文献速递青光眼

手动眼睑牵拉对回弹式眼压计测量眼压的影响

Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry.
PubMed 文献 · Br J Ophthalmol · 2018-02-02

研究纳入103例原发性开角型青光眼患者,前瞻性比较回弹式眼压计(RT)在有无眼睑牵拉(LM)条件下与Goldmann压平眼压计(GAT)的测量值。RT伴LM测得的眼压显著高于GAT,而RT不伴LM与GAT无显著差异;睑裂高度(PFH)越小,RT伴LM与GAT的差值越夸大。

要点

  • RT伴眼睑牵拉测得的眼压显著高于GAT(15.21±2.91 vs 13.97±2.80 mm Hg,P<0.001)。
  • RT不伴眼睑牵拉与GAT测得的眼压无显著差异(13.75±2.44 vs 13.97±2.80 mm Hg,P=0.096)。
  • 睑裂高度越小,RT伴眼睑牵拉与GAT的眼压差值越夸大(P<0.001)。

核心结论

  • 对眼睑施加牵拉时,RT测得的眼压显著偏高,且睑裂高度越小该趋势越明显;GAT与不伴眼睑牵拉的RT测值相关性良好。

研究设计与样本

  • 前瞻性研究,纳入103例原发性开角型青光眼患者,随机顺序进行RT伴眼睑牵拉、RT不伴眼睑牵拉和GAT三种测量,并测量睑裂高度。

主要发现

  • GAT测得的眼压均值为13.97±2.80 mm Hg,与RT不伴眼睑牵拉(13.75±2.44 mm Hg)无显著差异(P=0.096),但显著低于RT伴眼睑牵拉(15.21±2.91 mm Hg;P<0.001)。
  • Bland-Altman图中,RT伴眼睑牵拉相对GAT平均高估(均值-1.5),相对RT不伴眼睑牵拉平均高估(均值-1.2)。
  • 在高眼压(>20 mm Hg)亚组中,RT不伴眼睑牵拉测得的眼压显著低于GAT(P<0.001)。
  • 睑裂高度亚组分析显示,睑裂高度越小,GAT(P=0.014)与RT伴眼睑牵拉(P<0.001)之间的眼压差值越夸大。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND/AIMS: To investigate the effect of eyelid manipulation on the measurement of intraocular pressure (IOP) using two different tonometries (rebound tonometry (RT) vs Goldmann applanation tonometry (GAT)). METHODS: 103 patients with primary open-angle glaucoma were prospectively enrolled. For all of the patients, IOP measurements were performed in three different ways: (1) RT with lid manipulation (LM), (2) RT without LM and (3) GAT. The order of the three measurements was randomly selected. Additionally, the palpebral fissure height (PFH; elliptical space between upper and lower eyelids) was measured. RESULTS: The mean value of IOP measured by GAT was 13.97±2.80 mm Hg, which was not significantly different from that by RT without LM (13.75±2.44 mm Hg; P=0.096), but which was significantly lower than that by RT with LM (15.21±2.91 mm Hg; P<0.001). On a Bland-Altman plot, RT with LM was overestimated relative to GAT (mean: -1.5) and RT without LM (mean: -1.2). Among the high IOPs (>20 mm Hg), interestingly, those measured by RT without LM were significantly lower than those measured by GAT (P<0.001). In the subgroup analysis of PFH, the smaller the PFH, the more exaggerated the IOP difference between GAT (P=0.014) and RT with LM (P<0.001). CONCLUSION: RT-measured IOP was significantly exaggerated when manipulation was applied to the eyelid. This overall trend was more pronounced when PFH was small. GAT-measured IOP, meanwhile, showed a good correlation with IOP measured using RT without LM.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Baek SU, Ha A, Kim YK, et al. Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry. Br J Ophthalmol. 2018. PMID: 29420194. doi:10.1136/bjophthalmol-2017-311587

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。