脉冲模式对横向超声效率与碎核事件的影响
该体外实验室研究将猪晶状体硬化后切成2.0×2.0毫米碎块,使用AMO WhiteStar Signature Pro配合Ellips FX手柄和0.9毫米弯Dewey针头进行超声乳化,比较连续、6毫秒通断微脉冲和50毫秒通断长脉冲三种横向超声模式的效率与碎核事件。结果显示微脉冲较长脉冲效率高43%(P=0.00003)、较连续超声高42%(P=0.000387),碎核事件也较少但差异未达统计学显著。
要点
- 微脉冲较长脉冲效率高43%(P=0.00003)。
- 微脉冲较连续超声效率高42%(P=0.000387)。
- 微脉冲碎核事件较少,但差异未达统计学显著。
核心结论
- 6毫秒通6毫秒断的微脉冲横向三维超声较长脉冲和连续超声更高效,且产生的碎核事件更少。
研究设计与样本
- 体外实验室研究,在犹他大学John A. Moran眼科中心实验室进行。
- 猪晶状体经福尔马林硬化2小时并在基础盐溶液中平衡24小时,切成2.0×2.0毫米碎块后保存于基础盐溶液。
- 使用AMO WhiteStar Signature Pro机器配合Ellips FX手柄和0.9毫米、30度斜面弯Dewey针头进行超声乳化。
- 共进行3轮实验,每轮20枚晶状体,测量效率和碎核事件,3轮分别采用连续超声、6毫秒通断微脉冲和50毫秒通断长脉冲。
主要发现
- 微脉冲较长脉冲效率高43%(P=0.00003)。
- 微脉冲较连续超声效率高42%(P=0.000387)。
- 微脉冲较长脉冲和连续超声碎核事件更少,但该差异未达统计学显著。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To evaluate the effects of micropulse, long pulse, and continuous ultrasound on transverse ultrasound using Abbott Medical Optics' (AMO) WhiteStar Signature Pro with the Ellips FX handpiece. DESIGN: In vitro laboratory study. METHODS: This study was conducted at the John A. Moran Eye Center Laboratory, University of Utah, Salt Lake City, Utah, USA. Porcine lenses were hardened in formalin for 2 hours and equilibrated in basic salt solution (BSS) over a 24-hour period. The lenses were then cubed in 2.0 × 2.0-mm pieces. These pieces were stored in BSS until the time of experimentation. The AMO WhiteStar Signature Pro machine (Abbott Medical Optics) with the Ellips FX handpiece and a 0.9-mm bent Dewey tip with a 30-degree bevel (Microsurgical Technology Inc) were used for phacoemulsification. Three runs of 20 lenses each were performed, measuring efficiency and chatter. Transverse ultrasound varied in the 3 runs and included continuous, 6 ms on/off micropulse, and 50 ms on/off long pulse. RESULTS: Micropulse was more efficient than long pulse by 43% (P = .00003) and continuous by 42% (P = .000387). There were also less chatter events with micropulse than with long-pulse and continuous ultrasound. However, this difference did not reach significance. CONCLUSION: The 6 ms on and 6 ms off micropulse transverse 3-dimensional ultrasound is more efficient and produces fewer chatter events than both long-pulse and continuous ultrasound.
引用
- Wright AJ, Bohner AD, Bernhisel AA, et al. The Effect of Pulsing on Transverse Ultrasound Efficiency and Chatter. Am J Ophthalmol. 2017. PMID: 28860045. doi:10.1016/j.ajo.2017.08.012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。