使用光学相干断层扫描评估飞秒激光角膜基质内切口位置
回顾性病例系列纳入77例患者77眼,用光学相干断层扫描评估Catalys飞秒激光在白内障手术中制作的成对角膜基质内切口。实际未切割前部、未切割后部、中点深度及侧切角度均与计划参数存在显著差异,且同一眼内成对切口的中点深度与侧切角度无相关性。作者认为需进一步改进以确保该设备制作基质内切口的精确位置。
要点
- 实际未切割前部为17.2±5.8%,与计划20%显著不同(P<0.05)。
- 实际中点深度42.3±6.6%,较计划50%显著偏前(P<0.05)。
- 同一眼成对切口的中点深度与侧切角度无相关性(r=-0.063和-0.067,P>0.05)。
核心结论
- 基质内切口中点深度较计划50%显著偏前,同一眼内成对切口位置无相关性,需进一步改进以确保该设备制作基质内切口的精确位置。
研究设计与样本
- 回顾性病例系列,纳入77例患者77眼,使用Catalys飞秒激光制作成对基质内切口,术后3周或以上行光学相干断层扫描评估4项参数并与计划值比较。
主要发现
- 实际未切割前部为17.2±5.8%(范围7.2%-36.9%),与计划20%显著不同(P<0.05)。
- 实际未切割后部为32.5±8.8%(范围6.0%-57.9%),与计划20%显著不同(P<0.05)。
- 实际中点深度为42.3±6.6%(范围25.5%-65.4%),与计划50%显著不同(P<0.05)。
- 实际侧切角度为88.5°±5.6°(范围71°-114°),与计划90°显著不同(P<0.05)。
- 50眼成对切口间中点深度与侧切角度无相关性(r=-0.063和-0.067,P>0.05)。
关键数值与推荐
- 计划基质内切口参数为未切割前部20%、未切割后部20%、中点深度50%、侧切角度90°。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文结论指出需进一步改进以确保该设备制作基质内切口的精确位置。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To use optical coherence tomography (OCT) to evaluate the femtosecond laser intrastromal incisions made during cataract surgery to reduce corneal astigmatism. DESIGN: Retrospective case series. PARTICIPANTS: Seventy-seven eyes of 77 patients. METHODS: Paired intrastromal incisions were created using the Catalys femtosecond laser (Abbott Medical Optics, Inc., Santa Ana, CA). The planned intrastromal incision parameters were 20% uncut anterior, 20% uncut posterior, midpoint depth of 50%, and 90° side cut angle. Optical coherence tomography scans were obtained 3 weeks or more after surgery to assess these 4 parameters, and actual values were compared with intended values. MAIN OUTCOME MEASURES: Percentages of uncut anterior and posterior tissue, midpoint depth, and degrees of side cut angle. RESULTS: The mean values were 17.2±5.8% (range, 7.2%-36.9%) for uncut anterior, 32.5±8.8% (range, 6.0%-57.9%) for uncut posterior, and 42.3±6.6% (range, 25.5%-65.4%) for midpoint depth, which all were significantly different from the planned parameters (all P < 0.05). The mean side cut angle was 88.5°±5.6° (range, 71°-114°) and was significantly different from the planned side cut angle of 90° (P < 0.05). In 50 eyes that had paired intrastromal incisions scanned by the OCT, there was no correlation between the paired incisions for midpoint depth and side cut angle (correlation coefficient, r = -0.063 and -0.067, respectively; P > 0.05). CONCLUSIONS: The intrastromal incision midpoint depth was significantly more anterior than the planned depth of 50%. The locations of paired intrastromal incisions in each eye were not correlated. Further improvements are needed to ensure the precise location of the intrastromal incisions made with this device.