文献速递眼底病与葡萄膜炎白内障与屈光手术

附加式双人工晶状体系统首次人体植入:光学模拟、临床前验证与临床可行性

First-in-Human Implantation of an Add-On Dual Intraocular System Combining a Standard IOL and the SING-IMT™ Telescope: Optical Simulation, Preclinical Validation, and Clinical Feasibility.
PubMed 文献 · Retina · 2026-09-23

本研究评估囊袋内单焦人工晶状体联合睫状沟植入SING-IMT微型望远镜的双植入系统。光学模拟显示双植入配置在晚期AMD康复范围内维持视网膜成像质量,偏位0.5 mm内光学耐受良好,倾斜则快速损失中频对比度;猪眼实验证实睫状沟固定稳定;首次人体植入6个月随访显示远视力从20/200提升至20/125、近视力从56提升至15 Colenbrander单位、阅读速度从术前不可评估提升至20词/分,无器械相关不良事件。

要点

  • 双植入系统光学模拟视网膜弥散斑半径低于视网膜采样极限,预期视力约提升2行Snellen。
  • 偏位≤0.5 mm光学耐受良好,倾斜则导致中频对比度快速下降。
  • 首次人体植入6个月无器械相关不良事件,远近视力与阅读速度均改善。

核心结论

  • 囊袋内单焦人工晶状体联合睫状沟SING-IMT植入作为微创眼内放大策略,对特定晚期AMD人工晶状体眼患者具有可行性及短期安全性。

研究设计与样本

  • 包含光学模拟、猪眼临床前实验及首次人体植入的综合评价研究;首次人体植入随访6个月,采用超声生物显微镜与光学相干断层扫描评估前节。

主要发现

  • 光学模拟显示视网膜弥散斑均方根半径为18.7 µm(0°)与20.4 µm(2.5°),低于估计的视网膜采样极限(约49 µm)。
  • 猪眼实验证实睫状沟固定稳定,未破坏囊袋内人工晶状体或前节结构。
  • 首次人体植入无术中并发症;6个月远视力从20/200提升至20/125,近视力从56提升至15 Colenbrander单位,阅读速度从术前不可评估提升至20词/分,无器械相关不良事件。

关键数值与推荐

  • 光学模拟预期视力提升约2行Snellen。
  • 偏位达0.5 mm时光学耐受良好;倾斜则产生快速中频对比度损失。
  • 随访6个月,远视力20/200至20/125,近视力56至15 Colenbrander单位,阅读速度术前不可评估至20词/分。

临床怎么用

  • 提示对特定晚期AMD人工晶状体眼患者,睫状沟SING-IMT植入可作为微创眼内放大选择,需关注倾斜对对比度的影响并短期随访安全性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate a dual-implant system combining a capsular-bag monofocal intraocular lens with a sulcus-implanted small-incision new-generation implantable miniature telescope (SING-IMT). METHODS: Optical simulations quantified retinal spot size, modulation transfer function (MTF), and tolerance to misalignment. Preclinical porcine experiments evaluated surgical feasibility, anatomical stability, and early safety. First-in-human (FiH) implantation assessed surgical outcomes, anterior segment behaviour using ultrasound biomicroscopy and optical coherence tomography, functional vision, and safety over 6 months. RESULTS: Optical modelling demonstrated that the dual-implant configuration maintained retinal image quality within the rehabilitative regime for advanced AMD, with root-mean-square retinal spot radii of 18.7 µm (0°) and 20.4 µm (2.5°), below the estimated retinal sampling limit (∼49 µm), consistent with an expected visual acuity improvement of approximately 2 Snellen lines. Decentration up to 0.5 mm was optically well tolerated, whereas tilt produced rapid mid-frequency contrast loss. In vivo porcine studies confirmed stable sulcus seating without disruption of the capsular-bag IOL or anterior segment structures. FiH implantation was completed without intraoperative complications. At 6 months, distance visual acuity improved from 20/200 to 20/125, near acuity from 56 to 15 Colenbrander units, and reading speed from non-assessable preoperatively to 20 words/min, with no device-related adverse events. CONCLUSION: This integrated evaluation supports the feasibility and short-term safety of sulcus-based SING-IMT implantation as a minimally invasive intraocular magnification strategy for selected pseudophakic patients with advanced AMD.

来源:Retina,查看原文。版权归原作者所有。

引用

  1. Rizzo S, Sasso P, Nepita I, et al. First-in-Human Implantation of an Add-On Dual Intraocular System Combining a Standard IOL and the SING-IMT™ Telescope: Optical Simulation, Preclinical Validation, and Clinical Feasibility. Retina. 2026. PMID: 42776856. doi:10.1097/IAE.0000000000005023

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。