青光眼患者自我测量眼压的研究

Measurement of Intraocular Pressure by Patients With Glaucoma.
PubMed 文献 · JAMA Ophthalmol · 2017-01-01

这项观察性研究在2016年3月1日至12月30日于大学医院青光眼科纳入100例青光眼或高眼压患者,评估其使用回弹式眼压计进行自我眼压测量的能力与接受度。73%的患者达到完全成功标准,回弹式眼压计测值平均较Goldmann压平眼压计低2.66 mm Hg,多数患者认为操作容易且舒适。

要点

  • 100例青光眼或高眼压患者中73例(73%)达到完全成功标准。
  • 回弹式眼压计测值平均较Goldmann压平眼压计低2.66 mm Hg(95%一致性界限,-3.48至8.80 mm Hg)。
  • 自我测量与研究者测量的回弹式眼压计读数组内相关系数为0.903(95%CI,0.867-0.928)。

核心结论

  • 多数患者能够进行自我眼压测量,且该方法为患者所接受。
  • 自我眼压测量有潜力提高患者参与度,同时更完整地反映眼压随时间的变化。

研究设计与样本

  • 观察性研究,连续纳入100例青光眼或高眼压患者,纳入双眼。
  • 研究于2016年3月1日至12月30日在大学医院青光眼科进行。
  • 眼压分别用Goldmann压平眼压计和回弹式眼压计评估,患者接受标准化自我眼压测量培训后在观察下自行测量。

主要发现

  • 100例患者平均(SD)年龄67.5(10.9)岁,53%为女性。
  • 73/100例患者(73%)达到完全成功标准,另有6例能使用设备但眼压读数与临床医生所测相差大于5 mm Hg。
  • 回弹式眼压计测值平均较Goldmann压平眼压计低2.66 mm Hg(95%一致性界限,-3.48至8.80 mm Hg)。
  • 自我测量与研究者测量的回弹式眼压计读数相似,组内相关系数为0.903(95%CI,0.867-0.928)。
  • 79例成功或部分成功患者中,56例(71%)认为自我眼压测量容易,73例(92%)报告舒适,相似数量愿意将来进行自我眼压测量。

关键数值与推荐

  • 完全成功定义为技术良好且眼压读数与临床医生使用同一设备所测值相差在5 mm Hg以内。
  • 采用3项问卷评估患者对自我眼压测量的看法。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

IMPORTANCE: The ability of patients to measure their own intraocular pressure (IOP) would allow more frequent measurements and better appreciation of peak IOP and IOP fluctuation. OBJECTIVE: To examine whether patients with glaucoma can perform self-tonometry using a rebound tonometer and examine patient acceptability. DESIGN, SETTING, AND PARTICIPANTS: An observational study in which IOP was assessed using Goldmann applanation tonometry and a rebound tonometer. Consecutive patients were provided with a patient information sheet and those consenting to take part in the study received standardized self-tonometry training and were then instructed to measure their own IOP under observation. This study was conducted at a glaucoma clinic at a university hospital from March 1, 2016, to December 30, 2016, and included both eyes of 100 patients with glaucoma or ocular hypertension. MAIN OUTCOMES AND MEASURES: The percentage of patients who could successfully perform self-tonometry. Complete success was defined by a good technique and an IOP reading within 5 mm Hg of that obtained by a clinician using the same device. A 3-item questionnaire was used to examine perceptions of self-tonometry among patients. RESULTS: Among the 100 patients, the mean (SD) age was 67.5 (10.9) years (53% female). A total 73 of 100 patients (73%) met the complete success criteria. An additional 6 patients could use the device but had IOP readings greater than 5 mm Hg different from those obtained by the clinician. On average, IOP by the rebound tonometer was 2.66 mm Hg lower than Goldmann applanation tonometry (95% limits of agreement, -3.48 to 8.80 mm Hg). The IOPs with the rebound tonometer were similar whether obtained by self-tonometry or investigator, with excellent reproducibility with an intraclass correlation coefficient of 0.903 (95% CI, 0.867-0.928). A total of 56 of 79 successful or partially successful patients (71%) felt self-tonometry was easy, with 73 of 79 (92%) reporting self-tonometry to be comfortable, and a similar number happy to perform self-tonometry in the future. CONCLUSIONS AND RELEVANCE: Most patients could perform self-tonometry and the method was acceptable to patients. Self-tonometry has the potential to improve patient engagement, while also providing a more complete picture of IOP changes over time.