文献速递白内障与屈光手术

散光人工晶体再对齐计算器的准确性:基于矢量的双变量比较

Accuracy of Toric Intraocular Lens Realignment Calculators: A Vector-Based Bivariate Comparison of Current Calculation Methods.
PubMed 文献 · Am J Ophthalmol · 2026-07-11

本研究回顾性比较了五种散光人工晶体再对齐计算器在49眼残留散光中的表现。AstiFix、Gatinel-TOR和Ophtera-AA的预测误差分布几乎相同,轴向一致性极佳(ICC≥0.997);TorAlign和IOLCon-LPC的轴向推荐差异较大。

要点

  • AstiFix、Gatinel-TOR和Ophtera-AA三种计算器轴向推荐高度一致(ICC≥0.997)。
  • TorAlign轴向推荐显著不同,一致性较低(ICC=0.816-0.822)。
  • 重复生物测量未提高预测准确性。

核心结论

  • 基于再对齐前主觉验光的计算器(AstiFix、Gatinel-TOR、Ophtera-AA)提供高度一致的轴向推荐,未检测到系统偏差。
  • IOLCon-LPC和尤其是TorAlign显示出发散的轴向推荐和更高的总方差。

研究设计与样本

  • 回顾性比较准确性研究,纳入49例患者49眼。

主要发现

  • AstiFix、Gatinel-TOR和Ophtera-AA的双变量误差分布几乎相同,质心幅度小(0.14-0.15D),平均矢量PE低(0.43-0.45D)。
  • AstiFix、Ophtera-AA和Gatinel-TOR的轴向一致性极佳(ICC≥0.997),超过93%的眼在≤5°以内。
  • TorAlign轴向推荐显著不同,一致性较低(ICC=0.816-0.822)。
  • IOLCon-LPC与再对齐专用工具的一致性最低。
  • 各公式平均矢量PE无显著差异。

关键数值与推荐

  • 质心幅度0.14-0.15D,平均矢量PE 0.43-0.45D(AstiFix、Gatinel-TOR、Ophtera-AA)。
  • 轴向一致性ICC≥0.997,>93%眼在≤5°以内(AstiFix、Ophtera-AA、Gatinel-TOR)。
  • TorAlign轴向一致性ICC=0.816-0.822。
  • 基于中位数多边形分析,所有公式与零矢量无显著偏差(所有P≥.284)。

临床怎么用

  • 对于白内障术后残留散光需行人工晶体再对齐时,可优先选用AstiFix、Gatinel-TOR或Ophtera-AA,因其轴向推荐高度一致。
  • TorAlign和IOLCon-LPC的轴向推荐与专用工具差异较大,临床使用时需谨慎解读。

局限性

  • 回顾性研究设计。
  • 样本量较小(49眼)。
  • 仅使用主觉验光和有晶状体/无晶状体生物测量数据。
  • 未报告重复测量对预测准确性的改善效果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To compare currently available calculation methods using full vector and bivariate analysis of astigmatic prediction error (PE) for planning toric intraocular lens (IOL) realignment in eyes with residual astigmatism after phacoemulsification surgery. DESIGN: Retrospective comparative accuracy study. METHODS: Predicted postrealignment astigmatism and recommended rotation axes were calculated using AstigmatismFix (AstiFix), Gatinel Toric IOL Optimal Rotation Calculator (Gatinel-TOR), Ophtera AxisAligner (Ophtera-AA), Toric Aligner (TorAlign), and IOLCon Lens Power Calculator (IOLCon-LPC). Astigmatic PE was evaluated using vector magnitude and bivariate double-angle analysis (Wilcox-Holladay-Wang-Koch). Vector PEs, centroids, and marginal variances were compared. Agreement of recommended axes was assessed using intraclass correlation coefficients (ICC). Calculations were performed using subjective refraction, and phakic and pseudophakic biometric data where applicable. RESULTS: Forty-nine eyes of 49 patients were included. AstiFix, Gatinel-TOR, and Ophtera-AA showed nearly identical bivariate error distributions with small centroid magnitudes (0.14-0.15D), and low mean vector PE (0.43-0.45D). No formula demonstrated a statistically significant deviation from the zero-vector using median-based polygon analysis (all P ≥ .284). Axis agreement among AstiFix, Ophtera-AA, and Gatinel-TOR was excellent (ICC ≥ 0.997), with > 93% of eyes within ≤ 5°. TorAlign demonstrated significantly different axis recommendations and lower agreement (ICC = 0.816-0.822). IOLCon-LPC showed the lowest agreement with realignment-specific tools. Mean vector PE did not differ significantly among tested formulas. Repeated pseudophakic biometric measurements did not improve prediction accuracy. CONCLUSIONS: Realignment-specific calculators incorporating prerealignment subjective refraction (AstiFix, Gatinel-TOR, Ophtera-AA) provide highly consistent axis recommendations with no systematic bias detected by median-based polygon analysis. IOLCon-LPC and especially TorAlign demonstrated divergent axis recommendations and higher total variance.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Kaiser KP, Wendelstein JA, Sobhi M, et al. Accuracy of Toric Intraocular Lens Realignment Calculators: A Vector-Based Bivariate Comparison of Current Calculation Methods. Am J Ophthalmol. 2026. PMID: 42435830. doi:10.1016/j.ajo.2026.07.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。