文献速递青光眼

家用自测回弹式眼压计在青光眼与非青光眼患者中的测量准确性

[Accuracy of home-use self-rebound-tonometer for intraocular pressure measurement in glaucoma and non-glaucoma patients].
PubMed 文献 · Zhonghua Yan Ke Za Zhi · 2026-01-01

该横断面研究纳入中山大学中山眼科中心青光眼科门诊连续招募的69名受试者(132只眼),其中非青光眼15名(30只眼)、青光眼54例(102只眼),比较受试者自测、医师测量与Goldmann压平眼压计(GAT)三种眼压测量结果。三种测量值差异无统计学意义,家用自测回弹式眼压计与GAT具有较好一致性,提示其测量结果可靠。

要点

  • 受试者自测、医师测量与GAT眼压值分别为(15.74±4.40)、(15.70±5.22)、(15.88±5.10)mmHg,三者差异无统计学意义(F=0.12,P=0.874)。
  • 以GAT为标准,受试者自测平均偏差为(0.14±3.41)mmHg,与GAT相关性r=0.75(P<0.001)。
  • 受试者自测误差与GAT测量值正向关联(β=0.35,95%CI:0.21~0.49,P<0.001)。

核心结论

  • 家用自测回弹式眼压计在青光眼与非青光眼受试者中与Goldmann压平眼压计测量值具有较好一致性,测量结果可靠。

研究设计与样本

  • 横断面研究,2025年8月至11月于中山大学中山眼科中心青光眼科门诊连续招募非青光眼和青光眼受试者。
  • 共纳入69名受试者(132只眼),年龄(38.16±11.57)岁,其中非青光眼15名(30只眼)、青光眼54例(102只眼)。
  • 受试者经标准化培训与认证后依次进行受试者自测、医师测量及Goldmann压平眼压计测量。
  • 分析采用配对样本单因素方差分析、Pearson相关性系数、Bland-Altman及95%一致性界限分析,以及单变量和多变量回归分析。
  • 左右眼测量误差比较纳入双眼,其余统计分析每名受试者仅纳入右眼数据。

主要发现

  • 受试者自测、医师测量及GAT眼压值分别为(15.74±4.40)、(15.70±5.22)、(15.88±5.10)mmHg,三者差异无统计学意义(F=0.12,P=0.874)。
  • GAT与医师测量、GAT与受试者自测、医师测量与受试者自测的相关系数r分别为0.86、0.75、0.78(P均<0.001)。
  • 以GAT为标准,医师和受试者使用家用眼压计的偏差分别为(0.19±2.70)和(0.14±3.41)mmHg,医师与受试者使用家用眼压计测量的偏差为(-0.05±3.27)mmHg。
  • 受试者自测左右眼测量偏差分别为(-0.05±3.21)和(-0.03±2.72)mmHg,差异无统计学意义(P=0.872)。
  • 单因素分析中受试者自测误差与GAT测量值正向关联(β=0.35,95%CI:0.21~0.49,P<0.001);医师与受试者测量误差与最佳矫正视力(β=5.26,95%CI:2.04~8.48,P=0.002)及GAT测量值(β=0.24,95%CI:0.09~0.38,P=0.002)正向关联;青光眼与测量误差增大相关(β=2.10,95%CI:0.27~3.92,P=0.028)。

关键数值与推荐

  • 66名(95.65%)受试者通过家用自测回弹式眼压计标准化培训与认证(中文摘要载66名即95.7%受试者、126只眼即92.4%通过认证)。
  • GAT与家用眼压计测量值的Bland-Altman 95%一致性界限分析结果原文未列出具体界限数值。
  • 单因素回归中受试者自测误差与GAT测量值β=0.35(95%CI:0.21~0.49,P<0.001)。

临床怎么用

  • 家用自测回弹式眼压计经标准化培训认证后可用于青光眼与非青光眼人群家庭眼压自测,结果与GAT一致性较好。
  • 解读自测结果时应考虑GAT眼压水平、最佳矫正视力及是否为青光眼对测量误差的影响。
  • 受试者自测左右眼测量误差无显著差异,支持双眼家庭自测的可行性。

