短周期灭菌用于眼科手术器械连续同日白内障手术的评估
本研究由独立医疗器械验证测试实验室对3家制造商的超声乳化手柄及生物指示剂进行短周期灭菌测试,采用两种常见蒸汽灭菌器。结果显示所有接种样本均无目标微生物生长,支持在遵循制造商说明书的前提下,对连续同日白内障手术使用未包装短周期灭菌。
要点
- 短周期灭菌(暴露3.0-3.5分钟)后所有生物指示剂样本均无目标微生物生长。
- 干燥阶段中断不影响灭菌效果,处理器械残留水分无菌。
- 遵循说明书时,未包装短周期灭菌可用于连续同日白内障手术。
核心结论
- 研究支持在遵循两种经FDA批准的灭菌器说明书的前提下,对连续同日白内障手术使用未包装短周期灭菌。
- 当器械保存在覆盖的灭菌容器内以备连续手术使用时,不需要完整干燥阶段。
研究设计与样本
- 由独立医疗器械验证测试实验室对特定灭菌处理方法进行测试。
- 测试3家不同制造商的超声乳化手柄,以及适当的生物指示剂和对照,使用2种常见蒸汽灭菌器。
主要发现
- 两种灭菌器的所有接种测试样本均无目标微生物生长,无论完整干燥阶段是否中断。
- 处理器械后残留水分的移液样本和手柄拭子经3分钟转运/储存时间后也无微生物生长。
关键数值与推荐
- 短周期暴露时间为3.0-3.5分钟,用于未包装和容器装载,干燥时间仅1分钟以模拟连续病例使用。
- 生物指示剂接种高抗力嗜热脂肪地芽孢杆菌,每次灭菌验证测试一式三份进行。
- 测试包括包装和容器装载,采用完整干燥周期和7天储存时间以模拟延长储存。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The common practice of short-cycle sterilization for ophthalmic surgical instrumentation has come under increased regulatory scrutiny. This study was undertaken to evaluate the efficacy of short-cycle sterilization processing for consecutive same-day cataract procedures. DESIGN: Testing of specific sterilization processing methods by an independent medical device validation testing laboratory. PARTICIPANTS: Phaco handpieces from 3 separate manufacturers were tested along with appropriate biologic indicators and controls using 2 common steam sterilizers. METHODS: A STATIM 2000 sterilizer (SciCan, Canonsburg, PA) with the STATIM metal cassette, and an AMSCO Century V116 pre-vacuum sterilizer (STERIS, Mentor, OH) using a Case Medical SteriTite container (Case Medical, South Hackensack, NJ) rigid container were tested using phaco tips and handpieces from 3 different manufacturers. Biological indicators were inoculated with highly resistant Geobacillus stearothermophilus, and each sterility verification test was performed in triplicate. Both wrapped and contained loads were tested with full dry cycles and a 7-day storage time to simulate prolonged storage. In adherence with the manufacturers' instructions for use (IFU), short cycles (3.0-3.5-minute exposure times) for unwrapped and contained loads were also tested after only 1 minute of dry time to simulate use on a consecutive case. Additional studies were performed to demonstrate whether any moisture present in the load containing phaco handpieces postprocessing was sterile and would affect the sterility of the contents after a 3-minute transit/storage time. This approximated the upper limit of time needed to transfer a containment device to the operating room. MAIN OUTCOME MEASURES: Presence or absence of microbial growth from cultured test samples. RESULTS: All inoculated test samples from both sterilizers were negative for growth of the target organism whether the full dry phase was interrupted or not. Pipetted postprocessing moisture samples and swabs of the handpieces were also negative for growth after a 3-minute transit/storage time. CONCLUSIONS: These studies support the use of unwrapped, short-cycle sterilization that adheres to the IFU of these 2 popular Food and Drug Administration-cleared sterilizers for sequential same-day cataract surgeries. A full drying phase is not necessary when the instruments are kept within the covered sterilizer containment device for prompt use on a sequential case.
引用
- Chang DF, Hurley N, Mamalis N, et al. Evaluation of Ophthalmic Surgical Instrument Sterility Using Short-Cycle Sterilization for Sequential Same-Day Use. Ophthalmology. 2018. PMID: 29602568. doi:10.1016/j.ophtha.2018.03.012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。