气动眼压计与数字化Schiøtz眼压计测量房水流出易度的比较

Comparison of Aqueous Outflow Facility Measurement by Pneumatonography and Digital Schiøtz Tonography.
PubMed 文献 · Invest Ophthalmol Vis Sci · 2017-01-01

研究纳入28名健康参与者共56只眼,比较气动眼压计与数字化Schiøtz眼压计测量的房水流出易度。使用标准假设巩膜硬度系数时,Schiøtz眼压计测得流出易度0.29±0.09 μL/min/mm Hg,气动眼压计为0.24±0.08 μL/min/mm Hg(P<0.001);使用计算巩膜硬度系数时分别为0.23±0.08与0.21±0.07 μL/min/mm Hg(P=0.003)。两种设备测量的流出易度相关,但Schiøtz眼压计报告的值约高10%至20%,测量结果不能直接比较。

要点

  • 标准巩膜硬度系数下,Schiøtz眼压计测得的房水流出易度高于气动眼压计(P<0.001)。
  • 使用计算巩膜硬度系数时,两种设备测得的流出易度差异缩小(P=0.003)。
  • 两种设备测量的流出易度相关(假设巩膜硬度r=0.60,计算巩膜硬度r=0.70),但数值不能直接比较。

核心结论

  • 气动眼压计与Schiøtz眼压计测量的房水流出易度相关,但Schiøtz眼压计在使用标准方法时报告的流出易度约高10%至20%,两种设备的测量结果不能直接比较。

研究设计与样本

  • 研究纳入28名健康参与者共56只眼,年龄41至68岁,比较气动眼压计与数字化Schiøtz眼压计测量的房水流出易度。

主要发现

  • 使用标准假设巩膜硬度系数时,Schiøtz眼压计测得流出易度0.29±0.09 μL/min/mm Hg,气动眼压计为0.24±0.08 μL/min/mm Hg(P<0.001)。
  • 平均计算巩膜硬度系数为0.028±0.01 μL-1,使用该系数时Schiøtz眼压计测得流出易度0.23±0.08 μL/min/mm Hg,气动眼压计为0.21±0.07 μL/min/mm Hg(P=0.003)。
  • 假设巩膜硬度时两种设备测量的流出易度相关(r=0.60,P<0.001),计算巩膜硬度时也相关(r=0.70,P<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: It is not known if outflow facilities measured by pneumatonography and Schiøtz tonography are interchangeable. In this study we compared outflow facility measured by pneumatonography to outflow facility measured by digital Schiøtz tonography. METHODS: Fifty-six eyes from 28 healthy participants, ages 41 to 68 years, were included. Intraocular pressure (IOP) was measured in the sitting and supine positions with a pneumatonometer. With the subject in the supine position, IOP was recorded for 2 minutes by using a pneumatonometer with a 10-g weight and for 4 minutes by using a custom digital Schiøtz tonometer. Outflow facility was determined from the changes in pressure and intraocular volume and a standard assumed ocular rigidity coefficient for each instrument, respectively, and by using an ocular rigidity coefficient calculated by measuring pressure without and with a weight added to the pneumatonometer tip. RESULTS: The outflow facility was 0.29 ± 0.09 μL/min/mm Hg by Schiøtz tonography and 0.24 ± 0.08 μL/min/mm Hg by pneumatonography (P < 0.001) when using the standard assumed constant ocular rigidity coefficient. Mean calculated ocular rigidity coefficient was 0.028 ± 0.01 μL-1, and outflow facility determined by using this coefficient was 0.23 ± 0.08 μL/min/mm Hg by Schiøtz tonography and 0.21 ± 0.07 μL/min/mm Hg by pneumatonography (P = 0.003). Outflow facilities measured by the two devices were correlated when the ocular rigidity was assumed (r = 0.60, P < 0.001) or calculated (r = 0.70, P < 0.001). CONCLUSIONS: Outflow facilities measured by pneumatonography were correlated with those measured by Schiøtz tonography, but Schiøtz tonography reported approximately 10% to 20% higher facilities when using the standard method. When ocular rigidity was determined for each eye, differences were smaller. Measurements from these devices cannot be compared directly.