文献速递眼底病与葡萄膜炎

亚洲患者羟氯喹视网膜病变筛查的光学相干断层扫描方案

Optical Coherence Tomography Protocols for Screening of Hydroxychloroquine Retinopathy in Asian Patients.
PubMed 文献 · Am J Ophthalmol · 2017-09-28

本研究评估48只眼(24例患者)羟氯喹视网膜病变中光感受器缺陷的分布,比较不同光学相干断层扫描方案的检出敏感度。结果显示缺陷距中心凹平均最小距离1.84±1.26 mm,部分眼需宽视野扫描才能检出,12×9 mm²容积扫描敏感度最高(100%)。作者推荐对服用羟氯喹的亚洲患者采用宽视野OCT筛查。

要点

  • 光感受器缺陷距中心凹平均最小距离为1.84±1.26 mm。
  • 水平方向15只眼(31.3%)、垂直方向17只眼(35.4%)缺陷距离大于3 mm。
  • 12×9 mm²容积扫描检出视网膜病变的敏感度最高(100%)。

核心结论

  • 作者推荐对服用羟氯喹的亚洲患者采用宽视野OCT扫描进行筛查。

研究设计与样本

  • 诊断效度评估研究,纳入48只眼(24例患者)的羟氯喹视网膜病变眼。
  • 采用谱域OCT进行6 mm水平和垂直线扫描及6×6 mm²容积扫描。
  • 采用扫频源OCT进行9 mm线扫描及6×6 mm²和12×9 mm²容积扫描。

主要发现

  • 光感受器缺陷距中心凹的平均最小距离为1.84±1.26 mm(均值±标准差)。
  • 水平方向有15只眼(31.3%)、垂直方向有17只眼(35.4%)的光感受器缺陷距离大于3 mm,仅宽视野线扫描或容积扫描可检出这些眼的缺陷。
  • 9 mm线扫描检出羟氯喹视网膜病变显著优于6 mm扫描(P < .001)。
  • 宽视野容积扫描的敏感度显著高于标准容积扫描(P = .001)。
  • 12×9 mm²容积扫描检出视网膜病变的敏感度最高(100%)。

关键数值与推荐

  • 视网膜病变检出以光感受器缺陷为指标。
  • 9 mm线扫描检出羟氯喹视网膜病变显著优于6 mm扫描(P < .001)。
  • 宽视野容积扫描敏感度显著高于标准容积扫描(P = .001)。
  • 12×9 mm²容积扫描检出敏感度为100%。

临床怎么用

  • 作者推荐对服用羟氯喹的亚洲患者采用宽视野OCT扫描进行筛查。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To investigate the distribution of outer retinal changes in hydroxychloroquine (HCQ) retinopathy and explore optical coherence tomography (OCT) protocols to maximize the sensitivity of HCQ retinopathy detection in Asian patients. DESIGN: Diagnostic validity assessment. METHODS: Setting: Institutional. PATIENT POPULATION: Forty-eight eyes (24 patients) with HCQ retinopathy underwent 6-mm horizontal and vertical line scans and 6 × 6-mm2 volume scans using spectral-domain OCT (SD-OCT), and 9-mm line scans and 6 × 6-mm2 and 12 × 9-mm2 volume scans using swept-source OCT (SS-OCT). OBSERVATION PROCEDURES: Distances from the fovea to the defective photoreceptors were measured in the temporal, nasal, superior, and inferior directions from line scan OCT images. The sensitivity of retinopathy detection, indicated by photoreceptor defects, was compared among protocols. MAIN OUTCOME MEASURES: Detection of photoreceptor defects and distances from the fovea to the defects. RESULTS: The average minimum distance from the fovea to an area of photoreceptor defects was 1.84 ± 1.26 mm (mean ± standard deviation). The distances were greater than 3 mm horizontally and vertically in 15 (31.3%) and 17 (35.4%) eyes with HCQ retinopathy, respectively, and only wide-field line or volume scans could detect defects in the eyes. The 9-mm line scans detected HCQ retinopathy significantly better than 6-mm scans (P < .001); the sensitivity of the wide volume scan was significantly greater than the standard volume scan (P = .001). The 12 × 9-mm2 volume scan detected retinopathy with the greatest sensitivity (100%). CONCLUSIONS: Our study recommends a wide-field OCT scan to screen Asian patients taking HCQ medications.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Ahn SJ, Joung J, Lim HW, et al. Optical Coherence Tomography Protocols for Screening of Hydroxychloroquine Retinopathy in Asian Patients. Am J Ophthalmol. 2017. PMID: 28964805. doi:10.1016/j.ajo.2017.09.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。