kmax术前Scheimpflug 56.01±5.3 D vs 扫描裂隙55.04±5.1 D(P=.003),12月54.58±5.3 D vs 53.12±4.9 D(P<.0001),各时间点均显著不同但强相关(r=0.90-0.93)。
  • ΔKmax和RI两设备各时间点均无显著差异,但相关性差(ΔKmax r=0.41-0.53,RI r=0.29-0.48)。
  • 12月95%一致性界限:绝对Kmax 7.51 D,ΔKmax 8.61 D,RI 19.86 D。

关键数值与推荐

  • 标准CXL方案:3 mW/cm²持续30分钟。
  • 各时间点绝对Kmax设备间比较P值:术前P=.003,12月P<.0001。
  • 设备间相关系数:绝对Kmax r=0.90-0.93;ΔKmax r=0.41-0.53;RI r=0.29-0.48。
  • 12月95%一致性界限:绝对Kmax 7.51 D,ΔKmax 8.61 D,RI 19.86 D。

临床怎么用

  • 两设备均可分别用于监测角膜对CXL的反应。
  • 比较不同测量技术获得的结果时应谨慎。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate the correlation between anterior axial curvature difference maps following corneal cross-linking (CXL) for progressive keratoconus obtained from Scheimpflug-based tomography and Placido-based topography. DESIGN: Between-devices reliability analysis of randomized clinical trial data. METHODS: Corneal imaging was collected at a single-center institution preoperatively and at 3, 6, and 12 months postoperatively using Scheimpflug-based tomography (Pentacam; Oculus Inc, Lynnwood, Washington, USA) and scanning-slit, Placido-based topography (Orbscan II; Bausch & Lomb, Rochester, New York, USA) in patients with progressive keratoconus receiving standard protocol CXL (3 mW/cm2 for 30 minutes). Regularization index (RI), absolute maximum keratometry (K Max), and change in K Max (ΔK Max) were compared between the 2 devices at each time point. RESULTS: Fifty-one eyes from 36 patients were evaluated at all time points. Values were significantly different at all time points (56.01 ± 5.3 diopters [D] Scheimpflug vs 55.04 ± 5.1 D scanning-slit preoperatively [P = .003]; 54.58 ± 5.3 D Scheimpflug vs 53.12 ± 4.9 D scanning-slit at 12 months [P < .0001]) but strongly correlated between devices (r = 0.90-0.93) at all time points. The devices were not significantly different at any time point for either ΔK Max or RI but were poorly correlated at all time points (r = 0.41-0.53 for ΔK Max, r = 0.29-0.48 for RI). At 12 months, 95% limits of agreement were 7.51 D for absolute K Max, 8.61 D for ΔK Max, and 19.86 D for RI. CONCLUSIONS: Measurements using Scheimpflug and scanning-slit Placido-based technology are correlated but not interchangeable. Both devices appear reasonable for separately monitoring the cornea's response to CXL; however, caution should be used when comparing results obtained with one measuring technology to the other.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Lang PZ, Thulasi P, Khandelwal SS, et al. Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology. Am J Ophthalmol. 2018. PMID: 29729256. doi:10.1016/j.ajo.2018.04.018

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。