视网膜双折射扫描设备与传统自动验光设备的诊断准确性比较
这项前瞻性随机掩蔽诊断准确性队列研究纳入195名转诊至小儿眼科门诊的18岁及以下儿童,比较Spot Vision Screener(传统自动验光设备)与Blinq vision scanner(视网膜双折射扫描仪)检测弱视、斜视或显著屈光不正(VSRE)的性能。结果显示传统设备检测弱视和VSRE的敏感性更高,特异性无差异,视网膜双折射扫描仪的AUROC更低,尤其对弱视。
要点
- 传统自动验光设备检测弱视和VSRE的敏感性显著高于视网膜双折射扫描仪。
- 两种设备检测斜视的敏感性及所有疾病特异性均无差异。
- 视网膜双折射扫描仪对弱视的AUROC显著低于传统设备。
核心结论
- 传统自动验光设备检测弱视和屈光不正的总体诊断敏感性高于视网膜双折射扫描仪,所有疾病的特异性无差异。
- 传统自动验光设备可能更适用于全面的儿童视力筛查。
研究设计与样本
- 前瞻性、随机、掩蔽的诊断准确性队列研究,纳入195名18岁及以下新转诊至加州大学旧金山分校小儿眼科门诊的儿童。
- 139名参与者(平均年龄49.7月,标准差35.6月;75名女性,占54.0%)完成了两种光筛查测试和全面眼科检查。
主要发现
- 弱视、斜视和VSRE的患病率分别为15.1%、32.4%和30.1%。
- 传统自动验光设备检测弱视的敏感性比视网膜双折射扫描仪高29%(95%CI,-2%至50%;P=0.03)。
临床怎么用
- 在儿童视力筛查中,传统自动验光设备可能比视网膜双折射扫描仪更可靠地检测弱视和显著屈光不正。
局限性
- 本研究为单中心诊断准确性研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
IMPORTANCE: Photoscreening devices are commonly used to identify children at risk for eye disease, although they often demonstrate low specificity as a trade-off for high sensitivity. A newly introduced photoscreening device that uses a newer mechanism, retinal birefringence scanning, purports to detect amblyopia and strabismus directly, and may theoretically increase specificity. OBJECTIVE: To compare the diagnostic performance of 2 photoscreening devices, the Spot Vision Screener (Welch Allyn), hereafter referred to as a traditional autorefraction device, and the Blinq vision scanner (Rebion), hereafter referred to as a retinal birefringence scanner, for detecting amblyopia, strabismus, or visually significant refractive error (VSRE). DESIGN, SETTING, AND PARTICIPANTS: This was a prospective, randomized, masked, diagnostic accuracy cohort study. The study included consecutive children 18 years or younger referred as a new patient to the University of California, San Francisco, pediatric ophthalmology clinic between January 2024 and February 2025 for suspected vision problems. EXPOSURES: Each participant underwent screening with the traditional autorefraction device and the retinal birefringence scanner in randomized testing order, followed by a comprehensive eye examination. Sensitivity, specificity, and area under the receiver operating characteristic curve (AUROC) were calculated for each device relative to the clinical reference standard. MAIN OUTCOMES AND MEASURES: Sensitivity, specificity, and AUROC of both screening devices to detect amblyopia, strabismus, or VSRE. RESULTS: Of 195 enrolled participants (median [IQR] age, 50 [25-66] months), 139 (mean [SD] age, 49.7 [35.6] months; 75 female [54.0%]) completed both photoscreening tests. The prevalence of amblyopia, strabismus, and VSRE was 15.1%, 32.4%, and 30.1%, respectively. The traditional autorefraction device demonstrated higher sensitivity than the retinal birefringence scanner for detecting amblyopia (29%; 95% CI, -2% to 50%; P = .03) and VSRE (41%; 95% CI, 17%-59%; P < .001), with no difference in sensitivity for detecting strabismus identified (-6%; 95% CI, -28% to 18%; P = .69). No differences in specificity were identified. Across all vision disorders, the AUROC for the retinal birefringence scanner was lower than that for the traditional autorefraction device, particularly for amblyopia (-0.13; 95% CI, -0.23 to -0.03; P = .01). CONCLUSIONS AND RELEVANCE: In this single-site diagnostic accuracy study, the traditional autorefraction device demonstrated greater overall diagnostic sensitivity than the retinal birefringence scanner for detecting amblyopia and refractive errors, with no difference in specificity for all conditions. These findings suggest that the traditional autorefraction device may be more reliable for comprehensive pediatric vision screening.
引用
- Oatts JT, Takla P, Yu HN, et al. Diagnostic Accuracy of a Retinal Birefringence Scanning Device Compared With a Traditional Autorefraction Device. JAMA Ophthalmol. 2026. PMID: 42166152. doi:10.1001/jamaophthalmol.2026.1531
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。