便携式平板视野计与Humphrey视野分析仪的六个月纵向比较
多中心纵向观察性临床研究纳入60例稳定青光眼、高眼压或可疑青光眼患者,在基线和第2、4、6个月比较iPad视野应用Melbourne Rapid Fields(MRF)与Humphrey视野分析仪(HFA)24-2 SITA-standard及SITA-fast程序。MRF与HFA在4次访视中平均偏差(MD)呈强相关,重测信度良好;MRF速度与SITA-fast相似且显著快于SITA-standard。
要点
- MRF与SITA-fast的MD组内相关系数(ICC)在4次访视中为0.71至0.88,与SITA-standard为0.81至0.90。
- MRF的MD重测信度优秀,基线与6个月访视ICC为0.98(95%CI 0.96-0.99)。
- MRF测试时间4.6±0.1分钟,与SITA-fast的4.3±0.2分钟相似,显著快于SITA-standard的6.2±0.1分钟(P<.001)。
核心结论
- MRF在6个月内4次访视中与HFA强相关,具有良好重测信度,适用于常规视野检查不易获取的环境下监测视野。
研究设计与样本
- 多中心纵向观察性临床研究,纳入60例稳定青光眼、高眼压或可疑青光眼患者,计划在基线和第2、4、6个月访视。
主要发现
- MRF与SITA-fast的MD ICC在4次访视中为0.71至0.88,MRF与SITA-standard的MD ICC为0.81至0.90。
- MRF的MD重测信度优秀,基线与6个月访视ICC为0.98(95%CI 0.96-0.99);SITA-fast 6个月重测ICC为0.95,SITA-standard为0.93。
- MRF测试时间4.6±0.1分钟,SITA-fast 4.3±0.2分钟,SITA-standard 6.2±0.1分钟,MRF与SITA-fast速度相似且显著快于SITA-standard(P<.001)。
- MRF变化的点位数更少,但在发生变化的点上MRF的点位变异性大于SITA测试。
临床怎么用
- MRF适用于常规视野检查不易获取的环境下监测视野。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To establish the medium-term repeatability of the iPad perimetry app Melbourne Rapid Fields (MRF) compared to Humphrey Field Analyzer (HFA) 24-2 SITA-standard and SITA-fast programs. DESIGN: Multicenter longitudinal observational clinical study. METHODS: Sixty patients (stable glaucoma/ocular hypertension/glaucoma suspects) were recruited into a 6-month longitudinal clinical study with visits planned at baseline and at 2, 4, and 6 months. At each visit patients undertook visual field assessment using the MRF perimetry application and either HFA SITA-fast (n = 21) or SITA-standard (n = 39). The primary outcome measure was the association and repeatability of mean deviation (MD) for the MRF and HFA tests. Secondary measures were the point-wise threshold and repeatability for each test, as well as test time. RESULTS: MRF was similar to SITA-fast in speed and significantly faster than SITA-standard (MRF 4.6 ± 0.1 minutes vs SITA-fast 4.3 ± 0.2 minutes vs SITA-standard 6.2 ± 0.1 minutes, P < .001). Intraclass correlation coefficients (ICC) between MRF and SITA-fast for MD at the 4 visits ranged from 0.71 to 0.88. ICC values between MRF and SITA-standard for MD ranged from 0.81 to 0.90. Repeatability of MRF MD outcomes was excellent, with ICC for baseline and the 6-month visit being 0.98 (95% confidence interval: 0.96-0.99). In comparison, ICC at 6-month retest for SITA-fast was 0.95 and SITA-standard 0.93. Fewer points changed with the MRF, although for those that did, the MRF gave greater point-wise variability than did the SITA tests. CONCLUSIONS: MRF correlated strongly with HFA across 4 visits over a 6-month period, and has good test-retest reliability. MRF is suitable for monitoring visual fields in settings where conventional perimetry is not readily accessible.
引用
- Prea SM, Kong YXG, Mehta A, et al. Six-month Longitudinal Comparison of a Portable Tablet Perimeter With the Humphrey Field Analyzer. Am J Ophthalmol. 2018. PMID: 29550190. doi:10.1016/j.ajo.2018.03.009