单焦点、双焦点与景深延长型人工晶状体早期临床效果评估
非随机对照试验纳入63例白内障术后患者,比较四类人工晶状体植入后3个月的主观与客观视觉结局。EDOF型提供更优中距离视力及连续视程,双焦点客观光学性能更佳,非衍射型EDOF主观视觉功能评分呈较高趋势但组间VF-14差异无统计学意义。
要点
- 两组EDOF中距离视力优于单焦点和双焦点(P<0.001)。
- 双焦点CDVA最优且MTF截止频率与SR显著更高。
- 非衍射型EDOF球差更高但VF-14评分最高。
核心结论
- EDOF型人工晶状体可提供更优的中距离视力及更宽连续功能性视程。
- 双焦点人工晶状体客观光学性能更佳。
- 人工晶状体选择应综合功能性视程、光学质量及患者个体视觉需求。
研究设计与样本
- 非随机对照临床试验,纳入2025年4月至12月北京同仁眼科中心白内障手术患者67例,最终63例(63眼)完成3个月随访分析。
- 按植入人工晶状体类型分为单焦点组15眼、双焦点组17眼、衍射型EDOF组14眼、非衍射型EDOF组17眼。
- 术后3个月评估视力、离焦曲线、客观视觉质量及VF-14问卷。
主要发现
- 术后3个月UDVA四组分别为0.11±0.11、0.04±0.08、0.09±0.07及0.10±0.11,差异无统计学意义(P=0.054)。
- 双焦点组CDVA最优(-0.02±0.07),显著优于单焦点组(0.07±0.09)和衍射型EDOF组(0.05±0.05)(P=0.033)。
- 两组EDOF组DCIVA(衍射型0.12±0.07、非衍射型0.16±0.10)均优于单焦点组(0.32±0.18)和双焦点组(0.38±0.16)(P<0.001)。
- 单焦点组DCNVA最差(0.60±0.23),其余三组间差异无统计学意义。
- 非衍射型EDOF组球差显著高于其余三组(P<0.05),总高阶像差显著高于双焦点组和衍射型EDOF组(均P<0.05),但VF-14评分最高。
关键数值与推荐
- 衍射型EDOF组在2.00 D以内近视范围内维持≤0.2 logMAR视力,非衍射型EDOF组在1.50 D以内维持该水平。
- 术后等效球镜:单焦点组(-0.15±0.44)D、双焦点组(0.12±0.37)D呈正视,非衍射型EDOF组(-0.63±0.51)D、衍射型EDOF组(-0.73±0.41)D呈轻度近视(P<0.001)。
- 双焦点组MTF截止频率及SR显著优于单焦点组和非衍射型EDOF组(均P<0.05)。
- MTF截止频率和SR与眼内高阶像差独立相关。
临床怎么用
- 根据患者功能性视程需求选择人工晶状体:重视中距离视力可选EDOF型,重视客观光学质量可选双焦点型。
- 非衍射型EDOF虽球差较高,但主观视觉功能评分呈较高趋势,夜间驾驶及精细手工表现优异。
- 术后等效球镜设计需考虑不同类型人工晶状体的屈光状态差异。
局限性
- 非随机对照设计,样本量有限(最终63眼),随访仅3个月。
- 原文未报告各组基线等效球镜等全部基线特征的具体数值。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective: To compare the subjective and objective visual outcomes after implantation of monofocal, bifocal, and extended depth-of-focus (EDOF) intraocular lenses (IOLs). Methods: This non-randomized controlled clinical trial enrolled cataract patients scheduled for phacoemulsification and IOL implantation at Beijing Tongren Eye Center, Capital Medical University, between April and December 2025. According to the implanted IOL type, patients were assigned to monofocal, bifocal, diffractive EDOF, and non-diffractive EDOF groups. Visual acuity, defocus curves, objective visual quality, and the Visual Function Index-14 (VF-14) questionnaire were evaluated at 3 months postoperatively. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), distance-corrected intermediate visual acuity (DCIVA), and distance-corrected near visual acuity (DCNVA) were measured and recorded in logarithm of the minimum angle of resolution (logMAR). Defocus curves were obtained from +1.00 to -4.00 D in 0.50-D steps under best distance correction. Intraocular wavefront aberrations and modulation transfer function (MTF) were measured at a 4.0-mm pupil diameter using an optical path difference-based aberrometer. Objective scatter index, MTF cut-off frequency, and Strehl ratio (SR) were assessed using a double-pass optical quality analysis system. Intergroup comparisons were performed using one-way analysis of variance or the Kruskal-Wallis H test with Bonferroni post hoc correction. Pearson correlation analysis and linear regression models were used to identify factors associated with intraocular higher-order aberrations (HOAs). Results: A total of 67 patients (67 eyes) were enrolled. Except three patients lost to follow-up and one patient incomplete data, 63 patients (63 eyes) were included in the analysis, with 15 eyes in the monofocal group, 17 eyes in the bifocal group, 14 eyes in the diffractive EDOF group, and 17 eyes in the non-diffractive EDOF group. The mean age was (67.24±6.85) years. There were 17 men and 46 women. Baseline characteristics were comparable among the four groups (all P>0.05). At 3 months postoperatively, UDVA did not differ significantly among the groups (P=0.054). The bifocal group achieved the best CDVA (-0.02±0.07), which was