基于游戏的儿童视野检查:诊断效能与重复性研究
在曼彻斯特皇家眼科医院儿科门诊开展效度与信度研究,108名无眼病儿童(4-12岁)用Caspar's Castle系统检查单眼,其中45名用正常阈值、43名用人工缺陷阈值;另21名眼病儿童(4-16岁)与标准视野检查定性比较;106次检查可重复两次。人工缺陷检测的ROC曲线下面积为0.895,全部队列(106例)漏点数重复性系数为6.9(95%CI 6.16-8.07)。作者认为该系统展示出令人鼓舞的敏感度、特异度和信度,可能有助于解决幼儿视野检查的困难。
要点
- 人工缺陷检测的ROC曲线下面积为0.895。
- 21名眼病儿童中7名完成标准Humphreys视野检查,Caspar's Castle视野经专家判断与之一致。
- 全部106例儿童漏点数重复性系数为6.9(95%CI 6.16-8.07)。
核心结论
- Caspar's Castle基于游戏策略评估视野,展示出令人鼓舞的敏感度、特异度和信度。
- 该系统可能有助于解决当前幼儿视野检查的困难。
研究设计与样本
- 在曼彻斯特皇家眼科医院儿科眼科门诊开展效度与信度研究。
- 招募108名无眼病儿童(4-12岁),用Caspar's Castle系统检查单眼,采用正常阈值或人工改造以重现缺陷的阈值评估诊断效能。
- 另招募21名眼病儿童(4-16岁),将Caspar's Castle视野与已有视野检查进行定性比较。
- 共106次Caspar's Castle检查能够完成两次,通过重复性系数和Bland-Altman图确定重复性。
主要发现
- 无眼病儿童诊断测试中,45名完成正常阈值测试,43名完成人工缺陷阈值测试。
- 人工缺陷检测的ROC曲线下面积为0.895。
- 21名眼病儿童中7名完成标准Humphreys视野检查,Caspar's Castle视野经专家意见与这7名均对应一致。
- 全部儿童队列(106例)漏点数的重复性系数为6.9(95%CI 6.16-8.07)。
关键数值与推荐
- 人工缺陷检测的ROC曲线下面积为0.895。
- 全部队列(106例)漏点数重复性系数为6.9(95%CI 6.16-8.07)。
- 21名眼病儿童中7名完成标准Humphreys视野检查,Caspar's Castle视野与之一致。
临床怎么用
- 作者认为Caspar's Castle系统展示出令人鼓舞的敏感度、特异度和信度,可能有助于解决当前幼儿视野检查的困难。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To demonstrate utility of a game-based test ("Caspar's Castle") for the detection of visual field defects in children. METHODS: A validity and reliability study was carried out at Manchester Royal Eye Hospital Pediatric Ophthalmology Outpatients Department. We recruited 108 children with no eye pathology (aged 4-12 years) and examined a single eye with the Caspar's Castle system using either normal thresholds or thresholds artificially adapted to recreate defects to assess diagnostic utility. Number of peripheral stimuli missed was used to determine sensitivity and specificity of artificial defect detection and to plot receiver-operator characteristic curves. A further 21 children (aged 4-16 years) with pathology were recruited and Caspar's fields compared qualitatively with established field testing. A total of 106 of the Caspar's Castle examinations were able to be performed twice and repeatability was determined through coefficient of repeatability and Bland-Altman chart. RESULTS: In diagnostic testing using children with no eye pathology, 45 children completed a test using normal thresholds and 43 with tests using artificial defects. Area under receiver-operator characteristic curves for artificial defect detection was 0.895. Of the 21 children with pathology, seven had completed standard Humphreys field testing and Caspar's Castle fields corresponded with each of these by expert opinion. Coefficient of repeatability for number of points missed across all cohorts of children (106 patients) was 6.9 (95% confidence interval: 6.16-8.07). CONCLUSIONS: The Caspar's Castle system of assessing visual fields using novel game-based strategies demonstrates encouraging levels of sensitivity, specificity, and reliability. It could help address current difficulties in perimetry in young children.
引用
- Aslam TM, Ali ZC, Wang Y, et al. Diagnostic Performance and Repeatability of a Novel Game-Based Visual Field Test for Children. Invest Ophthalmol Vis Sci. 2018. PMID: 29625475. doi:10.1167/iovs.17-23546
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。