文献速递白内障与屈光手术

近视眼中多焦点人工晶状体的下方偏心

Inferior Decentration of Multifocal Intraocular Lenses in Myopic Eyes.
PubMed 文献 · Am J Ophthalmol · 2018-01-31

前瞻性队列研究评估28只近视眼与56只正视眼植入Tecnis ZMB00多焦点人工晶状体后1年的倾斜与偏心,发现近视组晶状体囊袋内下方偏心显著大于对照组,整体偏心更大,高阶像差更差、调制传递函数更差、主观症状更多。

要点

  • 近视组多焦点人工晶状体下方偏心大于对照组(-0.03±0.22 mm vs -0.21±0.29 mm,P=.002)。
  • 近视组整体偏心大于对照组(0.32±0.14 mm vs 0.40±0.18 mm,P=.023)。
  • 眼轴长度与垂直偏心负相关(r=-0.268,P=0.014),与整体偏心正相关(r=0.334,P=.002)。

核心结论

  • 近视眼植入多焦点人工晶状体后下方偏心更大并导致视觉质量下降,眼轴延长时人工晶状体与囊袋尺寸不匹配加剧不应被低估。

研究设计与样本

  • 前瞻性队列研究纳入28只近视眼和56只正视眼,行超声乳化联合Tecnis ZMB00多焦点人工晶状体植入,术后1年评估倾斜、偏心、高阶像差、调制传递函数及主观症状。

主要发现

  • 术后裸眼远视力、最佳矫正远视力和裸眼近视力两组间无差异。
  • 人工晶状体平均倾斜和水平偏心两组间无差异。
  • 近视组晶状体囊袋内下方偏心显著大于对照组(-0.03±0.22 mm vs -0.21±0.29 mm,P=.002)。
  • 整体偏心值对照组为0.32±0.14 mm,近视组为0.40±0.18 mm(P=.023)。
  • 眼轴长度与垂直偏心负相关(r=-0.268,P=0.014),与整体偏心正相关(r=0.334,P=.002)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To investigate the tilt and decentration of multifocal intraocular lenses (MfIOLs) implanted in myopic eyes. DESIGN: A prospective cohort study. METHODS: Twenty-eight myopic eyes and 56 emmetropic eyes were evaluated. Phacoemulsification with Tecnis ZMB00 MfIOL (Abbott Medical Optics, Santa Ana, California, USA) implantation was performed. At 1 year post-surgery, routine postoperative examinations were performed, and tilt and decentration of the MfIOLs, high-order aberrations, and modulation transfer function (MTF) were evaluated using the OPD-Scan III aberrometer (Nidek Co, Ltd, Gamagori, Japan). Subjective symptoms were assessed with a Quality of Vision questionnaire. RESULTS: Postoperative uncorrected distance visual acuity (VA), best-corrected distance VA, and uncorrected near VA did not differ between the 2 groups. The mean IOL tilt and horizontal decentration were not different between the control and myopic groups. However, the myopic group presented significantly inferior decentration in the capsular bag compared with the control group (-0.03 ± 0.22 mm vs -0.21 ± 0.29 mm, P = .002). The overall decentration values were 0.32 ± 0.14 mm in the controls and 0.40 ± 0.18 mm in the myopic group (P = .023). Axial length was negatively correlated with vertical decentration (r = -0.268, P = 0.014) and positively correlated with overall decentration (r = 0.334, P = .002). Worse aberration data, poorer MTF, and more subjective symptoms were also found in the myopic group than in the controls. CONCLUSION: Greater inferior decentration of MfIOLs and a consequent decrease in visual quality were found in myopic eyes, indicating that the increasing incompatibility between IOL and capsular bag size with axial length elongation should not be underestimated.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Zhu X, He W, Zhang Y, et al. Inferior Decentration of Multifocal Intraocular Lenses in Myopic Eyes. Am J Ophthalmol. 2018. PMID: 29355482. doi:10.1016/j.ajo.2018.01.007

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。