应用扫频源光学相干断层扫描血管成像检测年龄相关性黄斑变性中未治疗脉络膜新生血管
横断面分析156只眼(107只新生血管性年龄相关性黄斑变性、49只干性年龄相关性黄斑变性),比较扫频源光学相干断层扫描血管成像(SS-OCTA,Topcon DRI Triton,1050 nm)与荧光血管造影检测未治疗脉络膜新生血管(CNV)的检出率。SS-OCTA检出107只眼中81只眼的CNV,敏感度75.7%,49只干性眼均未检出CNV(特异度100%);色素上皮脱离超过400 μm时SS-OCTA未检出未治疗CNV有统计学意义(P=0.0238)。
要点
- SS-OCTA在107只新生血管性年龄相关性黄斑变性眼中检出81只眼CNV,敏感度75.7%。
- 49只干性年龄相关性黄斑变性眼均未检出CNV,特异度100%。
- 色素上皮脱离超过400 μm与SS-OCTA未检出未治疗CNV相关(P=0.0238)。
核心结论
- Topcon的SS-OCTA不能检出全部CNV病灶,大的色素上皮脱离与信号丢失相关。
- 荧光血管造影仍是金标准,但受试SS-OCTA设备可考虑作为年龄相关性黄斑变性中可行的附加诊断工具。
研究设计与样本
- 横断面分析,纳入98例患者的156只眼,其中新生血管性年龄相关性黄斑变性107只眼、干性年龄相关性黄斑变性49只眼。
- 患眼此前在三级视网膜中心通过多模式影像(荧光素血管造影和吲哚菁绿血管造影,Heidelberg Spectralis)诊断,随后用SS-OCTA(Topcon DRI Triton,1050 nm波长)4.5 mm×4.5 mm和6 mm×6 mm黄斑立方扫描评估。
- 主要结局指标为SS-OCTA在年龄相关性黄斑变性中的敏感度和特异度,并分析影响CNV检出率的潜在因素。
主要发现
- SS-OCTA在107只新生血管性年龄相关性黄斑变性眼中检出81只眼的CNV,敏感度为75.7%。
- 49只干性年龄相关性黄斑变性眼均未检出CNV,特异度为100%。
- SS-OCTA未检出未治疗CNV在色素上皮脱离超过400 μm时具有统计学意义(P=0.0238)。
关键数值与推荐
- SS-OCTA检测CNV的敏感度为75.7%(81/107只新生血管性年龄相关性黄斑变性眼)。
- SS-OCTA检测CNV的特异度为100%(49只干性年龄相关性黄斑变性眼均未检出CNV)。
- 色素上皮脱离超过400 μm时SS-OCTA未检出未治疗CNV的P=0.0238。
临床怎么用
- 作者认为荧光血管造影仍是金标准,受试SS-OCTA设备可考虑作为年龄相关性黄斑变性中可行的附加诊断工具。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To compare the detection rate of choroidal neovascularization (CNV) in treatment-naive neovascular age-related macular degeneration by swept source optical coherence tomography angiography (SS-OCTA, Topcon's DRI Triton) working at 1,050 nm wavelength versus fluorescence angiography. METHODS: Cross-sectional analysis of 156 eyes (107 neovascular age-related macular degeneration and 49 dry AMD) in 98 patients, previously diagnosed by multimodal imaging using fluorescein (FA) and indocyanine green angiography (Heidelberg's Spectralis) in a tertiary retina center, evaluated by SS-OCTA 4.5 mm × 4.5 mm and 6 mm × 6 mm macular cubes. Main outcome measures were sensitivity and specificity of SS-OCTA in AMD. Potential factors influencing CNV detection rate were analyzed. RESULTS: Swept source optical coherence tomography angiography detected CNV in 81 of 107 eyes, resulting in a sensitivity of 75.7%. In 49 eyes with dry AMD, no CNV could be identified (specificity 100%). A statistical significance was calculated for nondetection of treatment-naive CNV by SS-OCTA in pigment epithelial detachment over 400 μm (P = 0.0238). CONCLUSION: Topcon's SS-OCTA was not able to detect all CNV lesions. Large pigment epithelial detachments were associated with signal loss. Fluorescence angiography still remains the gold standard, but the tested SS-OCTA device can be considered as a feasible additional diagnostic tool in AMD.
引用
- Ahmed D, Stattin M, Graf A, et al. DETECTION OF TREATMENT-NAIVE CHOROIDAL NEOVASCULARIZATION IN AGE-RELATED MACULAR DEGENERATION BY SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY. Retina. 2018. PMID: 28902095. doi:10.1097/IAE.0000000000001832
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。