Dream-OCT与Pentacam测量角膜生物学参数的可重复性及一致性研究

[Repeatability of the new anterior segment optical coherence tomographic device and its consistency with the Scheimpflug imaging device in measuring corneal biometric parameters].
PubMed 文献 · Zhonghua Yan Ke Za Zhi · 2026-01-01

本研究纳入100例近视患者右眼,比较Dream-OCT与Pentacam测量角膜生物学参数的可重复性和一致性。两种仪器均具良好可重复性,WTW及角膜前、后表面屈光力和散光度数一致性良好可互换,但CCT、散光轴向及全角膜屈光力和散光度数一致性差,不建议相互替代。

要点

  • Dream-OCT与Pentacam测量角膜参数的ICC分别≥0.840和≥0.830,可重复性均良好。
  • 两仪器测量WTW及角膜前、后表面屈光力和散光度数一致性良好,临床可互换使用。
  • 两仪器测量CCT、散光轴向及全角膜屈光力和散光度数一致性差,不建议相互替代。

核心结论

  • Dream-OCT与Pentacam在角膜生物学参数测量中均展现良好可重复性。
  • 两仪器在WTW及角膜前、后表面屈光力和散光度数测量中一致性良好,可互换使用。
  • 两仪器在CCT、角膜散光轴向及全角膜屈光力和散光度数测量中一致性差,临床不建议相互替代。

研究设计与样本

  • 横断面研究,连续纳入2024年7至9月天津医科大学眼科医院行角膜屈光手术的近视患者右眼100只。
  • 纳入男性50例、女性50例,年龄(23.35±3.82)岁,用两种仪器各连续测量3次角膜生物学参数。
  • 采用ICC、配对t检验或Wilcoxon检验及Bland-Altman分析评估可重复性、差异性和一致性。

主要发现

  • Pentacam测量各角膜生物学参数ICC≥0.830,Dream-OCT测量ICC≥0.840,后者略高。
  • Dream-OCT测量CCT(537.76±25.89)μm小于Pentacam(550.05±25.26)μm,差异有统计学意义(P<0.05)。
  • Dream-OCT测量WTW(11.78±0.39)mm及角膜前、后表面和全角膜Ks、Kf、Km均大于Pentacam测量值,差异均有统计学意义(均P<0.05)。
  • 其余角膜生物学参数两仪器测量值差异无统计学意义(均P>0.05)。

关键数值与推荐

  • CCT的95%一致性界限(LOA)为-3.514~28.094 μm,较宽。
  • 角膜前表面散光轴向95%LOA为-29.504°~33.123°,后表面为-10.370°~11.650°,全角膜为-29.926°~34.169°,均较宽。
  • 全角膜Ks、Kf、Km的95%LOA分别为-1.770~0.439 D、-1.535~0.297 D、-1.479~0.303 D,散光度数为-0.917~1.063 D,均较宽。
  • WTW及角膜前、后表面Ks、Kf、Km及其散光度数的95%LOA较窄。

临床怎么用

  • 临床测量WTW及角膜前、后表面屈光力和散光度数时,Dream-OCT与Pentacam可互换使用。
  • 测量CCT、角膜散光轴向及全角膜屈光力和散光度数时,两仪器不可相互替代,需固定使用同一设备随访。

