在角膜水肿情况下使用飞秒激光进行可靠的前囊切开术
研究使用Ziemer LDV Z8液体界面飞秒激光平台,在猪眼模拟角膜清晰或水肿条件下以不同能量行前囊切开术,并在人眼白内障手术中比较手工与Z8前囊切开后的TUNEL检测。结果显示水肿角膜中提高能量可提升囊切开完成率,而囊膜形态、破裂强度及细胞死亡影响有限。
要点
- 角膜清晰时,能量90%、130%、150%下囊膜边缘形态与破裂强度无差异(P=0.42)。
- 角膜水肿时,90%能量仅50%囊切开成功,130%为75%,150%为100%。
- 人眼样本中130%能量组TUNEL阳性细胞增多,150%组未见增多(P<0.05)。
核心结论
- 由于飞秒纳焦平台能量低,即使递增能量对囊膜形态和强度影响极小,对细胞死亡影响可忽略。
- 提高能量似乎可增强角膜水肿时囊切开的一致性和完成能力。
研究设计与样本
- 离体猪眼36只,于摘除后小于6小时和大于24小时模拟角膜清晰与水肿,使用Ziemer LDV Z8飞秒激光(直径5 mm,能量90%、130%或150%)行囊切开术。
- 囊膜行扫描电镜评估,破裂强度用单柱万能测试系统测定;经伦理批准,23例患者在常规白内障手术中行手工或Z8囊切开后取囊膜行TUNEL检测。
主要发现
- 角膜清晰时,能量90%、130%、150%的破裂强度分别为120、113、118 mN,差异无统计学意义(P=0.42),边缘形态无差异。
- 角膜水肿时,90%能量囊切开成功率50%,130%为75%,150%为100%;破裂强度分别为112、133、114 mN,差异无统计学意义(P=0.3)。
- 人眼样本中,TUNEL阳性细胞数手工组0.0、90%能量组0.2、130%能量组2.1、150%能量组0.6,P<0.05,130%能量组增多而150%组未见增多。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To determine the effects of the Ziemer LDV Z8 liquid interface femtosecond laser platform during capsulotomy under different energy settings in the presence of corneal edema. METHODS: Cadaveric porcine eyes (n = 36) employed at less than 6 and greater than 24 post enucleation hours to simulate clear/edematous corneas, underwent capsulotomy with the Ziemer LDV Z8 femtosecond laser (5-mm diameter, energy 90%, 130%, or 150%). Lens capsules were removed for evaluation by scanning electron microscopy and rupture strengths determined by the single column universal testing system. Following ethical approval, 23 patients had lens capsules removed during routine cataract surgery following manual or Z8 capsulotomy and subjected to TUNEL assay. RESULTS: There was no difference in edge morphology or rupture strength (120, 113, and 118 mN at increasing energy, P = 0.42) in the clear cornea. Only 50% of capsulotomies succeeded at 90% energy in an edematous cornea, improving with increased energy (75% completion at 130%, 100% at 150%). Rupture strength in edematous corneas was not significantly different at 112, 133, and 114 mN for 90%, 130%, and 150%, respectively (P = 0.3). In human samples, increased TUNEL-positive cells were seen at 130% energy, but not at 150% (0.0 manual vs. 0.2 [90%] vs. 2.1 [130%] vs. 0.6 [150%], P < 0.05). CONCLUSIONS: Because of the low energy delivered by a femtosecond nanojoule platform, even incremental increases in energy appeared to have minimal effect on lens capsule morphology and strength and negligible influence on cell death. Furthermore, increasing energy appeared to enhance consistency and the ability to complete a capsulotomy in an edematous cornea.
引用
- Williams GP, George BL, Wong YR, et al. Performing Reliable Lens Capsulotomy in the Presence of Corneal Edema With a Femtosecond Laser. Invest Ophthalmol Vis Sci. 2017. PMID: 28873174. doi:10.1167/iovs.17-22227
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。