晶状体折射率对白内障患者光学生物测量的影响
本研究对7例白内障患者(56至77岁)术前用全眼OCT联合Scheimpflug成像测量晶状体折射率(NLens),并对比实测NLens与固定NLens(1.41)对眼轴长度(AL)和晶状体厚度(LT)的影响。结果显示个体NLens在1.400至1.434间变化,可导致AL和LT测量产生-0.070至0.027 mm的个体差异。结论强调NLens对精确AL和LT测量及人工晶状体度数计算至关重要。
要点
- 白内障眼个体间晶状体折射率(NLens)差异明显,范围为1.400至1.434。
- 使用实测NLens与固定NLens(1.41)相比,AL和LT测量值存在-0.070至0.027 mm的个体变化。
- 全眼OCT联合Scheimpflug成像可活体测量NLens并用于AL和LT计算。
核心结论
- 晶状体折射率在白内障患者个体间差异显著,对精确测量眼轴长度和晶状体厚度起关键作用。
- 全眼OCT联合Scheimpflug成像系统可用于活体测定NLens并计算AL和LT。
- 白内障手术中应使用精确的AL和LT测量值进行人工晶状体度数计算。
研究设计与样本
- 研究纳入7例白内障患者(年龄56至77岁,平均64.7岁),术前使用全眼OCT联合Scheimpflug成像系统采集图像。
- 通过自定义MatLab程序追踪从角膜到视网膜的中心光线,求解平均晶状体折射率(NLens)。
- 分别用实测NLens和固定NLens(1.41)转换光程为几何长度,计算并比较AL和LT。
主要发现
- 使用实测NLens时,平均AL为23.897 ± 0.854 mm,平均LT为4.309 ± 0.438 mm。
- 使用固定NLens(1.41)时,平均AL为23.918 ± 0.875 mm,平均LT为4.330 ± 0.461 mm。
- 个体眼NLens范围为1.400至1.434,与固定NLens相比导致AL和LT测量变化范围为-0.070至0.027 mm。
临床怎么用
- 白内障术前生物测量应考虑个体晶状体折射率差异,以提高AL和LT的准确性。
- 全眼OCT联合Scheimpflug成像可用于活体测量NLens,辅助人工晶状体度数计算。
局限性
- 样本量仅7例白内障患者,年龄56至77岁,结论推广需谨慎。
- 研究未报告重复性、再现性或长期随访数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Measurement of the axial length (AL) of the eye and the crystalline lens thickness (LT) relies on information of the refractive index of the crystalline lens for converting optical path-length (OPL) to geometric size. The purpose of this study was to measure the lens refractive index (NLens) in vivo for patients with cataract and evaluate the influence of the NLens on AL and LT measurements. METHODS: Cataract eyes of 7 patients, aged 56 to 77 years old (mean = 64.7 years), were imaged with a combined whole-eye optical coherence tomography (OCT) and Scheimpflug Imaging (SI) system before surgery. The images were processed using a custom MatLab program to trace the central light-ray from the cornea to the retina and to solve the average lens refractive index (NLens). The OPLs of other ocular segments, except the lens, were converted to geometric lengths by using the refractive indices from a model eye. The AL and LT were calculated with the measured NLens and also a constant NLens (1.41) for comparison. RESULTS: For the patients with cataract, mean AL and LT were 23.897 ± 0.854 and 4.309 ± 0.438 mm, respectively, when the measured NLens was used for OPL conversion, and changed to 23.918 ± 0.875 mm and 4.330 ± 0.461 mm, respectively, when the NLens was set at 1.41. The lens NLens from individual eyes varied from 1.400 to 1.434 among the patients, and caused a substantial individual variation in the AL and LT measurements, as compared with those with a constant NLens in a range from -0.070 to 0.027 mm. CONCLUSIONS: The Scheimpflug imaging-equipped whole-eye OCT system provides a useful technique to measure AL and LT with the NLens determined for the cataract eye in vivo. The NLens varies substantially from individual eye to individual eye among the patients with cataract, and plays a critical role for deriving accurate AL and LT measurements. During cataract surgery the accurate AL and LT measurements should be used for IOL power calculation.
引用
- He JC, Hoover A, Gameiro G, et al. Influence of Refractive Index of the Crystalline Lens on Optical Biometric Measurements for Cataract Patients. Invest Ophthalmol Vis Sci. 2026. PMID: 42644678. doi:10.1167/iovs.67.10.61
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。