局限性

  • 本研究仅纳入近视患者右眼,样本年龄较轻,结论外推需谨慎。
  • 原文未报告其他局限性信息。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objective: To investigate the repeatability and consistency of corneal biological parameters measured by the new anterior segment optical coherence tomographic device (Dream-OCT) and the Scheimpflug imaging device (Pentacam). Methods: This was a cross-sectional study. Clinical data of myopic patients (right eyes) who underwent corneal refractive surgery at the Eye Hospital of Tianjin Medical University from July to September 2024 were collected using the consecutive enrollment method, including corneal biological parameters measured continuously three times by Dream-OCT and Pentacam. The central corneal thickness (CCT), horizontal corneal diameter (WTW), steep keratometry (Ks), flat keratometry (Kf), and mean keratometry (Km) of the anterior corneal surface, posterior corneal surface, and total cornea were recorded, as well as the astigmatism degree, astigmatism axis of the anterior corneal surface, posterior corneal surface, and total cornea, and the astigmatism degree component J0 at the 0° or 90° axis and the astigmatism degree component J45 at the 45° or 135° axis. The intraclass correlation coefficient (ICC), paired t test, or Wilcoxon test, and Bland-Altman analysis were used to assess the repeatability, difference, and agreement of the data obtained by the two devices. Results: A total of 100 patients with myopia (100 right eyes) were enrolled, including 50 males and 50 females, with an average age of (23.35±3.82) years. The ICC of various corneal biological parameters measured by Pentacam and Dream-OCT were ≥0.830 and ≥0.840, respectively, with the ICC values of Dream-OCT being slightly higher than those of Pentacam. The CCT measured by Dream-OCT [(537.76±25.89) μm] was lower than that measured by Pentacam [(550.05±25.26) μm], while the WTW [(11.78±0.39) mm], Ks [(43.97±1.31), (-6.48±0.21), (43.64±1.37) D], Kf [(42.63±1.10), (-6.08±0.20), (42.39±1.22) D], and Km [(43.27±1.23), (-6.29±0.20), (43.00±1.24) D] of the anterior corneal surface, posterior corneal surface, and total cornea measured by Dream-OCT were all higher than the corresponding values measured by Pentacam, with statistically significant differences (all P<0.05). There were no statistically significant differences in the other corneal biological parameters between the two instruments (all P>0.05). The 95% limits of agreement between the two instruments were wide for CCT (-3.514 to 28.094 μm), anterior corneal astigmatism axis (-29.504° to 33.123°), posterior corneal astigmatism axis (-10.370° to 11.650°), total corneal astigmatism axis (-29.926° to 34.169°), total corneal Ks (-1.770 to 0.439 D), Kf (-1.535 to 0.297 D), Km (-1.479 to 0.303 D), and astigmatism power (-0.917 to 1.063 D). In contrast, the 95% limits of agreement were narrow for WTW and Ks, Kf, Km, and astigmatism power of the anterior and posterior corneal surfaces. Conclusions: Both Dream-OCT and Pentacam demonstrated good repeatability in measuring corneal biometric parameters. The two instruments showed excellent consistency in measuring WTW, as well as the refractive power and astigmatic power of the anterior and posterior corneal surfaces, and thus can be used interchangeably in clinical practice. However, their agreement was poor for measurements of CCT, astigmatism axis, total corneal refractive power, and astigmatic power, suggesting that the two devices are not interchangeable for these parameters. 目的: 探讨新型眼前节相干光层析成像仪(Dream-OCT)与眼前节三维分析仪(Pentacam)测量角膜生物学参数的可重复性和一致性。 方法: 横断面研究。采用连续纳入法收集2024年7至9月在天津医科大学眼科医院行角膜屈光手术的近视眼患者(右眼)临床资料,包括使用Dream-OCT和Pentacam连续测量3次的角膜生物学参数,记录角膜中央厚度(CCT)、角膜水平直径(WTW),角膜前和后表面及全角膜的陡峭轴(Ks)、平坦轴(Kf)及平均角膜屈光力(Km),角膜前和后表面及全角膜的散光度数、散光轴向以及角膜散光度数0°或90°轴向分量J0和45°或135°轴向分量J45。分别采用组内相关系数(ICC)、配对样本t检验或Wilcoxon符号秩检验、Bland-Altman分析法对2个仪器测量数据的可重复性、差异性和一致性进行统计分析和评估。 结果: 共纳入100例近视眼患者(100只右眼),其中男性50例,女性50例;年龄为(23.35±3.82)岁。使用Pentacam测量各角膜生物学参数的ICC≥0.830,使用Dream-OCT测量角膜生物学参数的ICC≥0.840。Dream-OCT测量的CCT[(537.76±25.89)μm]小于Pentacam测量值[(550.05±25.26)μm],而WTW[(11.78±0.39)mm]及角膜前和后表面及全角膜的Ks[(43.97±1.31)、(-6.48±0.21)、(43.64±1.37)D]、Kf[(42.63±1.10)、(-6.08±0.20)、(42.39±1.22)D]、Km[(43.27±1.23)、(-6.29±0.20)、(43.00±1.24)D]均大于 Pentacam测量值,差异均有统计学意义(均P<0.05);其余角膜生物学参数2个仪器测量值的差异无统计学意义(均P>0.05)。2个仪器测量 CCT(-3.514~28.094 μm)、角膜前表面散光轴向(-29.504°~33.123°)、角膜后表面散光轴向(-10.370°~11.650°)以及全角膜散光轴向(-29.926~34.169°),全角膜Ks(-1.770~0.439 D)、Kf(-1.535~0.297 D)、Km(-1.479~0.303 D)和散光度数(-0.917~1.063 D)的95%一致性界限(LOA)较宽,而WTW以及角膜前和后表面Ks、Kf、Km及其散光度数的95%LOA较窄。 结论: Dream-OCT和Pentacam在角膜生物学参数测量中均展现出较好的可重复性。2个仪器测量的WTW、角膜前和后表面屈光力和散光度数具有良好的一致性,临床实践中可互换使用;而在CCT、角膜散光轴向以及全角膜屈光力和散光度数的测量中,2个仪器的一致性差,临床应用中不建议相互替代。.