使用新型视频眼镜进行斜视测量
这项横断面方法比较研究纳入41例≥6岁斜视患者和17名健康志愿者,比较新型斜视视频眼镜与Hess屏试验测量9个注视方向眼偏斜的一致性。水平与垂直偏斜的组内相关系数分别为0.83(95%CI 0.77–0.88)和0.76(95%CI 0.68–0.82),视频眼镜还可用于伴有视觉抑制的共同性斜视患者及6岁儿童。
要点
- 视频眼镜与Hess屏测量水平偏斜的组内相关系数为0.83(95%CI 0.77–0.88)。
- 垂直偏斜的组内相关系数为0.76(95%CI 0.68–0.82)。
- 视频眼镜可用于Hess屏不适用的视觉抑制患者及6岁儿童。
核心结论
- 新型视频眼镜简单、快速、准确测量眼偏斜,结果与常规Hess屏试验高度相关,甚至可用于不适合Hess屏的视觉抑制患者,并适用于6岁儿童。
研究设计与样本
- 横断面方法比较研究,将新型斜视视频眼镜与现有参考标准Hess屏试验进行比较。
- 纳入41例因先天性或获得性麻痹性或共同性斜视导致眼位不正的≥6岁成人和儿童患者,以及17名健康志愿者。
- 所有参与者均接受双眼红外视频眼镜(内置激光目标投射和液晶显示快门交替遮盖眼睛)和常规Hess屏试验检查。
- 两种测试均在水平与垂直偏心度0±15°的9点目标网格上测量眼偏斜。
主要发现
- 视频眼镜与Hess屏试验在水平和垂直偏斜测量上均显示良好一致性,水平组内相关系数0.83(95%CI 0.77–0.88),垂直0.76(95%CI 0.68–0.82)。
- 两种方法均在9点网格上再现了特征性斜视模式。
- 与Hess屏试验不同,视频眼镜测量甚至可用于伴有视觉抑制的共同性斜视患者。
关键数值与推荐
- 水平偏斜组内相关系数0.83(95%CI 0.77–0.88)。
- 垂直偏斜组内相关系数0.76(95%CI 0.68–0.82)。
- 测量在水平与垂直偏心度0±15°的9点目标网格上进行。
临床怎么用
- 原文结论指出该设备简单、快速、准确,可用于视觉抑制患者及6岁儿童,但未据此给出具体推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess the validity of a novel, simplified, noninvasive test for strabismus using video goggles. DESIGN: Cross-sectional method comparison study in which the new test, the strabismus video goggles, is compared with the existing reference standard, the Hess screen test. PARTICIPANTS: We studied 41 adult and child patients aged ≥6 years with ocular misalignment owing to congenital or acquired paralytic or comitant strabismus and 17 healthy volunteers. METHODS: All participants were tested with binocular infrared video goggles with built-in laser target projection and liquid crystal display shutters for alternate occlusion of the eyes and the conventional Hess screen test. In both tests, ocular deviations were measured on a 9-point target grid located at 0±15° horizontal and vertical eccentricity. MAIN OUTCOME MEASURES: Horizontal and vertical ocular deviations at 9 different gaze positions of each eye were measured by the strabismus video goggles and the Hess screen test. Agreement was quantified as the intraclass correlation coefficient. Secondary outcomes were the utility of the goggles in patients with visual suppression and in children. RESULTS: There was good agreement between the strabismus video goggles and the Hess screen test in the measurements of horizontal and vertical deviation (intraclass correlation coefficient horizontal 0.83, 95% confidence interval [0.77, 0.88], vertical 0.76, 95% confidence interval [0.68, 0.82]). Both methods reproduced the characteristic strabismus patterns in the 9-point grid. In contrast to Hess screen testing, strabismus video goggle measurements were even possible in patients with comitant strabismus and visual suppression. CONCLUSIONS: The new device is simple and is fast and accurate in measuring ocular deviations, and the results are closely correlated with those obtained using the conventional Hess screen test. It can even be used in patients with visual suppression who are not suitable for the Hess screen test. The device can be applied in children as young as 6 years of age.