significantly better than that of the monofocal group (0.07±0.09) and the diffractive EDOF group (0.05±0.05) (P=0.033). The bifocal group also demonstrated significantly higher MTF cut-off frequency and SR than the monofocal and non-diffractive EDOF groups (both P<0.05). Both EDOF groups showed significantly better DCIVA than the monofocal and bifocal groups (both P<0.001). On defocus curve analysis, the diffractive EDOF group maintained visual acuity of 0.2 logMAR or better within a defocus range of up to -2.00 D, whereas the non-diffractive EDOF group maintained this level within -1.50 D. The monofocal group exhibited the poorest DCNVA (0.60±0.23), while no significant difference was observed among the other three groups. Postoperative spherical equivalent refraction was close to emmetropia in the monofocal [(-0.15±0.44) D] and bifocal [(0.12±0.37) D] groups, whereas mild myopia was observed in the diffractive EDOF [(-0.73±0.41) D] and non-diffractive EDOF [(-0.63±0.51) D] groups (P<0.001). The spherical aberration was significantly higher in the non-diffractive EDOF group than in the other three groups (all P<0.05). Nevertheless, this group achieved the highest VF-14 score and showed better performance in night driving and fine handwork tasks. Correlation and regression analyses demonstrated that the MTF cut-off and SR were independently associated with intraocular HOAs. Conclusions: EDOF IOLs provided superior intermediate visual acuity and a broader continuous functional range of vision. The non-diffractive EDOF group showed a trend toward better subjective visual function, although VF-14 scores did not differ significantly among the four groups. Bifocal IOLs demonstrated superior objective optical performance. Selection of IOLs should be individualized according to the functional visual range, optical quality, and patient-specific visual demands. 目的: 比较单焦点、双焦点及景深延长型(EDOF)人工晶状体(IOL)植入术后主观与客观视觉结局。 方法: 非随机对照临床试验。纳入2025年4月至12月在首都医科大学北京同仁眼科中心拟行超声乳化白内障吸除联合IOL植入术的白内障患者,根据植入IOL类型分为单焦点组、双焦点组、衍射型EDOF组和非衍射型EDOF组。术后3个月评估视力、离焦曲线、客观视觉质量和视觉功能指数14量表(VF-14)。采用标准对数视力表检测未矫正远视力(UDVA)、矫正远视力(CDVA)、远矫正中间视力(DCIVA)和远矫正近视力(DCNVA),以最小分辨角的对数记录;离焦曲线在最佳远矫正基础上以0.50 D为间隔从-4.00~+1.00 D测定;眼内波前像差及调制传递函数(MTF)采用基于光程差的波前像差仪在4.0 mm瞳孔直径下测量;客观散射指数(OSI)、MTF截止频率及斯特列尔比(SR)采用双通道客观光学质量分析仪进行系统评估。组间比较行单因素方差分析或Kruskal-Wallis H检验,事后行Bonferroni多重比较;变量相关性采用Pearson相关分析,影响因素采用线性回归模型。 结果: 本研究共纳入白内障患者67例,其中3例未完成术后3个月随访,1例资料缺失,未纳入最终分析。最终63例(63只眼)完成研究并纳入统计分析,年龄为(67.24±6.85)岁;男性17例,女性46例。其中单焦点组15例、双焦点组17例、衍射型EDOF组14例及非衍射型EDOF组17例。研究对象4个组间年龄、性别分布等差异均无统计学意义(均P>0.05),组间具有可比性。术后3个月,4个组UDVA分别为0.11±0.11、0.04±0.08、0.09±0.07及0.10±0.11,差异无统计学意义(P=0.054)。双焦点组CDVA最优(-0.02±0.07),显著优于单焦点组(0.07±0.09)和衍射型EDOF组(0.05±0.05)(P=0.033),且MTF截止频率及SR显著优于单焦点组和非衍射型EDOF组(均P<0.05)。两组EDOF组DCIVA(衍射型EDOF组0.12±0.07、非衍射型EDOF组0.16±0.10)均优于单焦点组(0.32±0.18)和双焦点组(0.38±0.16)(P<0.001),衍射型EDOF组在2.00 D以内的近视范围内、非衍射型EDOF组在1.50 D以内的近视范围内维持≤0.2的视力。DCNVA方面,单焦点组最差(0.60±0.23),其余3个组间差异无统计学意义。术后等效球镜度数单焦点组(-0.15±0.44)D和双焦点组(0.12±0.37)D呈正视状态,非衍射型EDOF组(-0.63±0.51)D和衍射型EDOF组(-0.73±0.41)D呈轻度近视状态(P<0.001)。非衍射型EDOF组球差显著高于其余3个组(P<0.05),总高阶像差显著高于双焦点组和衍射型EDOF组(均P<0.05),但VF-14评分最高,夜间驾驶及精细手工方面表现优异。相关性及回归分析进一步证实,MTF截止频率和SR与眼内高阶像差独立相关。 结论: EDOF型IOL可提供更优的中距离视力及连续功能性视程,非衍射型EDOF 组患者主观视觉功能评分呈较高趋势,但四组间VF-14评分差异无统计学意义;双焦点IOL客观光学性能更佳。IOL的选择应综合考虑功能性视程、光学质量及患者的个体视觉需求。.
引用
- Chen T, Li EJ, Wang ZY, et al. [Evaluation of early clinical outcomes after implantation of monofocal, bifocal, and extended depth-of-focus intraocular lenses]. Zhonghua Yan Ke Za Zhi. 2026. PMID: 42552122. doi:10.3760/cma.j.cn112142-20260315-00102
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。