环曲面亲水性丙烯酸酯人工晶状体旋转稳定性与屈光结局
单中心非对照干预试验纳入120眼(80例)年龄相关性白内障伴术前角膜散光>1.0 D者,行超声乳化植入RayOne EMV Toric IOL。114眼完成6个月随访,术后6个月平均绝对旋转2.44°±2.34°,19眼(16.7%)旋转>5°,92%眼残余散光≤0.50 D。
要点
- 术后1周内旋转稳定性良好,但1周后旋转倾向增加,6个月内16.7%人工晶状体旋转>5°。
- 术前角膜散光1.68±0.66 D降至术后6个月残余散光0.27±0.26 D。
- 92%眼术后6个月残余屈光散光≤0.50 D,屈光结局良好。
核心结论
- RayOne EMV Toric IOL在术后第一周内表现出良好的旋转稳定性,但1周后旋转倾向增加,共19枚(16.7%)旋转超过5°,其中1例从手术结束至6个月旋转超过10°。
- 屈光结局良好,92%眼残余屈光散光≤0.50 D。
研究设计与样本
- 单中心、非对照、干预性临床试验,纳入80例患者共120眼,年龄相关性白内障且术前角膜散光>1.0 D,行超声乳化联合RayOne EMV Toric IOL植入。
- 最终114眼(75例)完成分析,通过手术结束时及术后1小时、1周、1个月、6个月的逆光照明图像评估旋转稳定性,术后6个月评估主觉验光。
主要发现
- 从手术结束至术后6个月平均绝对旋转为2.44°±2.34°(范围0.04至10.60)。
- 各随访间期旋转:手术结束至1小时0.89°±1.10°(0.00至8.37),1小时至1周0.58°±0.49°(0.00至2.76),1周至1个月1.19°±1.45°(0.02至7.17),1个月至6个月1.11°±1.41°(0.00至7.96)。
- 术前角膜散光1.68±0.66 D(1.00至3.99),术后6个月残余屈光散光0.27±0.26 D(0.00至1.25)。
- 19眼(16.7%)人工晶状体旋转超过5°,其中1例从手术结束至6个月旋转超过10°。
- 92%眼残余屈光散光≤0.50 D。
关键数值与推荐
- 术前角膜散光>1.0 D为纳入标准。
- 旋转稳定性以手术结束时为基线,与术后1小时、1周、1个月、6个月逆光照明图像比较。
- 残余屈光散光≤0.50 D的比例为92%。
临床怎么用
- 对于年龄相关性白内障伴角膜散光>1.0 D者,RayOne EMV Toric IOL可提供良好的术后早期旋转稳定性和屈光结局。
- 术后1周后仍需关注人工晶状体旋转风险,部分眼可能旋转>5°甚至>10°,可能影响散光矫正效果。
局限性
- 单中心、非对照设计,无对照组。
- 最终分析114眼,失访6眼(120眼纳入)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess rotational stability and refractive outcomes of a toric hydrophilic acrylic intraocular lens (IOL). DESIGN: Single-center, noncomparative, interventional clinical trial. METHODS: In this study, 120 eyes from 80 patients with age-related cataract and preoperative corneal astigmatism greater than 1.0 diopter (D) were enrolled. Patients underwent phacoemulsification with implantation of a RayOne EMV Toric IOL. Rotational stability was evaluated relative to baseline measurements at the end of surgery (EOS) and compared to retroillumination images acquired at 1 hour (1 h), 1 week (1 w), 1 month (1 m), and 6 months (6 m), postoperatively. Subjective manifest refraction was assessed at the 6 m follow-up visit. RESULTS: A total of 114 eyes from 75 patients were available for the final analysis. Mean absolute rotation from EOS to 6 m was 2.44° ± 2.34° (range: 0.04; 10.60). Rotation between follow-up visits was 0.89° ± 1.10° (0.00; 8.37) from EOS to 1 h, 0.58° ± 0.49° (0.00; 2.76) from 1 h to 1 w, 1.19° ± 1.45° (0.02; 7.17) from 1 w to 1 m, and 1.11° ± 1.41° (0.00; 7.96) from 1 m to 6 m. Preoperative corneal astigmatism was 1.68 ± 0.66 D (1.00; 3.99), which was reduced to a postoperative refractive astigmatism of 0.27 ± 0.26 D (0.00; 1.25) at 6 m. CONCLUSIONS: The RayOne EMV Toric IOL demonstrated good rotational stability within the first postoperative week. However, an increased tendency to rotation after 1 w was observed, leading to a total of 19 (16.7%) IOLs rotating more than 5°, including one case exceeding 10° from EOS to 6 m. Refractive outcomes were favorable, with 92% of eyes achieving ≤0.50 D of residual refractive astigmatism.