光学相干断层扫描布鲁赫膜开口评估与立体视盘照相估算杯盘比的差异比较
本研究为可靠性分析,纳入99例患者共195只眼(116只青光眼、79只疑似青光眼),比较更新版OCT布鲁赫膜开口算法与青光眼专家立体视盘照相读取的垂直和水平杯盘比。OCT计算的杯盘比显著大于照片读取(P < .001),两种方法一致性差(ICC 0.50–0.63),29%和35%的眼垂直和水平杯盘比差异≥0.2。
要点
- OCT布鲁赫膜开口算法测得的垂直和水平杯盘比均显著大于青光眼专家照片读取(P < .001)。
- 两种方法垂直杯盘比差异≥0.2者占29%,水平杯盘比差异≥0.2者占35%,平均差异均为0.3。
- 读取者与OCT之间ICC为0.50–0.63,中等和大视盘眼垂直杯盘比差异与杯面积弱相关。
核心结论
- OCT与临床医生照片读取在杯盘比评估上一致性差。
- 两种方法垂直杯盘比差异在中和大视盘眼中取决于杯面积。
- 临床实践中就杯盘比下结论时应考虑这些差异。
研究设计与样本
- 可靠性分析,纳入99例患者共195只眼(116只青光眼、79只疑似青光眼),同一就诊日完成立体视盘照相和OCT扫描。
- 视盘照片由2名设盲青光眼专家读取估算垂直和水平杯盘比。
- 采用组内相关系数(ICC)和Bland-Altman图评估照片读取与OCT估算杯盘比的一致性。
主要发现
- 按视盘大小分层前后,OCT计算的垂直和水平杯盘比均显著大于照片读取(P < .001)。
- 两种方法杯盘比差异≥0.2者:垂直杯盘比占29%,水平杯盘比占35%,各自平均差异均为0.3。
- 读取者与OCT之间ICC为0.50–0.63。
- 中视盘眼垂直杯盘比差异大小与杯面积弱相关(r2 = 0.10,P = .008),大视盘眼亦弱相关(r2 = 0.09,P = .007)。
临床怎么用
- 原文结论指出,临床实践中就杯盘比下结论时应考虑OCT与照片读取之间的这些差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To compare the vertical and horizontal cup-to-disc ratio (VCDR, HCDR) by an updated optical coherence tomography (OCT) Bruch membrane opening (BMO) algorithm and stereoscopic optic disc photograph readings by glaucoma specialists. DESIGN: Reliability analysis. METHODS: A total of 195 eyes (116 glaucoma and 79 glaucoma suspect) of 99 patients with stereoscopic photographs and OCT scans of the optic discs taken during the same visit were compared. Optic disc photographs were read by 2 masked glaucoma specialists for VCDR and HCDR estimation. Intraclass correlation coefficient (ICC) and Bland-Altman plots were used to assess the agreement between photograph reading and OCT in estimating CDR. RESULTS: OCT images computed significantly larger VCDR and HCDR than photograph reading before and after stratifying eyes based on disc size (P < .001). The difference in CDR estimates between the 2 methods was equal to or greater than 0.2 in 29% and 35% of the eyes for VCDR and HCDR, respectively, with a mean difference of 0.3 in each case. The ICCs between the readers and OCT ranged between 0.50 and 0.63. The size of disagreement in VCDR correlated weakly with cup area in eyes with medium (r2 = 0.10, P = .008) and large (r2 = 0.09, P = .007) discs. CONCLUSIONS: OCT and photograph reading by clinicians agree poorly in CDR assessment. The difference in VCDR between the 2 methods was depended on cup area in medium and large discs. These differences should be considered when making conclusions regarding CDRs in clinical practice.