3 mm×3 mm与6 mm×6 mm光相干断层扫描血管成像扫描范围在非增殖型糖尿病视网膜病变评估中的比较
本研究纳入27例患者共46只眼,用RTVue XR Avanti OCT的3 mm×3 mm与6 mm×6 mm「Angio Retina」扫描方案成像,比较两种扫描范围对中心凹无血管区(FAZ)、FAZ重塑、微动脉瘤、毛细血管无灌注及伪影的显示能力。3 mm×3 mm显示FAZ与FAZ重塑优于6 mm×6 mm,6 mm×6 mm检出微动脉瘤更易,毛细血管无灌注与伪影发生率两种扫描范围无显著差异。作者认为两种扫描范围在评估此类患者时均有应用价值。
要点
- 3 mm×3 mm显示FAZ 89% vs 6 mm×6 mm 59%(P=0.001),前者更优。
- 6 mm×6 mm检出微动脉瘤57% vs 3 mm×3 mm 35%(P=0.003),前者更易。
- 毛细血管无灌注与运动、重影伪影发生率两种扫描范围无显著差异。
核心结论
- 3 mm×3 mm扫描显示FAZ及FAZ重塑优于6 mm×6 mm,可能因其扫描密度更高;6 mm×6 mm检出微动脉瘤更易,可能因其扫描面积更大;作者认为两种扫描在评估这些患者时均有应用价值。
研究设计与样本
- 采用RTVue XR Avanti OCT的3 mm×3 mm与6 mm×6 mm「Angio Retina」扫描方案对患者进行成像,评估清晰勾勒FAZ、FAZ重塑、微动脉瘤、毛细血管无灌注、运动伪影与重影伪影的能力。
- 共纳入27例患者的46只眼。
主要发现
- 3 mm×3 mm扫描89% vs 6 mm×6 mm扫描59%清晰勾勒FAZ(P=0.001)。
- 3 mm×3 mm扫描80% vs 6 mm×6 mm扫描43%显示FAZ重塑(P=0.0002)。
- 6 mm×6 mm扫描57% vs 3 mm×3 mm扫描35%检出微动脉瘤(P=0.003)。
- 3 mm×3 mm扫描87% vs 6 mm×6 mm扫描89%检出毛细血管无灌注(P=0.99)。
- 两种扫描范围间伪影发生率无显著差异:运动伪影P=0.29,重影伪影P=0.29。
关键数值与推荐
- 清晰勾勒FAZ:3 mm×3 mm扫描89% vs 6 mm×6 mm扫描59%(P=0.001)。
- 显示FAZ重塑:3 mm×3 mm扫描80% vs 6 mm×6 mm扫描43%(P=0.0002)。
- 检出微动脉瘤:6 mm×6 mm扫描57% vs 3 mm×3 mm扫描35%(P=0.003)。
- 检出毛细血管无灌注:3 mm×3 mm扫描87% vs 6 mm×6 mm扫描89%(P=0.99)。
- 伪影发生率:运动伪影P=0.29,重影伪影P=0.29,两种扫描范围无显著差异。
临床怎么用
- 作者认为,在评估非增殖型糖尿病视网膜病变患者时,3 mm×3 mm与6 mm×6 mm扫描均有应用价值。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To evaluate the utility of different optical coherence tomography angiography scan protocols in evaluating retinal changes in non-proliferative diabetic retinopathy. METHODS: Patients were imaged with the RTVue XR Avanti OCT 3 mm × 3 mm and 6 mm × 6 mm "Angio Retina" scan protocols. Ability to clearly delineate the foveal avascular zone (FAZ), FAZ remodeling, microaneurysms, capillary nonperfusion, motion, and doubling artifacts were evaluated. RESULTS: Forty-six eyes from 27 patients were enrolled. Eighty-nine percent of 3 mm × 3 mm versus 59% of 6 mm × 6 mm scans clearly delineated the FAZ (P = 0.001). Eighty percent of 3 mm × 3 mm versus 43% of 6 mm × 6 mm scans demonstrated FAZ remodeling (P = 0.0002). Microaneurysms were detected by 57% of 6 mm × 6 mm and 35% of 3 mm × 3 mm scans (P = 0.003). Capillary nonperfusion was detected in 87% of 3 mm × 3 mm scans versus 89% of 6 mm × 6 mm scans (P = 0.99). No significant differences were noted in the incidence of artifacts between the scan sizes (motion artifact P = 0.29 and doubling artifact P = 0.29). CONCLUSION: 3 mm × 3 mm scan delineated FAZ and remodeling better than 6 mm × 6 mm scan, likely because of its higher scan density. 6 mm × 6 mm scans detected microaneurysms more readily than 3 mm × 3 mm, likely because of its larger scan area. There were utility for both 3 mm × 3 mm and 6 mm × 6 mm scans when evaluating these patients.