文献速递眼底病与葡萄膜炎

不同扫频源光学相干断层扫描血管成像扫描模式测量新生血管性年龄相关性黄斑变性病灶面积的比较

Comparison of Neovascular Lesion Area Measurements From Different Swept-Source OCT Angiographic Scan Patterns in Age-Related Macular Degeneration.
PubMed 文献 · Invest Ophthalmol Vis Sci · 2017-01-01

本研究对比了新生血管性年龄相关性黄斑变性患眼使用不同扫描范围(3×3、6×6、9×9、12×12 mm)的扫频源光学相干断层扫描血管成像(SS-OCTA)测量病灶面积的一致性。小病灶组(3×3 mm内)和较大病灶组(6×6 mm内)各纳入30眼。结果显示不同扫描模式测量的病灶面积相似,提示SS-OCTA可用于定量随访脉络膜新生血管的扩大。

要点

  • 小病灶组自动测量平均面积:3×3、6×6、12×12 mm扫描分别为1.11、1.14、1.27 mm²(ANOVA P<0.001)。
  • 较大病灶组自动测量平均面积:6×6、9×9、12×12 mm扫描分别为5.43、5.53、5.49 mm²(ANOVA P=0.435)。
  • 不同扫描模式测量的病灶面积相似,提示SS-OCTA可用于定量随访脉络膜新生血管扩大。

核心结论

  • 不同扫描模式测量的病灶面积相似,提示SS-OCTA成像可用于定量随访脉络膜新生血管随疾病进展的扩大。

研究设计与样本

  • 研究纳入新生血管性年龄相关性黄斑变性患者,使用100-kHz SS-OCTA仪器(PLEX Elite 9000)进行3×3、6×6、9×9和12×12 mm视野扫描。
  • 第1组纳入完全位于3×3 mm扫描范围内的小病灶,第2组纳入完全位于6×6 mm扫描范围内的较大病灶。

主要发现

  • 第1组共纳入26例患者的30眼,第2组共纳入25例患者的30眼。
  • 第1组自动测量平均病灶面积:3×3 mm扫描为1.11 mm²(SD=0.78),6×6 mm扫描为1.14 mm²(SD=0.80),12×12 mm扫描为1.27 mm²(SD=0.82);ANOVA P<0.001;事后比较3×3与6×6 mm扫描P=0.184,其余两对比较P<0.001。
  • 第2组自动测量平均病灶面积:6×6 mm扫描为5.43 mm²(SD=2.56),9×9 mm扫描为5.53 mm²(SD=2.48),12×12 mm扫描为5.49 mm²(SD=2.65);ANOVA P=0.435;事后比较6×6与9×9 mm扫描P=0.062,6×6与12×12 mm扫描P=0.553,9×9与12×12 mm扫描P=0.654。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: We compared area measurements for the same neovascular lesions imaged using swept source optical coherence tomography angiography (SS-OCTA) and enlarging scan patterns. METHODS: Patients with neovascular age-related macular degeneration were imaged using a 100-kHz SS-OCTA instrument (PLEX Elite 9000). The scanning protocols included the 3 × 3, 6 × 6, 9 × 9, and 12 × 12 mm fields of view. Two groups were studied. Group 1 included small lesions contained within the 3 \( \times \) 3 mm scan, and Group 2 included larger lesions that were fully contained within the 6 \( \times \) 6 mm scan. RESULTS: A total of 30 eyes of 26 patients were enrolled in Group 1 and 30 eyes of 25 patients were enrolled in Group 2. In Group 1, the automated mean lesion area measurements were 1.11 (SD = 0.78), 1.14 (SD = 0.80), and 1.27 (SD = 0.82) mm2 for the 3 \( \times \) 3, 6 \( \times \) 6, and 12 \( \times \) 12 mm scans, respectively (ANOVA P < 0.001; post hoc comparisons, P = 0.184, 3 \( \times \) 3 vs. 6 \( \times \) 6 mm; P < 0.001 for the other two pairs). In Group 2, the automated mean lesion area measurements were 5.43 (SD = 2.56), 5.53 (SD = 2.48), and 5.49 (SD = 2.65) mm2 for the 6 \( \times \) 6, 9 \( \times \) 9, and 12 \( \times \) 12 mm scans, respectively (ANOVA P = 0.435; post-hoc comparisons, P = 0.062, 6 \( \times \) 6 vs. 9 \( \times \) 9 mm; P = 0.553, 6 \( \times \) 6 vs. 12 \( \times \) 12 mm; P = 0.654, 9 \( \times \) 9 vs. 12 \( \times \) 12 mm). CONCLUSIONS: The similarity in lesion area measurements across different scan patterns suggests that SS-OCTA imaging can be used to follow quantitatively the enlargement of choroidal neovascularization as the disease progresses.

来源:Invest Ophthalmol Vis Sci,查看原文。版权归原作者所有。

引用

  1. Zheng F, Zhang Q, Motulsky EH, et al. Comparison of Neovascular Lesion Area Measurements From Different Swept-Source OCT Angiographic Scan Patterns in Age-Related Macular Degeneration. Invest Ophthalmol Vis Sci. 2017. PMID: 28986595. doi:10.1167/iovs.17-22506

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。