热装置(精准脉冲囊切开术)创建人囊切开术的超微结构完整性

Ultrastructural Integrity of Human Capsulotomies Created by a Thermal Device.
PubMed 文献 · Ophthalmology · 2017-10-23

这项前瞻性多中心病例系列研究收集了2名术者用精准脉冲囊切开术(PPC)装置行囊切开患者的标本,行扫描电镜观察囊膜边缘。结果显示囊膜边缘总体均匀并外翻,但部分区域边缘不规则、呈毛边状,裂隙灯下也可见毛边外观,作者认为可能与热能耗散有关,其完整性与临床相关性需进一步评估。

要点

  • PPC装置能创建可重复、居中且精确的圆形囊切开。
  • 电镜下囊切开边缘总体均匀并外翻,部分区域边缘不规则、毛边状。
  • 术后裂隙灯可见前囊膜边缘毛边外观,其完整性与临床相关性需进一步评估。

核心结论

  • PPC装置能够创建可重复、居中且精确的圆形囊切开。
  • 离体人囊切开标本的超微结构总体显示囊切开边缘外翻,但部分病例伴有囊膜边缘不规则、毛边区域,可能由热能耗散所致。
  • 术后外观与扫描电镜特征提示需进一步评估PPC的完整性与临床相关性。

研究设计与样本

  • 前瞻性、多中心病例系列。
  • 连续纳入接受PPC装置囊切开的患者,由2名术者完成,随后行常规超声乳化白内障手术。
  • 所有囊切开标本均收集用于扫描电镜(SEM)检查。

主要发现

  • 术后裂隙灯检查可见前囊膜边缘毛边外观。
  • 扫描电镜显示囊膜边缘总体均匀并卷曲,但囊膜边缘间杂有不规则、毛边外观的区域。
  • 研究将上述表现与手工连续环形撕囊进行了比较。

临床怎么用

  • 作者认为,术后外观与扫描电镜特征提示需进一步评估PPC的完整性与临床相关性。

局限性

  • 术后外观与扫描电镜特征提示需进一步评估PPC的完整性与临床相关性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To investigate the ultrastructural features of anterior capsulotomy performed with a thermal device, the precision pulse capsulotomy (PPC). DESIGN: Prospective, multicenter case series. PARTICIPANTS: Consecutive patients undergoing capsulotomy with the PPC device. METHODS: Prospective study of patients undergoing capsulotomy with the PPC by 2 surgeons, followed up by routine phacoemulsification cataract surgery, was undertaken. All capsulotomy specimens were collected for scanning electron microscopy (SEM). Observations were made regarding uniformity of the capsular edge and the presence of irregularities that may compromise integrity. Comparisons were made with manual continuous curvilinear capsulorrhexis. MAIN OUTCOME MEASURES: Ultrastructural features of PPC and presence of irregularities. RESULTS: Frayed appearance of the anterior capsule edge was noted in postoperative visits under slit-lamp examination. Scanning electron microscopy sampling showed a generally uniform rolled capsular edge, but interspersed with areas of irregularity with frayed appearance at the capsule margin. CONCLUSIONS: The PPC device is capable of creating reproducible, central, and precise circular capsulotomy. The ultrastructural features in ex vivo human capsulotomy specimens generally show eversion of the capsulotomy edge, but in some cases, this was accompanied by areas of irregular capsule margin with frayed edges, likely caused by dissipated thermal energy. The postoperative appearance and SEM features warrant further assessment of the PPC integrity and clinical correlation.

来源:Ophthalmology,查看原文。版权归原作者所有。

引用

  1. Hooshmand J, Abell RG, Goemann K, et al. Ultrastructural Integrity of Human Capsulotomies Created by a Thermal Device. Ophthalmology. 2017. PMID: 29074029. doi:10.1016/j.ophtha.2017.09.018

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。