黄斑积液降低增强深度光学相干断层扫描脉络膜厚度测量的可重复性
该回顾性可靠性分析纳入124只眼,评估黄斑区不同类型视网膜积液对增强深度成像光学相干断层扫描(EDI-OCT)脉络膜厚度(CT)测量可重复性的影响。结果显示,黄斑积液会降低CT测量的可重复性,尤其在视网膜下积液和脉络膜较厚的眼中更为明显;中心凹下CT或平均CT需分别达到74.1 μm或63.9 μm的差异,才可能检测到真正的临床变化。
要点
- 黄斑积液会降低EDI-OCT脉络膜厚度测量的可重复性。
- 视网膜下积液对可重复性的影响最差,脉络膜越厚可重复性越差。
- 中心凹下CT需74.1 μm、平均CT需63.9 μm的差异才可能反映真实临床变化。
核心结论
- 黄斑积液会降低脉络膜厚度测量的可重复性,尤其在伴有视网膜下积液和脉络膜较厚的眼中更为明显。
- 在黄斑积液的眼中,中心凹下脉络膜厚度需74.1 μm或平均脉络膜厚度需63.9 μm的差异,才可能检测到真正的临床变化。
研究设计与样本
- 回顾性可靠性分析,共分析124只眼。
- 通过半自动分割分析EDI-OCT图像中的脉络膜-巩膜交界,测量中心凹下脉络膜厚度及中央3 mm水平段平均值,并记录人口学数据、黄斑中心厚度和积液类型。
- 采用组内相关系数(ICC)和可重复性系数(CR)评估测量者内和测量者间可重复性。
主要发现
- 124只眼中,60只眼(48.4%)为糖尿病黄斑水肿,32只眼(25.8%)为新生血管性年龄相关性黄斑变性,32只眼(25.8%)为其他原因导致的积液。
- 测量者间中心凹下脉络膜厚度的ICC为0.95、CR为74.1 μm;平均脉络膜厚度的ICC为0.96、CR为63.9 μm。
- 不同积液原因之间CR相似,但视网膜下积液的CR较视网膜内积液或视网膜色素上皮下积液更差。
- CR随脉络膜厚度增加而变差,但不受视网膜厚度影响。
- 测量者内可重复性总体高于测量者间,并呈现相同趋势。
关键数值与推荐
- 测量者间中心凹下脉络膜厚度ICC为0.95、CR为74.1 μm。
- 测量者间平均脉络膜厚度ICC为0.96、CR为63.9 μm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To determine if different types of retinal fluid in the central macula affect the reproducibility of choroidal thickness (CT) measurements on enhanced depth imaging optical coherence tomography (EDI-OCT). DESIGN: Retrospective reliability analysis. METHODS: EDI-OCT images were obtained and the choroidal-scleral junction was analyzed through semiautomated segmentation. CT was measured at the fovea and averaged across the central 3-mm horizontal segment. Demographic data, central macular thickness, and type of fluid present were recorded. Intragrader and intergrader repeatability were assessed using the intraclass correlation coefficient (ICC) and coefficient of repeatability (CR). RESULTS: Of 124 eyes analyzed, 60 (48.4%) had diabetic macular edema, 32 (25.8%) had neovascular age-related macular degeneration, and 32 (25.8%) had other causes of fluid. Intergrader ICC (CR) was 0.95 (74.1 μm) and 0.96 (63.9 μm) for subfoveal and average CT, respectively. CR was similar across various causes of retinal fluid, but was worst for subretinal fluid compared to intraretinal or sub-retinal pigment epithelial fluid. CR also worsened with increasing choroidal thickness, but was not affected by retinal thickness. Intragrader repeatability was generally greater than intergrader values, and followed the same trend. CONCLUSIONS: The presence of macular fluid reduces CT measurement reproducibility, particularly in eyes with subretinal fluid and greater choroidal thickness. A difference of 74.1 μm in subfoveal CT or 63.9 μm in average CT may be necessary to detect true clinical change in eyes with macular fluid.