光学相干断层扫描血管成像正面图像测量浅层黄斑视网膜血管密度的重复性与再现性
这项横断面研究纳入21名健康志愿者(42眼)和22例视网膜疾病患者(22眼),每眼用3台仪器各重复扫描3次(共9次采集),评估光学相干断层扫描血管成像(OCTA)衍生的浅层视网膜血管长度密度(VLD)和灌注密度(PD)的重复性与再现性。结果显示,连续3次扫描的组内相关系数(ICC)VLD为0.82至0.98、PD为0.83至0.95,变异系数(CV)均低于6%,但一致性随扫描模式和位置而变化。
要点
- 连续3次扫描的ICC:VLD 0.82至0.98,PD 0.83至0.95;CV:VLD 2.2%至5.9%,PD 2.4%至5.9%。
- 再现性ICC为0.62至0.95,所有组CV均小于6%。
- 3×3-mm模式内环ICC为0.92(95%CI 0.85-0.96)至0.96(95%CI 0.93-0.98);6×6-mm模式外环ICC为0.93(95%CI 0.86-0.97)至0.96(95%CI 0.92-0.98)。
核心结论
- OCTA图像可获取浅层视网膜血管的VLD和PD,具有高重复性和再现性,但可随扫描模式和位置而变化。
研究设计与样本
- 横断面研究,纳入21名健康志愿者(42眼)和22例视网膜疾病患者(22眼),包括14例年龄相关性黄斑变性、3例视网膜前膜、2例糖尿病视网膜病变、2例近视性黄斑变性和1例视网膜静脉阻塞。
- 参与者招募时间为2016年9月1日至2016年11月31日。
- 每眼用3台独立仪器各重复扫描3次,共9次采集;眼睛随机分配至3×3-mm或6×6-mm扫描模式。
- 若任何一次采集的信号强度小于7,则该眼被排除后续分析。
- 重复性定义为同一设备内测量的一致性,再现性定义为同类型设备间测量的一致性,通过组内相关系数(ICC)和变异系数评估。
主要发现
- 共纳入21名健康志愿者(8男13女;平均[SD]年龄36[6]岁)和22例视网膜疾病患者(15男7女;平均[SD]年龄79[9]岁);其中40只正常眼(42眼中)和15只视网膜疾病眼(22眼中)符合信号强度标准并纳入分析。
- 连续3次扫描的ICC:VLD为0.82至0.98,PD为0.83至0.95;CV:VLD为2.2%至5.9%,PD为2.4%至5.9%。
- 再现性方面,ICC为0.62至0.95,所有组CV均小于6%。
- 3×3-mm模式内环一致性最高,ICC为0.92(95%CI 0.85-0.96)至0.96(95%CI 0.93-0.98)。
- 6×6-mm模式外环一致性最高,ICC为0.93(95%CI 0.86-0.97)至0.96(95%CI 0.92-0.98)。
关键数值与推荐
- 信号强度小于7的采集被排除后续分析。
- 重复性通过ICC和CV评估,ICC范围VLD 0.82至0.98、PD 0.83至0.95,CV范围VLD 2.2%至5.9%、PD 2.4%至5.9%。
- 再现性通过ICC和CV评估,ICC范围0.62至0.95,CV所有组小于6%。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
IMPORTANCE: The repeatability and reproducibility of quantitative metrics from optical coherence tomographic angiography (OCTA) must be assessed before these data can be confidently interpreted in clinical research and practice. OBJECTIVE: To evaluate the repeatability and reproducibility of OCTA-derived retinal vascular quantitative metrics. DESIGN, SETTING AND PARTICIPANTS: In this cross-sectional study, 21 healthy volunteers (42 eyes) and 22 patients with retinal disease (22 eyes), including 14 with age-related macular degeneration, 3 with epiretinal membrane, 2 with diabetic retinopathy, 2 with myopic macular degeneration, and 1 with retinal vein occlusion, were enrolled. Participants were recruited from September 1 through November 31, 2016. Each eye underwent 3 repeated scans with 3 instruments for a total of 9 acquisitions. Eyes were randomly assigned to scanning with a 3 × 3-mm or 6 × 6-mm pattern. Eyes were excluded from subsequent analysis if any acquisition had a signal strength of less than 7. Repeatability (defined as the agreement in measurements within a device) and reproducibility (defined as the agreement between devices of the same type) were assessed by intraclass correlation coefficient (ICC) and coefficient of variation. EXPOSURES: All eyes underwent scanning using 3 separate devices. MAIN OUTCOMES AND MEASURES: Vessel length density (VLD) and perfusion density (PD) of the superficial retinal vasculature. RESULTS: A total of 21 healthy volunteers (8 men and 13 women; mean [SD] age, 36 [6] years) and 22 patients with retinal disease (15 men and 7 women; mean [SD] age, 79 [9] years) underwent evaluation. Of these, 40 of 42 normal eyes and 15 of 22 eyes with retinal disease met signal strength criteria and were included in this analysis. The ICC among the 3 consecutive scans ranged from 0.82 to 0.98 for VLD and from 0.83 to 0.95 for PD. The coefficient of variation (CV) ranged from 2.2% to 5.9% for VLD and from 2.4% to 5.9% for PD. For reproducibility, the ICC ranged from 0.62 to 0.95 and the CV was less than 6% in all groups. The agreement was highest for the 3 × 3-mm pattern in the inner ring (ICC range, 0.92 [95% CI, 0.85-0.96] to 0.96 [95% CI, 0.93-0.98]) and 6 × 6-mm pattern in the outer ring (ICC range, 0.93 [95% CI, 0.86-0.97] to 0.96 [95% CI, 0.92-0.98]). CONCLUSIONS AND RELEVANCE: Vessel length density and PD of the superficial retinal vasculature can be obtained from OCTA images with high levels of repeatability and reproducibility but can vary with scan pattern and location.