两种折射型扩展景深人工晶状体设计的个体内比较

Intraindividual Comparison of Two Refractive Extended-Depth-of-Focus Intraocular Lens Designs.
PubMed 文献 · Am J Ophthalmol · 2026-04-25

前瞻性随机个体内对照研究纳入40例双侧年龄相关性白内障患者,一眼植入波前塑形折射型EDF IOL(DEN00V),对侧眼植入扇区分段折射型EDF IOL(AN6V)。37例完成3个月随访,主要终点UIVA在AN6V组显著更优,其余多数视力指标及对比敏感度无显著组间差异。

要点

  • 两种折射型EDF IOL均提供扩展功能性视力范围并保持对比敏感度。
  • 主要终点UIVA在AN6V组显著优于DEN00V组。
  • CDVA和UNVA的差异经多重比较校正后未保持显著。

核心结论

  • 两种折射型EDF IOL均表现出扩展的功能性视力范围、保留的对比敏感度和极少的光现象。
  • 主要终点UIVA观察到统计学显著差异,支持AN6V。

研究设计与样本

  • 前瞻性、随机、患者和检查者设盲的个体内比较研究。
  • 共40例(80眼)双侧年龄相关性白内障患者接受超声乳化联合折射型EDF IOL植入,一眼植入DEN00V(Johnson & Johnson Vision),对侧眼植入AN6V(Teleon)。
  • 术后3个月比较单眼CDVA/UDVA、DIVA/UIVA、DNVA/UNVA、屈光结果、离焦曲线、对比敏感度、光晕测量及问卷主观视觉质量。
  • 37例(74眼)纳入分析。

主要发现

  • 4 m平均UDVA、66 cm DIVA和40 cm UDNVA分别为DEN00V组0.05±0.10、0.17±0.11、0.31±0.11 logMAR,AN6V组0.06±0.10、0.15±0.10、0.32±0.11 logMAR,均无显著差异。
  • 主要终点UIVA在DEN00V组为0.18±0.13 logMAR,AN6V组为0.12±0.09 logMAR,差异有统计学意义。
  • CDVA在DEN00V组为0.01±0.07 logMAR,AN6V组为-0.02±0.07 logMAR(P = .01);UNVA在DEN00V组为0.32±0.12 logMAR,AN6V组为0.25±0.10 logMAR(P = .01)。
  • 明视和暗视条件下对比敏感度显示名义差异,但经多重比较校正后未保持统计学显著。

临床怎么用

  • 对于希望获得扩展视力范围的双侧白内障患者,两种折射型EDF IOL均可考虑。
  • 若优先考虑中距离视力(UIVA),AN6V在主要终点上表现更优。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To compare intraindividual differences in visual performance of two refractive extended-depth-of-focus (EDF) intraocular lenses (IOLs) targeted for emmetropia. DESIGN: Prospective, randomized, patient- and examiner-masked intraindividual comparative study. METHODS: A total of 40 patients (80 eyes) with bilateral age-related cataract underwent phacoemulsification with implantation of refractive EDF IOL in each eye: a wavefront-shaped refractive EDF IOL (DEN00V, Johnson & Johnson Vision) in one eye and a sector-segmented EDF IOL (AN6V, Teleon) in the fellow eye. After 3 months, monocular distance- corrected and uncorrected distance (CDVA/UDVA) intermediate (DIVA/UIVA) and near (DNVA/UNVA) visual acuity, refractive outcomes, defocus curves, contrast sensitivity, halometry, and subjective visual quality using questionnaires were compared. RESULTS: In 37 patients (74 eyes), mean UDVA at 4 m, DIVA at 66 cm and UDNVA at 40 cm were 0.05 ± 0.10, 0.17 ± 0.11, and 0.31 ± 0.11 logMAR for the DEN00V and 0.06 ± 0.10, 0.15 ± 0.10, and 0.32 ± 0.11 logMAR for the AN6V, respectively, with no significant differences. A statistically significant difference was observed for the primary endpoint, UIVA, with 0.18 ± 0.13 logMAR for the DEN00V and 0.12 ± 0.09 logMAR for the AN6V. Nominal differences were observed for CDVA and UNVA with 0.01 ± 0.07 and 0.32 ± 0.12 logMAR for the DEN00V and -0.02 ± 0.07 (P = .01) and 0.25 ± 0.10 (P = .01) logMAR for the AN6V. Further nominal differences could be shown for contrast sensitivity under photopic and mesopic conditions. However, these differences did not remain statistically significant after adjustment for multiple comparisons. CONCLUSIONS: Both IOLs demonstrated an extended range of functional vision, preserved contrast sensitivity, and minimal photopic phenomena. A statistically significant difference was observed for the primary endpoint, UIVA, favoring the AN6V.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Zeilinger J, Palkovits S, Ruiss M, et al. Intraindividual Comparison of Two Refractive Extended-Depth-of-Focus Intraocular Lens Designs. Am J Ophthalmol. 2026. PMID: 42044763. doi:10.1016/j.ajo.2026.04.014