便携式脑机接口检测青光眼:客观评估视野功能损害
这项病例对照研究纳入33例青光眼患者的62只眼和17名健康参与者的30只眼,比较便携式脑机接口设备与标准自动视野计区分青光眼眼与健康眼的能力。全局脑机接口多焦稳态视觉诱发电位参数的受试者工作特征曲线下面积为0.92,大于标准自动视野计平均偏差的0.81、平均敏感度的0.80和模式标准差0.77,后两者与脑机接口差异有统计学意义。作者认为该设备可能有助于评估与视野刺激相关的脑电反应,并区分青光眼性神经病变眼与健康眼。
要点
- 全局脑机接口多焦稳态视觉诱发电位参数区分青光眼眼与健康眼的曲线下面积为0.92(95%CI 0.86-0.96)。
- 标准自动视野计平均偏差曲线下面积0.81(95%CI 0.72-0.90),平均敏感度0.80(95%CI 0.69-0.88),模式标准差0.77(95%CI 0.66-0.87)。
- 脑机接口与标准自动视野计的扇区测量之间未见统计学显著差异。
核心结论
- 脑机接口设备可能有助于评估与视野刺激相关的脑电反应。
- 该设备在临床环境中可区分青光眼性神经病变眼与健康眼。
- 作者认为需进一步研究脑机接口设备用于家庭测试及检测随时间变化的视野功能损害的可行性。
研究设计与样本
- 病例对照研究,纳入33例青光眼患者的62只眼和17名健康参与者的30只眼。
- 青光眼基于对视盘立体照片的设盲分级诊断。
- 所有参与者在3个月内接受脑机接口设备与标准自动视野计测试。
- 脑机接口设备整合可穿戴无线干电极脑电图和眼电图系统及手机式头戴显示器,检测与视野刺激相关的多焦稳态视觉诱发电位。
- 在10名青光眼参与者的20只眼中进行重复测试以评估重复性,共3次测量,间隔一周。
主要发现
- 全局脑机接口多焦稳态视觉诱发电位参数的受试者工作特征曲线下面积为0.92(95%CI 0.86-0.96)。
- 标准自动视野计平均偏差曲线下面积0.81(95%CI 0.72-0.90),平均敏感度0.80(95%CI 0.69-0.88),模式标准差0.77(95%CI 0.66-0.87)。
- 脑机接口全局参数曲线下面积大于标准自动视野计平均敏感度(P=0.03)和模式标准差(P=0.01)。
- 脑机接口与标准自动视野计的扇区测量之间未见统计学显著差异。
- 全局和扇区参数的组内相关系数为0.74至0.92,平均变异系数为3.03%至7.45%。
关键数值与推荐
- 33例青光眼参与者中19例(58%)为白人,12例(36%)为黑人,2例(6%)为亚裔。
- 17名健康眼参与者中9例(53%)为白人,8例(47%)为黑人,无亚裔。
临床怎么用
- 原文未据此给出推荐。
局限性
- 作者认为需进一步研究脑机接口设备用于家庭测试及检测随时间变化的视野功能损害的可行性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
IMPORTANCE: The current assessment of visual field loss in diseases such as glaucoma is affected by the subjectivity of patient responses and the lack of portability of standard perimeters. OBJECTIVE: To describe the development and initial validation of a portable brain-computer interface (BCI) for objectively assessing visual function loss. DESIGN, SETTING, AND PARTICIPANTS: This case-control study involved 62 eyes of 33 patients with glaucoma and 30 eyes of 17 healthy participants. Glaucoma was diagnosed based on a masked grading of optic disc stereophotographs. All participants underwent testing with a BCI device and standard automated perimetry (SAP) within 3 months. The BCI device integrates wearable, wireless, dry electroencephalogram and electrooculogram systems and a cellphone-based head-mounted display to enable the detection of multifocal steady state visual-evoked potentials associated with visual field stimulation. The performances of global and sectoral multifocal steady state visual-evoked potentials metrics to discriminate glaucomatous from healthy eyes were compared with global and sectoral SAP parameters. The repeatability of the BCI device measurements was assessed by collecting results of repeated testing in 20 eyes of 10 participants with glaucoma for 3 sessions of measurements separated by weekly intervals. MAIN OUTCOMES AND MEASURES: Receiver operating characteristic curves summarizing diagnostic accuracy. Intraclass correlation coefficients and coefficients of variation for assessing repeatability. RESULTS: Among the 33 participants with glaucoma, 19 (58%) were white, 12 (36%) were black, and 2 (6%) were Asian, while among the 17 participants with healthy eyes, 9 (53%) were white, 8 (47%) were black, and none were Asian. The receiver operating characteristic curve area for the global BCI multifocal steady state visual-evoked potentials parameter was 0.92 (95% CI, 0.86-0.96), which was larger than for SAP mean deviation (area under the curve, 0.81; 95% CI, 0.72-0.90), SAP mean sensitivity (area under the curve, 0.80; 95% CI, 0.69-0.88; P = .03), and SAP pattern standard deviation (area under the curve, 0.77; 95% CI, 0.66-0.87; P = .01). No statistically significant differences were seen for the sectoral measurements between the BCI and SAP. Intraclass coefficients for global and sectoral parameters ranged from 0.74 to 0.92, and mean coefficients of variation ranged from 3.03% to 7.45%. CONCLUSIONS AND RELEVANCE: The BCI device may be useful for assessing the electrical brain responses associated with visual field stimulation. The device discriminated eyes with glaucomatous neuropathy from healthy eyes in a clinically based setting. Further studies should investigate the feasibility of the BCI device for home-based testing as well as for detecting visual function loss over time.