角膜新生血管成像:光学相干断层扫描血管成像与荧光血管造影比较
本研究纳入15例角膜新生血管患者,比较OCTA与ICGA对血管参数的评估。ICGA主观图像质量评分优于OCTA(3.3±0.9 vs 2.1±1.2,P=0.002)。两种设备对血管参数的一致性为轻度至中度,OCTA在捕捉小血管方面精确性较低。
要点
- ICGA主观图像质量评分优于OCTA(3.3±0.9 vs 2.1±1.2,P=0.002)。
- ICGA与OCTA对血管参数的一致性为轻度至中度。
- OCTA可提供红细胞流动证据和深度信息,辅助ICGA用于治疗规划。
核心结论
- 与ICGA相比,当前OCTA系统在捕捉角膜新生血管复合体中的小血管方面精确性较低,OCTA分割软件需要验证研究。
- OCTA通过提供红细胞流动证据及深度信息,可与ICGA互补,可能有助于角膜新生血管的治疗规划。
研究设计与样本
- 纳入角膜新生血管延伸至少3 mm进入角膜的患者,所有患者在同一次就诊时接受角膜成像。
- 使用AngioVue OCTA系统(Optovue,Inc.)配合长角膜适配模块(CAM-L)记录图像;使用Heidelberg系统(HRA2扫描激光检眼镜;Heidelberg Engineering)配合荧光滤光片记录ICGA图像。
- 图像由两名独立观察者进行质量分级,比较设备间的血管参数(面积、数量、直径、分支和终点、迂曲度),使用Bland-Altman图评估参数差异。
主要发现
- 共纳入15例角膜新生血管患者,主要与微生物性角膜炎相关。
- ICGA平均主观图像质量评分为3.3±0.9,优于OCTA的2.1±1.2(P=0.002)。
- ICGA图像的观察者间一致性几乎完美(κ=0.83),OCTA图像为实质性一致(κ=0.69)。
- ICGA与OCTA对所有研究血管参数分级的一致性为轻度至中度。
- 血管直径差异为-8.98 μm(P=0.01,95%一致性界限[LOA]:-15.89至-2.07),分支数量差异为25.93(P=0.09,95% LOA:-4.31至56.17),终点数量差异为49(P=0.05,95% LOA:0.78至97.22)。
关键数值与推荐
- 纳入标准:角膜新生血管延伸至少3 mm进入角膜。
- ICGA平均主观图像质量评分3.3±0.9 vs OCTA 2.1±1.2(P=0.002)。
- 观察者间一致性:ICGA κ=0.83,OCTA κ=0.69。
- 血管直径差异-8.98 μm(P=0.01,95% LOA:-15.89至-2.07)。
- 分支数量差异25.93(P=0.09,95% LOA:-4.31至56.17);终点数量差异49(P=0.05,95% LOA:0.78至97.22)。
临床怎么用
- OCTA可提供红细胞流动证据及深度信息,与ICGA互补,可能有助于角膜新生血管的治疗规划。
- 作者认为当前OCTA系统在捕捉小血管方面精确性较低,OCTA分割软件需要验证研究。
局限性
- OCTA分割软件需要验证研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The purpose of this study was to compare optical coherence tomography angiography (OCTA) and indocyanine green angiography (ICGA) for the assessment of corneal neovascularization (CoNV). METHODS: Patients with CoNV extending at least 3 mm into the cornea were included. All patients underwent corneal imaging at the same visit. Images were recorded using the AngioVue OCTA system (Optovue, Inc.) with the long corneal adaptor module (CAM-L). ICGA images were recorded with fluorescent filters using the Heidelberg system (HRA2 Scanning Laser Ophthalmoscope; Heidelberg Engineering). Images were graded for quality by two independent observers. Vessel parameters: area, number, diameter, branch and end points, and tortuosity, were compared between devices. Bland-Altman plots were used to assess differences between parameters. RESULTS: Fifteen patients with CoNV predominantly associated with microbial keratitis were included. Mean subjective image quality score was better for ICGA (3.3 ± 0.9) than for OCTA (2.1 ± 1.2, P = 0.002), with almost perfect interobserver agreement for ICGA images (κ = 0.83) and substantial agreement for OCTA images (κ = 0.69). Agreement of grading of all investigated vessel parameters between ICGA and OCT images was slight to moderate, with significant differences found for vessel diameter (-8.98 μm, P = 0.01, 95% limits of agreement [LOA]: -15.89 to -2.07), number of branch (25.93, P = 0.09, 95% LOA: -4.31 to 56.17), and terminal points (49, P = 0.05, 95% LOA: 0.78 to 97.22). CONCLUSION: Compared with ICGA, current OCTA systems are less precise in capturing small vessels in CoNV complexes, and validation studies are needed for OCTA segmentation software. OCTA, however, complements ICGA by providing evidence of red blood cell flow, which together with depth information, may be helpful when planning treatment of CoNV.