局限性

  • 为单中心横断面研究,样本量69名受试者(132只眼),非青光眼受试者仅15名(30只眼)。
  • 原文未报告Bland-Altman 95%一致性界限的具体数值。
  • 原文未报告多因素分析中除GAT测量值外其他因素的独立作用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objective: To compare the measurement results of a home-use self-tonometer and the Goldmann applanation tonometer (GAT) in glaucoma and non-glaucoma subjects, evaluate the accuracy of the home-use self-tonometer, and investigate potential factors influencing the measurement outcomes. Methods: It was a cross-sectional study. Non-glaucoma and glaucoma subjects were consecutively recruited from the Glaucoma Clinic of the Zhongshan Ophthalmic Center, Sun Yat-sen University, between August 2025 and November 2025. After receiving standardized training and certification for the home-use self-tonometer, subjects underwent three sequential intraocular pressure (IOP) measurements: self-measurement by the subject, measurement by a physician (using the home tonometer), and GAT measurement. Differences, correlations, and agreement among the three IOP measurements were analyzed using paired one-way analysis of variance (ANOVA), Pearson correlation coefficients, and Bland-Altman analysis with 95% limits of agreement. Paired t-tests were used to analyze whether there were differences in measurement errors between the left and right eyes during subject self-measurement. Univariate and multivariate regression analysis were performed to evaluate factors associated with measurement errors of the home-use self-rebound tonometer Both eyes were included when analyzing the difference in measurement errors between left and right eyes; for all other statistical analyses, only data from one eye per subject were included. Results: A total of 69 subjects (132 eyes) were recruited, with a mean age of (38.16±11.57) years, including 15 non-glaucoma subjects (30 eyes) and 54 glaucoma patients (102 eyes). Sixty-six subjects (95.65%) passed the certification. The IOP values obtained from subject self-measurement, physician measurement, and GAT were (15.74±4.40) mmHg, (15.70±5.22) mmHg (1 mmHg=0.133 kPa), and (15.88±5.10) mmHg, respectively, with no statistically significant differences among the three methods (F=0.12, P=0.874). Strong correlations were observed between GAT measurements and physician measurements, between GAT measurements and subject self-measurements, and between physician measurements and subject self-measurements (r=0.86, 0.75, and 0.78, respectively; all P<0.001). Using GAT values as the reference, the mean biases between GAT and physician measurements and between GAT and subject self-measurements were (0.19±2.70) mmHg and (0.14±3.41) mmHg, respectively, while the mean bias between the physician and the subject using the home-use self-tonometer was (-0.05±3.27) mmHg. Compared with GAT measurements, the measurement errors of subject self-measurement in the right and left eyes were (-0.05±3.21) mmHg and (-0.03±2.72) mmHg, respectively, with no significant difference between eyes (P=0.872). In univariate linear regression analysis, the measurement error of subject self-measurement was positively associated with GAT-measured IOP (β=0.35, 95%CI: 0.21-0.49, P<0.001). The measurement differences between physician and subject measurements using the home-use self-rebound tonometer were positively associated with best-corrected visual acuity (BCVA) and GAT-measured IOP (β=5.26 and 0.24, respectively; 95%CI: 2.04-8.48 and 0.09-0.38, respectively; both P=0.002). Glaucoma was associated with increased measurement differences between physician and subject measurements using the home-use self-rebound tonometer (β=2.10, 95%CI: 0.27-3.92, P=0.028). In multivariate linear regression analysis, GAT-measured IOP was an independent factor associated with physician measurement error (β=-0.29, 95%CI: 0.06-0.55, P=0.022). Conclusion: The home-use self-rebound-tonometer showed good agreement with the Goldmann applanation tonometer in both glaucoma and non-glaucoma subjects, indicating that the measurement results obtained using the home-use self-rebound tonometer are reliable. 目的: 对比青光眼与非青光眼患者使用家用自测回弹式眼压计和Goldmann压平眼压计的测量结果,评估家用自测回弹式眼压计测量结果的准确性。 方法: 横断面研究。2025年8月至11月期间在中山大学中山眼科中心青光眼科门诊患者中连续招募非青光眼和青光眼受试者。受试者经家用自测回弹式眼压计标准化培训与认证后,依次进行3种眼压测量:受试者自测、医师测量及Goldmann压平眼压计测量。采用配对样本单因素方差分析、Pearson相关性系数、Bland-Altman及95%一致性界限分析3种眼压测量值之间的差异、相关性及一致性;使用配对t检验分析受试者自测眼压左右眼间测量误差是否存在差异;采用单变量及多变量回归分析评估家用自测回弹式眼压计测量误差的影响因素。在分析左右眼间测量误差差异时纳入受试者的双眼进行分析,其余统计分析如受试者双眼入组,选取右眼数据。 结果: 共招募69名受试者(132只眼),年龄(38.16±11.57)岁,其中15名(30只眼)非青光眼受试者,54例(102只眼)青光眼患者。其中,66名(95.7%)受试者126只眼(92.4%)通过认证。受试者自测、医师测量及Goldmann压平眼压计测量的眼压值分别为(15.74±4.40)、(15.70±5.22)和(15.88±5.10)mmHg(1 mmHg=0.133 kPa),三者间差异无统计学意义(F=0.12,P=0.874)。压平眼压计测量与医师测量及受试者自测、受试者自测与医师测量所得眼压值之间具有强相关性(r分别为0.86、0.75、0.78,P均<0.001)。以压平眼压计测量值为标准,医师和受试者使用家用自测回弹式眼压计值的偏差分别为(0.19±2.70)和(0.14±3.41)mmHg,而医师和受试者使用家用自测回弹式眼压计测量的偏差为(-0.05±3.27)mmHg。与压平眼压计测量值相比,受试者自测左右眼间测量偏差分别为(-0.05±3.21)和(-0.03±2.72)mmHg,差异无统计学意义(P=0.872)。单因素线性回归分析显示,受试者自测的误差与压平眼压测量值存在正向关联(β=0.35,95%CI:0.21~0.49,P<0.001),医师和受试者使用家用自测回弹式眼压计测量误差与最佳矫正视力、压平眼压测量值存在正向关联(β分别为5.26及0.24,95%CI分别为2.04~8.48及0.09~0.38,P均为0.002);患有青光眼是医师与受试者使用家用自测眼压计测量误差增大的影响因素(β=2.10,95%CI:0.27~3.92,P=0.028)。多因素线性回归分析显示,压平眼压测量值是医师测量误差的独立正向相关因素(β=-0.29,95%CI:0.06~0.55,P=0.022)。 结论: 青光眼与非青光眼患者使用家用自测回弹式眼压计与Goldmann压平眼压计的眼压测量值具有较好的一致性,家用自测回弹式眼压计的测量结果可靠。.

来源:Zhonghua Yan Ke Za Zhi,查看原文。版权归原作者所有。

引用

  1. Zhuo KC, Luo N, Yan YT, et al. [Accuracy of home-use self-rebound-tonometer for intraocular pressure measurement in glaucoma and non-glaucoma patients]. Zhonghua Yan Ke Za Zhi. 2026. PMID: 42706142. doi:10.3760/cma.j.cn112142-20251114-00483

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。