持续释放环孢素A的沟槽型青光眼引流装置减少兔眼术后瘢痕形成

Grooved Glaucoma Drainage Devices That Continuously Deliver Cyclosporine A Decrease Postsurgical Scar Formation in Rabbit Eyes.
PubMed 文献 · Invest Ophthalmol Vis Sci · 2017-01-01

研究在18只兔眼(每组6眼)中植入普通青光眼引流装置(GDD)、沟槽型GDD(G-GDD)及带环孢素A-聚乳酸-羟基乙酸共聚物(CsA-PLGA)复合涂层的沟槽型GDD(CsA@G-GDD),随访至12周。结果显示G-GDD与CsA@G-GDD组滤过泡存活时间显著更长、眼压显著更低,CsA@G-GDD组滤过泡显著更高、滤过通道更通畅,瘢痕形成少于普通GDD组。作者认为带CsA-PLGA涂层的G-GDD可能成为现有青光眼滤过装置的替代选择。

要点

  • G-GDD与CsA@G-GDD组滤过泡存活时间显著长于GDD组,眼压显著低于GDD组(均P<0.001)。
  • CsA@G-GDD组滤过泡高度显著高于GDD组和G-GDD组(P=0.003)。
  • CsA@G-GDD组前房造影显示通畅滤过通道多于GDD组(P=0.032),G-GDD与CsA@G-GDD组术后瘢痕形成较GDD组少。

核心结论

  • 与普通GDD相比,带CsA-PLGA涂层的G-GDD抑制术后瘢痕形成并提高手术成功率,可能成为现有青光眼滤过装置的替代选择。

研究设计与样本

  • 在18只兔眼(每组6眼)中分别植入普通GDD、G-GDD和CsA@G-GDD,随访至植入后12周,比较眼压、滤过泡存活时间、滤过泡形态及前房反应。
  • 采用Indiana滤过泡外观分级量表量化滤过泡形态,前房造影检查滤过通路是否阻塞并测定造影剂引流时间与弥散面积,苏木精-伊红染色与免疫组化评估瘢痕形成。

主要发现

  • G-GDD组与CsA@G-GDD组滤过泡存活时间显著长于GDD组,眼压显著低于GDD组(均P<0.001)。
  • CsA@G-GDD组滤过泡高度显著高于GDD组和G-GDD组(P=0.003)。
  • 前房造影显示CsA@G-GDD组通畅滤过通道多于GDD组(P=0.032)。
  • G-GDD组与CsA@G-GDD组术后瘢痕形成范围小于GDD组。

关键数值与推荐

  • 滤过泡存活时间与眼压:G-GDD组和CsA@G-GDD组均优于GDD组(均P<0.001)。
  • 滤过泡高度:CsA@G-GDD组高于GDD组和G-GDD组(P=0.003)。
  • 通畅滤过通道数:CsA@G-GDD组多于GDD组(P=0.032)。

临床怎么用

  • 作者认为带CsA-PLGA涂层的G-GDD可能成为现有青光眼滤过装置的替代选择。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To study the functional outcomes of filtration surgery by implanting improved glaucoma drainage devices (GDDs) comprising surface grooves and a composite coating of cyclosporine A (CsA) and PLGA in experimental rabbit eyes. METHODS: Improved GDDs were designed and prepared by modifying normal GDD with surface grooves and a CsA-PLGA composite coating. Normal GDDs, grooved GDDs (G-GDDs), and G-GDDs with a CsA-PLGA composite coating (CsA@G-GDD) were implanted into 18 rabbit eyes (six eyes per group). The intraocular pressure (IOP), bleb survival time, bleb morphology, and anterior chamber reactions were assessed for up to 12 weeks after GDD implantation. The IOPs were compared statistically among the three groups. Bleb morphology was quantified using the Indiana Bleb Appearance Grading Scale. Anterior chamber radiography was performed to check whether the filtrating pathway was blocked, and determine the drainage time and diffusion area of the contrast agent. Hematoxylin and eosin staining and immunohistochemistry were conducted to assess how the GDDs slowed or prevented scar formation. RESULTS: The improved GDDs were successfully prepared and implanted in 18 rabbit eyes without severe surgical complications. Bleb survival time was significantly longer and IOP was significantly lower in the G-GDD and CsA@G-GDD groups compared with the GDD group (all, P < 0.001). Blebs were significantly higher in the CsA@G-GDD group than in the GDD and G-GDD groups (P = 0.003). Anterior chamber radiography revealed more unobstructed filtration channels in the CsA@G-GDD group than in the GDD group (P = 0.032). Postsurgical scar formation was less extensive in the G-GDD and CsA@G-GDD groups than in the GDD group. CONCLUSIONS: Compared with the normal GDDs, G-GDDs with a CsA-PLGA coating inhibited postsurgical scar formation and improved the surgical success rate, and might represent an alternative to existing glaucoma filtration devices.